A novel calibration method incorporating nonlinear optimization and ball‐bearing markers for cone‐beam CT with a parameterized trajectory

PURPOSE Cone-beam (CB) CT is a powerful noninvasive imaging modality, and is widely used in many applications. Accurate geometric parameters are essential for high-quality image reconstruction. Usually, a CBCT system with higher spatial resolution, particularly on the order of microns or nanometers, will be more sensitive to the parametric accuracy. Here, we propose a novel calibration method combining a simple phantom containing ball bearing markers and an advanced optimization procedure. This method can be applied to CBCT with reproducible geometry and frame-to-frame invariant geometric parameters. METHODS Our proposed simplex-simulated annealing procedure minimizes the cost function that associates the geometrical parameters with the degree to which the back projections of the ball bearings in projections from various viewing angles converge, and the global minimum of the cost function corresponds to the actual geometric parameters. Specifically, six geometric parameters can be directly obtained by minimizing the cost function, and the last parameter, the distance from source to rotation axis (SRD), can be obtained using prior knowledge of the phantom - the spacing between the two ball bearings. RESULTS Numerical simulation was performed to validate that the proposed method with various noise levels. With the proposed method, the mean errors and standard deviations can be reduced to ∼10% and less than 1/3 of a competing benchmark method in the case of strong Gaussian noise (sigma = 200% of the pixel size) and large tilt angle (tilt angle =  - 4 ∘ ). The calibration experiments with micro-CT and high-resolution CT scanners demonstrate that the proposed method recovers imaging parameters accurately, leading to superior image quality. CONCLUSION The proposed method can obtain accurate geometric parameters of a CBCT system with a circular trajectory. While in the case of micro-CT the proposed method has a performance comparable to the competing method, for high-resolution CT, which is more sensitive to geometric calibration, the proposed method demonstrates higher calibration accuracy and more robustness than the benchmark algorithm.

[1]  F. Noo,et al.  Direct determination of geometric alignment parameters for cone-beam scanners. , 2009, Physics in medicine and biology.

[2]  Lars-Åke Johansson,et al.  Flapless, CBCT-guided osteotome sinus floor elevation with simultaneous implant installation. I: radiographic examination and surgical technique. A prospective 1-year follow-up. , 2012, Clinical oral implants research.

[3]  Abel Zhou,et al.  Validation of a Monte Carlo code system for grid evaluation with interference effect on Rayleigh scattering , 2018, Physics in medicine and biology.

[4]  Benjamin M. W. Tsui,et al.  Estimation of geometrical parameters for fan beam tomography , 1987 .

[5]  R Aamir,et al.  MARS spectral molecular imaging of lamb tissue: data collection and image analysis , 2013, 1311.4528.

[6]  A Uneri,et al.  Geometric calibration using line fiducials for cone-beam CT with general, non-circular source-detector trajectories , 2017, Medical Imaging.

[7]  T Varslot,et al.  Reliable automatic alignment of tomographic projection data by passive auto-focus. , 2011, Medical physics.

[8]  Jing Li,et al.  Interior tomography in microscopic CT with image reconstruction constrained by full field of view scan at low spatial resolution. , 2018, Physics in medicine and biology.

[9]  G T Gullberg,et al.  Estimation of geometrical parameters and collimator evaluation for cone beam tomography. , 1990, Medical physics.

[10]  Lothar R. Schad,et al.  Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems , 2018, International Journal of Computer Assisted Radiology and Surgery.

[11]  W A Kalender,et al.  Simultaneous misalignment correction for approximate circular cone-beam computed tomography. , 2008, Physics in medicine and biology.

[12]  A Del Guerra,et al.  An optimization-based method for geometrical calibration in cone-beam CT without dedicated phantoms. , 2008, Physics in medicine and biology.

[13]  Normand Robert,et al.  The geometric calibration of cone-beam systems with arbitrary geometry. , 2009, Physics in medicine and biology.

[14]  Dan Xia,et al.  Geometric Calibration of a Micro-CT System and Performance for Insect Imaging , 2011, IEEE Transactions on Information Technology in Biomedicine.

[15]  Thilo Michel,et al.  Contrast agent recognition in small animal CT using the Medipix2 detector , 2009 .

[16]  Yoshiyuki Noto,et al.  Reduction of dark-band-like metal artifacts caused by dental implant bodies using hypothetical monoenergetic imaging after dual-energy computed tomography. , 2013, Oral surgery, oral medicine, oral pathology and oral radiology.

[17]  N J Pelc,et al.  Alignment of a volumetric tomography system. , 2001, Medical physics.

[18]  A Uneri,et al.  A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories , 2018, Physics in medicine and biology.

[19]  G. Wang,et al.  A general cone-beam reconstruction algorithm , 1993, IEEE Trans. Medical Imaging.

[20]  M. F. Cardoso,et al.  The simplex-simulated annealing approach to continuous non-linear optimization , 1996 .

[21]  David A Jaffray,et al.  Accurate technique for complete geometric calibration of cone-beam computed tomography systems. , 2005, Medical physics.

[22]  Yiannis Kyriakou,et al.  Image features for misalignment correction in medical flat-detector CT. , 2012, Medical physics.

[23]  John A. Nelder,et al.  A Simplex Method for Function Minimization , 1965, Comput. J..

[24]  K. Hoffmann,et al.  Self-calibration of a cone-beam micro-CT system. , 2008, Medical physics.

[25]  R Clackdoyle,et al.  Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography. , 2000, Physics in medicine and biology.

[26]  Dilworth Y Parkinson,et al.  Using high resolution computed tomography to visualize the three dimensional structure and function of plant vasculature. , 2013, Journal of visualized experiments : JoVE.

[27]  Kai Yang,et al.  A geometric calibration method for cone beam CT systems. , 2006, Medical physics.

[28]  Xiaoquan Yang,et al.  Online Geometric Calibration of Cone-Beam Computed Tomography for Arbitrary Imaging Objects , 2013, IEEE Transactions on Medical Imaging.

[29]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[30]  Willi A Kalender,et al.  Geometric misalignment and calibration in cone-beam tomography. , 2004, Medical physics.