Analysis of the Cyclotron Facility Calibration and Aircraft Results Obtained by LIULIN-3M Instrument

The LIULIN-3M instrument is a further development of the LIULIN dosimeter-radiometer, which has been used on the NffR space station in the 1988-1994 time period, The LIULIN-3M is designed for continuous monitoring of the radiation environment during the BION-12 satellite flight in 1999. A semiconductor detector with 1 mm thickness and 1 cm(exp 2) area is used in the instrument. Pulse high analysis technique is used for measurement of the energy losses in the detector. The final data sets from the instrument are the flux and the dose rate for the exposition time and 256 channels of LET spectra if a non-nal coincidence of the particles to the detector is considered. The LIULIN-3M instrument was calibrated by proton fluxes with different energies at the Indiana University Cyclotron Facility in June 1997 and was used for space radiation measurements during commercial aircraft flights. Obtained calibration and flight results are analyzed in the paper.

[1]  K O'Brien,et al.  Atmospheric cosmic rays and solar energetic particles at aircraft altitudes. , 1996, Environment international.

[2]  Dachev TsP,et al.  Space radiation dosimetry with active detections for the scientific program of the second Bulgarian cosmonaut on board the Mir space station. , 1989, Advances in space research : the official journal of the Committee on Space Research.