OBJECTIVES
Focused assessment with sonography in trauma (FAST) can define life-threatening injuries in austere settings with remote real-time review by experienced physicians. This study evaluates vest-mounted microwave, satellite, and LifeLink communications technology for image clarity and diagnostic accuracy during remote transmission of FAST examinations.
METHODS
Using a SonoSite, FAST was obtained on three patients with pericardial and intraperitoneal effusions and two control subjects in a remotely located U.S. Army Combat Support Hospital. A miniature vest-mounted video transmitter attached to the SonoSite sent wireless ultrasound video 20 m to a receiving antenna. The signal was then transferred over VSAT satellite systems at 512 kilobaud per second (kbps), INMARSAT satellite systems at 64 kbps, and over LifeLink on a moving ambulance through a metropolitan wireless traffic-management network. Clarity and absence or presence of effusions were recorded by 15 staff emergency physicians.
RESULTS
Average sensitivity, specificity, and accuracy were 87% (95% confidence interval [CI]=79% to 95%), 85% (95% CI=81% to 89%), and 86% (95% CI=82% to 90%) for the Premier Wireless Vest; 98% (95% CI=97% to 99%), 83% (95% CI=75% to 91%), and 86% (95% CI=82% to 90%) for VSAT; 95% (95% CI=94% to 96%), 70% (95% CI=58% to 82%), and 75% (95% CI=70% to 80%) for INMARSAT; and 82% (95% CI=73% to 91%), 83% (95% CI=74% to 92%), and 82% (95% CI=78% to 86%) for LifeLink with clarity of 3.0 (95% CI=2.7 to 3.3), 2.9 (95% CI=2.6 to 3.2), 1.3 (95% CI=1.2 to 1.4), and 2.1 (95% CI=1.8 to 2.4), respectively.
CONCLUSIONS
Accuracy correlated with clarity. Roaming vest transmission of FAST provides interpretable, diagnostic imagery at the distances used in this study. VSAT provided the best clarity and diagnostic value with the lighter, more portable INMARSAT serving a lesser role for remote clinical interpretation. LifeLink performed well, and further infrastructure improvements may increase clarity and accuracy.
[1]
B. Durham.
Emergency medicine physicians saving time with ultrasound.
,
1996,
The American journal of emergency medicine.
[2]
G. Rozycki,et al.
Surgeon-performed ultrasound for the assessment of truncal injuries: lessons learned from 1540 patients.
,
1998,
Annals of surgery.
[3]
Bernard J. Rubal,et al.
W IRELESS AND S ATELLITE T RANSMISSION OF P REHOSPITAL F OCUSED A BDOMINAL S ONOGRAPHY FOR T RAUMA
,
2003
.
[4]
E. Ruiz,et al.
Emergency department echocardiography improves outcome in penetrating cardiac injury.
,
1992,
Annals of emergency medicine.