A method for using Josephson voltage standards for direct characterization of high performance digitizers to establish AC voltage and current traceability to SI

A method for traceability to SI for ac voltage and current based on high performance digitizers is presented. In contrast to the existing thermal-based methods, the proposed method utilizes direct traceability to quantum-based waveforms via the use of Josephson voltage systems. This allows not only a simplification of the traceability chain and reduced measurement times but also offers the potential for analysis of the ac voltage and current waveform spectral content, a feature which is not possible using thermal methods. Scaling of current and voltage is achieved by the use of current shunts and resistive voltage dividers respectively. Target operating ranges are up to 1 A and 100 V with a frequency range up to 1 kHz for both. The corresponding target uncertainty for this traceability route is 1 μV V−1 and 2 μA A−1 up to frequencies of 1 kHz. The traceability chain is described and various components are characterized to validate their suitability for this task. It is demonstrated that these uncertainty targets can be met under certain conditions. The use of multi-tone calibration waveforms is investigated to further reduce measurement time. An uncertainty analysis method based on simulation using real component performance data is demonstrated.

[1]  Mun-Seog Kim,et al.  Differential sampling of AC waveforms based on a programmable Josephson voltage standard using a high-precision sampler , 2021, Metrologia.

[2]  Petar Mostarac,et al.  The AC Amplitude Measurement Characteristics of High-resolution Digitizers based on Calibration with Thermal Voltage Converter and Swerlein Algorithm , 2021, 2021 IEEE International Instrumentation and Measurement Technology Conference (I2MTC).

[3]  R. Behr,et al.  An AC quantum voltmeter for frequencies up to 100 kHz using sub-sampling , 2021, Metrologia.

[4]  N. Flowers-Jacobs,et al.  Calibration of an AC Voltage Source Using a Josephson Arbitrary Waveform Synthesizer at 4 V , 2020, 2020 Conference on Precision Electromagnetic Measurements (CPEM).

[5]  M. Sira Noise and stability of Fluke 8588A in digitizing mode , 2020, Conference on Precision Electromagnetic Measurements.

[6]  Bryan Waltrip,et al.  Comparison of AC Power Referenced to Either PJVS or JAWS , 2020, 2020 Conference on Precision Electromagnetic Measurements (CPEM).

[7]  Ali Tangel,et al.  Using Programmable Josephson Voltage Standard for Static and Dynamic Gain Characterization of Integrating ADC , 2020, IEEE Transactions on Instrumentation and Measurement.

[8]  Mun-Seog Kim,et al.  Measurement configurations for differential sampling of AC waveforms based on a programmable Josephson voltage standard: effects of sampler bandwidth on the measurements , 2020, Metrologia.

[9]  Roman Malarić,et al.  Comparison of Digitizers for High-Precision Sampling Power Meters , 2020, IEEE Transactions on Instrumentation and Measurement.

[10]  Ralf Behr,et al.  Characterization of a precision modular sinewave generator , 2020, Measurement Science and Technology.

[11]  Samuel P. Benz,et al.  AC Voltage Measurements to 120 V With a Josephson Arbitrary Waveform Synthesizer and an Inductive Voltage Divider , 2019, IEEE Transactions on Instrumentation and Measurement.

[12]  Martin Sira,et al.  A Novel Method for Calibration of ADC Using JAWS , 2019, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[13]  R. Behr,et al.  Characterization of an analog-to-digital converter frequency response by a Josephson arbitrary waveform synthesizer , 2019, Measurement Science and Technology.

[14]  Alain Rüfenacht,et al.  Impact of the latest generation of Josephson voltage standards in ac and dc electric metrology , 2018, Metrologia.

[15]  R. Behr,et al.  The AC Quantum Voltmeter Used for AC Current Calibrations , 2018, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[16]  M. F. Beug,et al.  Calibration of an Inductive Voltage Divider Using Pulse-Driven Josephson Arrays , 2018, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[17]  G.W.C. Wijayasundara,et al.  Calibration of AC Voltage for a High Speed Sampling ADC Board Using Thermal Voltage Converters , 2018, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[18]  F. Sandoval,et al.  Study of Keysight 3458A Temperature Coefficient for Different Aperture Times in DCV Sampling Mode , 2018, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[19]  M. Sira,et al.  Precision Buffer with Low Input Capacitance , 2018, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[20]  M. Sira,et al.  Uncertainty of Phasor Measurement Unit Calculated by Means of Monte Carlo Method , 2018, 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018).

[21]  S. Ertürk,et al.  Metrological measurements using programmable Josephson Voltage Standard , 2017, 2017 10th International Conference on Electrical and Electronics Engineering (ELECO).

[22]  Rado Lapuh,et al.  Keysight 3458A Noise Performance in DCV Sampling Mode , 2017, IEEE Transactions on Instrumentation and Measurement.

