The use of on-line image verification to estimate the variation in radiation therapy dose delivery.

PURPOSE On-line radiotherapy imaging systems provide data that allow us to study the geometric nature of treatment variation. It is more clinically relevant to examine the resultant dosimetric variation. In this work, daily beam position as recorded by the on-line images is used to recalculate the treatment plan to show the effect geometric variation has on dose. METHODS AND MATERIALS Daily 6 MV or 18 MV x-ray portal images were acquired using a fiberoptic on-line imaging system for 12 patients with cancers in the head and neck, thoracic, and pelvic regions. Each daily on-line portal image was aligned with the prescription simulation image using a template of anatomical structures defined on the latter. The outline of the actual block position was then superimposed on the prescription image. Daily block positions were cumulated to give a summary image represented by the block overlap isofrequency distribution. The summary data were used to analyze the amount of genometric variation relative to the prescription boundary on a histogram distribution plot. Treatment plans were recalculated by considering each aligned portal image as an individual beam. RESULTS On-Line Image Verification (OLIV) data can differentiate between systematic and random errors in a course of daily radiation therapy. The data emphasize that the type and magnitude of patient set-up errors are unique for individual patients and different clinical situations. Head and neck sites had the least random variation (average 0-100% block overlap isofrequency distribution width = 7 mm) compared to thoracic (average 0-100% block overlap isofrequency distribution width = 12 mm) or pelvic sites (average 0-100% block overlap isofrequency distribution width = 14 mm). When treatment delivery is analyzed case by case, systematic as well as random errors are represented. When the data are pooled by anatomical site, individuality of variations is lost and variation appears random. Recalculated plans demonstrated dosimetric deviations from the original plans. The differences between the two dosimetric distributions were emphasized using a technique of plan subtraction. This allowed quick identification of relative "hot and cold spots" in the recalculated plans. The magnitude and clinical significance of dosimetric variation was unique for each patient. CONCLUSIONS OLIV data are used to study geometric uncertainties because of the unique nature for individual patients. Dose recalculation is helpful to illustrate the dosimetric consequences of set-up errors.

[1]  M. Goitein,et al.  Accuracy of radiation field alignment in clinical practice. , 1985, International journal of radiation oncology, biology, physics.

[2]  T E Schultheiss,et al.  Can modest escalations of dose be detected as increased tumor control? , 1992, International journal of radiation oncology, biology, physics.

[3]  A G Haus,et al.  The value of frequent treatment verification films in reducing localization error in the irradiation of complex fields , 1976, Cancer.

[4]  M. Williams,et al.  Dose-response relationships for human tumors: implications for clinical trials of dose modifying agents. , 1984, International journal of radiation oncology, biology, physics.

[5]  J. van de Geijn,et al.  EXTDOS 71. Revised and expanded version of EXTDOS, a program for treatment planning in external beam therapy. , 1972 .

[6]  A Brahme,et al.  Dosimetric precision requirements in radiation therapy. , 1984, Acta radiologica. Oncology.

[7]  J. Cox,et al.  Weekly localization films and detection of field placement errors. , 1978, International journal of radiation oncology, biology, physics.

[8]  M Goitein,et al.  Calculation of the uncertainty in the dose delivered during radiation therapy. , 1985, Medical physics.

[9]  G E Hanks,et al.  Patterns of care study: Hodgkin's disease relapse rates and adequacy of portals , 1983, Cancer.

[10]  R Mohan,et al.  The role of uncertainty analysis in treatment planning. , 1991, International journal of radiation oncology, biology, physics.

[11]  M W Vannier,et al.  A method to analyze 2-dimensional daily radiotherapy portal images from an on-line fiber-optic imaging system. , 1991, International journal of radiation oncology, biology, physics.

[12]  J. W. Epstein,et al.  On-line radiotherapy imaging with an array of fiber-optic image reducers. , 1990, International journal of radiation oncology, biology, physics.

[13]  J. Leong,et al.  Implementation of random positioning error in computerised radiation treatment planning systems as a result of fractionation. , 1987, Physics in medicine and biology.

[14]  J W Wong,et al.  Study of treatment variation in the radiotherapy of head and neck tumors using a fiber-optic on-line radiotherapy imaging system. , 1991, International journal of radiation oncology, biology, physics.

[15]  A G Haus,et al.  Localization error in the radiotherapy of Hodgkin's disease and malignant lymphoma with extended mantle fields , 1974, Cancer.