Incoherent-scatter computed tomography with monochromatic synchrotron x ray: feasibility of multi-CT imaging system for simultaneous measurement-of fluorescent and incoherent scatter x rays

We describe a new system of incoherent scatter computed tomography (ISCT) using monochromatic synchrotron X rays, and we discuss its potential to be used in in vivo imaging for medical use. The system operates on the basis of computed tomography (CT) of the first generation. The reconstruction method for ISCT uses the least squares method with singular value decomposition. The research was carried out at the BLNE-5A bending magnet beam line of the Tristan Accumulation Ring in KEK, Japan. An acrylic cylindrical phantom of 20-mm diameter containing a cross-shaped channel was imaged. The channel was filled with a diluted iodine solution with a concentration of 200 /spl mu/gI/ml. Spectra obtained with the system's high purity germanium (HPGe) detector separated the incoherent X-ray line from the other notable peaks, i.e., the iK/sub /spl alpha// and K/sub /spl beta/1/ X-ray fluorescent lines and the coherent scattering peak. CT images were reconstructed from projections generated by integrating the counts In the energy window centering around the incoherent scattering peak and whose width was approximately 2 keV. The reconstruction routine employed an X-ray attenuation correction algorithm. The resulting image showed more homogeneity than one without the attenuation correction.

[1]  R. L. Clarke,et al.  A new method for measurement of bone mineral content using both transmitted and scattered beams of gamma-rays , 1973, Physics in medicine and biology.

[2]  Geoffrey Harding,et al.  X-ray imaging with Compton-scatter radiation , 1983 .

[3]  Charles L. Lawson,et al.  Solving least squares problems , 1976, Classics in applied mathematics.

[4]  J. Kosanetzky,et al.  Scattered X-ray beam nondestructive testing , 1989 .

[5]  U Neitzel,et al.  Coherent scatter in radiographic imaging: a Monte Carlo simulation study. , 1985, Physics in medicine and biology.

[6]  U Neitzel,et al.  Elastic scatter computed tomography. , 1985, Physics in medicine and biology.

[7]  D N Slatkin,et al.  Single-and dual-energy CT with monochromatic synchrotron x-rays. , 1997, Physics in medicine and biology.

[8]  M M Goodsitt,et al.  The Composition of Bone Marrow for a Dual-Energy Quantitative Computed Tomography Technique: A Cadaver and Computer Simulation Study , 1994, Investigative radiology.

[9]  K Doi,et al.  The validity of Monte Carlo simulation in studies of scattered radiation in diagnostic radiology. , 1983, Physics in medicine and biology.

[10]  A Fenster,et al.  Tomographic imaging of the angular-dependent coherent-scatter cross section. , 1997, Medical physics.

[11]  P. G. Lale The examination of internal tissues, using gamma-ray scatter with a possible extension to megavoltage radiography. , 1959, Physics in medicine and biology.

[12]  U Neitzel,et al.  X-ray diffraction measurements of some plastic materials and body tissues. , 1987, Medical physics.

[13]  F. T. Farmer,et al.  A new approach to the determination of anatomical cross-sections of the body by Compton scattering of gamma-rays. , 1971, Physics in medicine and biology.

[14]  Tetsuya Yuasa,et al.  Reconstruction method for fluorescent X-ray computed tomography by least-squares method using singular value decomposition , 1997 .

[15]  J. H. Hubbell,et al.  Atomic form factors, incoherent scattering functions, and photon scattering cross sections , 1975 .

[16]  U Neitzel,et al.  X-ray diffraction computed tomography. , 1987, Medical physics.

[17]  P. G. Lale The examination of internal tissues by high-energy scattered x-radiation. , 1968, Radiology.

[18]  A. L. Hanson,et al.  Interaction of keV photons with matter and new applications , 1992 .

[19]  G. Harding,et al.  Status and outlook of coherent-x-ray scatter imaging. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[20]  W. Heitler,et al.  The quantum theory of radiation , 1936 .

[21]  Tetsuya Yuasa,et al.  Fluorescent x-ray computed tomography with synchrotron radiation using fan collimator , 1996, Medical Imaging.