[23]  Helko E. van den Brom,et al.  AC–DC Calibrations With a Pulse-Driven AC Josephson Voltage Standard Operated in a Small Cryostat , 2017, IEEE Transactions on Instrumentation and Measurement.

[24]  Shih-Fang Chen,et al.  Calibration of inductive voltage dividers at power frequencies using an AC-PJVS , 2016, 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016).

[25]  Stanislav Maslan,et al.  QWTB — Software tool box for sampling measurements , 2016, 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016).

[26]  M. Sira,et al.  Uncertainty of the signal parameter estimation from sampled data , 2016, Conference on Precision Electromagnetic Measurements.

[27]  R. Lapuh,et al.  Stable arbitrary waveform generator as a transfer standard for ADC calibration , 2016, 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016).

[28]  R. Behr,et al.  Towards a 1 V Josephson Arbitrary Waveform Synthesizer , 2015, IEEE Transactions on Applied Superconductivity.

[29]  Stanislav Maslan,et al.  A validation of a THD measurement equipment with a 24-bit digitizer , 2014, 29th Conference on Precision Electromagnetic Measurements (CPEM 2014).

[30]  Ralf Behr,et al.  An ac quantum voltmeter based on a 10 V programmable Josephson array , 2013 .

[31]  Yasutaka Amagai,et al.  Low-Frequency Characterization in Thermal Converters Using AC-Programmable Josephson Voltage Standard System , 2013, IEEE Transactions on Instrumentation and Measurement.

[32]  R. Behr,et al.  Precision Comparison of Sine Waveforms With Pulse-Driven Josephson Arrays , 2013, IEEE Transactions on Applied Superconductivity.

[33]  Takahiro Yamada,et al.  Application of a Josephson quantum voltage source to the measurement of microsecond timescale settling time on the Agilent 3458A 8 digit voltmeter , 2012 .

[34]  Ralf Behr,et al.  Quantum-referenced voltage waveform synthesiser , 2011 .

[35]  Dimitrios Georgakopoulos,et al.  Precision AC–DC Difference Measurement System Based on a Programmable Josephson Voltage Standard , 2011, IEEE Transactions on Instrumentation and Measurement.

[36]  Paul S. Wright,et al.  Characterization of Metrological Grade Analog-to-Digital Converters Using a Programmable Josephson Voltage Standard , 2011, IEEE Transactions on Instrumentation and Measurement.

[37]  Gert Rietveld,et al.  Characterization of a Wideband Digitizer for Power Measurements up to 1 MHz , 2010, IEEE Transactions on Instrumentation and Measurement.

[38]  R. Iuzzolino,et al.  Sub-$\mu{\rm m}$ SNS Josephson Junction Arrays for the Josephson Arbitrary Waveform Synthesizer , 2009, IEEE Transactions on Applied Superconductivity.

[39]  Harald Moser,et al.  Primary AC Power Standard Based on Programmable Josephson Junction Arrays , 2007, IEEE Transactions on Instrumentation and Measurement.

[40]  S. Benz,et al.  Resonance-free low-pass filters for the AC Josephson voltage standard , 2006, IEEE Transactions on Applied Superconductivity.

[41]  A. Sosso,et al.  Application of DMM Linearity to dc Volt Traceability , 2004, 2004 Conference on Precision Electromagnetic Measurements.

[42]  Samuel P. Benz,et al.  AC coupling technique for Josephson waveform synthesis , 2001 .

[43]  D. Reymann,et al.  Using power spectra and Allan variances to characterise the noise of Zener-diode voltage standards , 2000 .

[44]  A. Sosso,et al.  Calibration of multimeters as voltage ratio standards , 2000, Conference on Precision Electromagnetic Measurements. Conference Digest. CPEM 2000 (Cat. No.00CH37031).

[45]  Harald Moser,et al.  A new scheme for generating and measuring active, reactive, and apparent power at power frequencies with uncertainties of 2.5×10 -6 , 1998, IEEE Trans. Instrum. Meas..

[46]  Samuel P. Benz,et al.  A pulse‐driven programmable Josephson voltage standard , 1996 .

[47]  J. J. Snyder,et al.  An Ultra-High Resolution Frequency Meter , 1981 .

[48]  E. S. Williams Thermal voltage converters and comparator for very accurate ac voltage measurements , 1971 .

[49]  J. J. Hill,et al.  Theory, design and measurement of inductive voltage dividers , 1968 .

[50]  Mise en pratique for the definition of the ampere and other electric units in the SI , 2018 .

[51]  Ralf Behr,et al.  Characterization of a high-resolution analog-to-digital converter with a Josephson AC voltage source , 2005, IEEE Transactions on Instrumentation and Measurement.

[52]  E. Iso Guide to the Expression of Uncertainty in Measurement , 1993 .