HeartWare HVAD Flow Estimator Accuracy for Left and Right Ventricular Support.

This study investigated the accuracy of the HeartWare HVAD flow estimator for left ventricular assist device (LVAD) support and biventricular assist device (BiVAD) support for modes of reduced speed (BiVAD-RS) and banded outflow (BiVAD-B). The HVAD flow estimator was evaluated in a mock circulatory loop under changes in systemic and pulmonary vascular resistance, heart rate, central venous pressure, and simulated hematocrit (correlated to viscosity). A difference was found between mean estimated and mean measured flow for LVAD (0.1 ± 0.3 L/min), BiVAD-RS (-0.1 ± 0.2 L/min), and BiVAD-B (0 ± 0.2 L/min). Analysis of the flow waveform pulsatility showed good correlation for LVAD (r2 = 0.98) with a modest spread in error (0.7 ± 0.1 L/min), while BiVAD-RS and BiVAD-B showed similar spread in error (0.7 ± 0.3 and 0.7 ± 0.2 L/min, respectively), with much lower correlation (r2 = 0.85 and r2 = 0.60, respectively). This study demonstrated that the mean flow error of the HVAD flow estimator is similar when the device is used in LVAD, BiVAD-RS, or BiVAD-B configuration. However, the instantaneous flow waveform should be interpreted with caution, particularly in the cases of BiVAD support.

[1]  T. Kolias,et al.  Doppler-derived dP/dt and -dP/dt predict survival in congestive heart failure. , 2000, Journal of the American College of Cardiology.

[2]  Robert L Kormos,et al.  Seventh INTERMACS annual report: 15,000 patients and counting. , 2015, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[3]  Einly Lim,et al.  Theoretical foundations of a Starling-like controller for rotary blood pumps. , 2012, Artificial organs.

[4]  P. Macdonald,et al.  Beat-to-beat detection of aortic valve opening in HeartWare left ventricular assist device patients. , 2018, Artificial organs.

[5]  George M Pantalos,et al.  Intraoperative evaluation of the HeartMate II flow estimator. , 2009, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[6]  T. Pirbodaghi,et al.  Left Ventricular Volume Unloading with Axial and Centrifugal Rotary Blood Pumps , 2015, ASAIO journal.

[7]  U. Steinseifer,et al.  An advanced mock circulation loop for in-vitro cardiovascular device evaluation. , 2020, Artificial organs.

[8]  Ashish S Shah,et al.  Right heart dysfunction after left ventricular assist device implantation: a comparison of the pulsatile HeartMate I and axial-flow HeartMate II devices. , 2008, The Annals of thoracic surgery.

[9]  Axel Haverich,et al.  Successful use of the HeartWare HVAD rotary blood pump for biventricular support. , 2010, The Journal of thoracic and cardiovascular surgery.

[10]  M. Wheeler,et al.  Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device , 2017, World journal of cardiology.

[11]  A. Cheung,et al.  American Association for Thoracic Surgery/International Society for Heart and Lung Transplantation guidelines on selected topics in mechanical circulatory support. , 2020, The Journal of thoracic and cardiovascular surgery.

[12]  James F. Antaki,et al.  A Control System for Rotary Blood Pumps Based on Suction Detection , 2009, IEEE Transactions on Biomedical Engineering.

[13]  R. Higgins,et al.  The Society of Thoracic Surgeons Intermacs database annual report: Evolving indications, outcomes, and scientific partnerships. , 2019, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[14]  D. Zimpfer,et al.  Continuous monitoring of cardiac rhythms in left ventricular assist device patients. , 2014, Artificial organs.

[15]  S. Schulte-Eistrup,et al.  Changes in Total Cardiac Output and Oxygen Extraction During Exercise in Patients Supported With an HVAD Left Ventricular Assist Device , 2018, Artificial organs.

[16]  Robert L Kormos,et al.  Eighth annual INTERMACS report: Special focus on framing the impact of adverse events. , 2017, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[17]  Shunsuke Saito,et al.  Biventricular support using implantable continuous-flow ventricular assist devices. , 2011, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[18]  D. Burkhoff,et al.  Accuracy of the HVAD Pump Flow Estimation Algorithm , 2016, ASAIO journal.

[19]  Martin Schweiger,et al.  Biventricular Circulatory Support With Two Miniaturized Implantable Assist Devices , 2011, Circulation.

[20]  H. Schima,et al.  Suction events during left ventricular support and ventricular arrhythmias. , 2007, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[21]  RolandHetzer,et al.  Biventricular Circulatory Support With Two Miniaturized Implantable Assist Devices , 2011 .

[22]  Roland Hetzer,et al.  Long-term biventricular support with the heartware implantable continuous flow pump. , 2010, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[23]  Daniel L. Timms,et al.  Replication of the Frank-Starling response in a mock circulation loop , 2011, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[24]  D. Humphreys,et al.  Longitudinal structural, functional, and cellular myocardial alterations with chronic centrifugal continuous-flow left ventricular assist device support. , 2017, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[25]  Victor L Poirier,et al.  Design Features, Developmental Status, and Experimental Results With the Heartmate III Centrifugal Left Ventricular Assist System With a Magnetically Levitated Rotor , 2007, ASAIO journal.

[26]  S. Russell,et al.  Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. , 2010, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[27]  J. Teuteberg,et al.  Right ventricular failure after left ventricular assist devices. , 2015, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[28]  F. Casas,et al.  Development of a pump flow estimator for rotary blood pumps to enhance monitoring of ventricular function. , 2012, Artificial organs.

[29]  J. Michael Textbook of Medical Physiology , 2005 .