Impact of GEM foil hole geometry on GEM detector gain

Detailed 3D imaging of Gas Electron Multiplier (GEM) foil hole geometry was realized. Scanning White Light Interferometry was used to examine six topological parameters of GEM foil holes from both sides of the foil. To study the effect of the hole geometry on detector gain, the ANSYS and Garfield ++ software were employed to simulate the GEM detector gain on the basis of SWLI data. In particular, the effective gain in a GEM foil with equally shaped holes was studied. The real GEM foil holes exhibited a 4% lower effective gain and 6% more electrons produced near the exit electrode of the GEM foil than the design anticipated. Our results indicate that the GEM foil hole geometry affects the gain performance of GEM detectors.

[1]  G. Bencivenni,et al.  The operational experience of the triple-GEM detectors of the LHCb muon system: Summary of 2 years of data taking , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[2]  L. Ropelewski,et al.  Performance of GEM detectors in high intensity particle beams , 2001 .

[3]  R. Orava,et al.  Foil geometry effects on GEM characteristics , 2001 .

[4]  Edward Hæggström,et al.  Quasidynamic calibration of stroboscopic scanning white light interferometer with a transfer standard , 2013 .

[5]  R. Veenhof,et al.  Penning transfer in argon-based gas mixtures , 2010 .

[6]  M. Posik,et al.  Optical and electrical performance of commercially manufactured large GEM foils , 2015 .

[7]  L. Fernandes,et al.  Operation of a single-GEM in noble gases at high pressures , 2007 .

[8]  T. Hildén,et al.  Study of GEM-foil defects with optical scanning system , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[9]  Fabio Sauli,et al.  The gas electron multiplier (GEM): Operating principles and applications , 2016 .

[10]  S. Bachmanna,et al.  Performance of GEM detectors in high intensity particle beams , 2001 .

[11]  U. Becker,et al.  Test and evaluation of new GEMs with an automatic scanner , 2006 .

[12]  High pressure operation of the triple-GEM detector in pure Ne, Ar and Xe , 2001, physics/0103082.

[13]  Leszek Ropelewski,et al.  EUROPEAN LABORATORY FOR PARTICLE PHYSICS CERN-EP/99-48 12.04.1999 CHARGE AMPLIFICATION AND TRANSFER PROCESSES IN THE GAS ELECTRON MULTIPLIER , 2001 .

[14]  F. Sauli GEM: A new concept for electron amplification in gas detectors , 1997 .

[15]  E. Hæggström,et al.  Calibrating an optical scanner for quality assurance of large area radiation detectors , 2014 .

[16]  R. Turpeinen,et al.  Optical quality assurance of GEM foils , 2015 .