Novel Hand-Held Device for Exhaled Nitric Oxide-Analysis in Research and Clinical Applications

Changes in expired nitric oxide (NO) occur in airway inflammation and have proved to be important in the monitoring of inflammatory disease processes such as asthma. We set out to develop a novel hand-held NO-analyzer with a performance comparable to the present more costly and complex chemiluminescence instruments. The new device is based on a specially designed electrochemical sensor, where we have developed a novel sampling and analysis technology, compensating for the relatively slow response properties of the electrochemical sensor technique. A Lowest Detection Limit in NO-analysis from reference gas tests of less than 3 ppb and a response time of 15 seconds together with an average precision in human breath measurements of 1.4 ppb were obtained. We also show an agreement with the existing 'gold standard' FENO measurement technique, within 0.5 ppb in a group of 19 subjects together with a high linearity and accuracy compared to reference gases. The new analyzer enables affordable monitoring of inflammatory airway diseases in research and routine clinical practice.

[1]  Forthcoming articles in International Journal of Clinical Monitoring and Computing , 1985 .

[2]  J. Cowan,et al.  Diagnosing asthma: comparisons between exhaled nitric oxide measurements and conventional tests. , 2004, American journal of respiratory and critical care medicine.

[3]  S Moncada,et al.  Endogenous nitric oxide is present in the exhaled air of rabbits, guinea pigs and humans. , 1991, Biochemical and biophysical research communications.

[4]  I. Adcock,et al.  Relationship between exhaled nitric oxide and mucosal eosinophilic inflammation in children with difficult asthma, after treatment with oral prednisolone. , 2001, American journal of respiratory and critical care medicine.

[5]  P. Biban,et al.  Effect of atmospheric nitric oxide (NO) on measurements of exhaled NO in asthmatic children , 1998, Pediatric pulmonology.

[6]  A. B. Robinson,et al.  Quantitative analysis of urine vapor and breath by gas-liquid partition chromatography. , 1971, Proceedings of the National Academy of Sciences of the United States of America.

[7]  S. Kharitonov,et al.  Exhaled and nasal nitric oxide measurements : recommendations , 1997 .

[8]  B. Krotoszynski,et al.  Characterization of human expired air: a promising investigative and diagnostic technique. , 1977, Journal of chromatographic science.

[9]  Arthur S Slutsky,et al.  Marked flow-dependence of exhaled nitric oxide using a new technique to exclude nasal nitric oxide. , 1997, American journal of respiratory and critical care medicine.

[10]  P. Barnes,et al.  Reproducibility of exhaled nitric oxide measurements in healthy and asthmatic adults and children , 2003, European Respiratory Journal.

[11]  M. Ennis,et al.  Exhaled nitric oxide correlates with airway eosinophils in childhood asthma , 2002, Thorax.

[12]  Joseph R. Stetter,et al.  The properties and applications of amperometric gas sensors , 1992 .

[13]  E. Weitzberg,et al.  Increased amount of nitric oxide in exhaled air of asthmatics. , 1993, The European respiratory journal.