A novel gamma-ray detector with submillimeter resolutions using a monolithic MPPC array with pixelized Ce:LYSO and Ce:GGAG crystals

Abstract We have developed a large-area monolithic Multi-Pixel Photon Counter (MPPC) array consisting of 4×4 channels with a three-side buttable package. Each channel has a photosensitive area of 3×3 mm 2 and 3600 Geiger mode avalanche photodiodes (APDs). For typical operational gain of 7.5×10 5 at +20 °C, gain fluctuation over the entire MPPC device is only ± 5.6 % , and dark count rates (as measured at the 1 p.e. level) amount to ≤ 400 kcps per channel. We first fabricated a gamma-ray camera consisting of the MPPC array with one-to-one coupling to a Ce-doped ( Lu , Y ) 2 ( SiO 4 ) O (Ce:LYSO) crystal array (4×4 array of 3×3×10 mm 3 crystals). Energy and time resolutions of 11.5 ± 0.5 % (FWHM at 662 keV) and 493 ± 22 ps were obtained, respectively. When using the charge division resistor network, which compiles signals into four position-encoded analog outputs, the ultimate positional resolution is estimated as 0.19 mm in both X and Y directions, while energy resolution of 10.2 ± 0.4 % (FWHM) was obtained. Finally, we fabricated submillimeter Ce:LYSO and Ce-doped Gd 3 Ga 3 Al 2 O 12 (Ce:GGAG) scintillator matrices each consisting of 1.0×1.0, 0.7×0.7 and 0.5×0.5 mm 2 pixels, to further improve the spatial resolution. In all types of Ce:LYSO and Ce:GGAG matrices, each crystal was clearly resolved in the position histograms when irradiated by a 137 Cs source. The energy resolutions for 662 keV gamma-rays for each Ce:LYSO and Ce:GGAG scintillator matrix were ≤ 14.3 % . These results suggest excellent potential for its use as a high spatial medical imaging device, particularly in positron emission tomography (PET).

[1]  K. Kamada,et al.  Composition Engineering in Cerium-Doped (Lu,Gd)3(Ga,Al)5O12 Single-Crystal Scintillators , 2011 .

[2]  P. J. Simpson,et al.  Proof-of-principle study of a small animal PET/field-cycled MRI combined system using conventional PMT technology , 2010 .

[3]  S. Dolinsky,et al.  Multi-Pixel Photon Counters for TOF PET Detector and Its Challenges , 2009, IEEE Transactions on Nuclear Science.

[4]  Eiji Sugiyama,et al.  Design and performance from an integrated PET/MRI system for small animals , 2010, Annals of nuclear medicine.

[5]  Tadashi Watabe,et al.  Development of a Si-PM-based high-resolution PET system for small animals , 2010, Physics in medicine and biology.

[6]  David J. Schlyer,et al.  Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph , 2007 .

[7]  I. Kandarakis,et al.  Evaluation of the light emission efficiency of LYSO:Ce scintillator under X-ray excitation for possible applications in medical imaging , 2006 .

[8]  D. Wolski,et al.  Multi Pixel Photon Counters (MPPC) as an Alternative to APD in PET Applications , 2010, IEEE Transactions on Nuclear Science.

[9]  Simon R. Cherry,et al.  Simple charge division readouts for imaging scintillator arrays using a multi-channel PMT , 1995 .

[10]  H. Ikeda,et al.  Development of an APD-Based PET Module and Preliminary Resolution Performance of an Experimental Prototype Gantry , 2010, IEEE Transactions on Nuclear Science.

[11]  K. Kamada,et al.  Development of a gamma-ray imager using a large area monolithic 4 × 4 MPPC array for a future PET scanner , 2012 .

[12]  H. Ikeda,et al.  Development of a large-area monolithic 4×4 MPPC array for a future PET scanner employing pixelized Ce:LYSO and Pr:LuAG crystals , 2011 .

[13]  William W. Moses,et al.  Trends in PET imaging , 2001 .

[14]  R. Pani,et al.  Preliminary study of different readout strategies for a positron emission mammograph head , 2006 .

[15]  S. Siegel,et al.  A sub-millimeter resolution PET detector module using a multi-pixel photon counter array , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[16]  Manuel Desco,et al.  Performance evaluation of SiPM detectors for PET imaging in the presence of magnetic fields , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[17]  Randy Wojcik,et al.  Initial tests of a prototype MRI-compatible PET imager , 2006 .