Development of Positron Emission Tomography With Wobbling and Zooming for High Sensitivity and High-Resolution Molecular Imaging

Demands for in-vivo human molecular imaging with high resolution and high sensitivity in positron emission tomography (PET) require several new design formulae. A classical problem of the PET design, however, was the trade-off between sensitivity and resolution. To satisfy both requirements, the brain-body convertible PET with wobbling and zooming is proposed. The features of this new proposed system are wobble sampling for high-resolution imaging and zooming mode for high sensitivity, especially for the brain dedicated imaging. For the high resolution, wobbling with a linear interpolation and line spread function (LSF) deconvolution reconstruction algorithm was introduced. The result of the proposed system provided resolution up to 1.56 mm full width at half maximum (FWHM) in the brain mode and resulting in the detector-to-resolution ratio (DRR) was 2.47. For both brain phantom and in-vivo rat brain imaging, the proposed system demonstrated superior image quality compared with the commercial PET systems. The newly designed PET with wobbling and zooming also demonstrated the possibility of developing practically usable high-resolution human brain PET-MRI fusion system, especially for the neuroscience research.

[1]  C. J. Thompson,et al.  Real-Time Data Rebinning in PET to Obtain Uniformly Sampled Projections , 1985, IEEE Transactions on Nuclear Science.

[2]  Shuping Xie,et al.  A dedicated PET system for human brain and head/neck imaging , 2013, 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC).

[3]  Quanzheng Li,et al.  National Electrical Manufacturers Association and Clinical Evaluation of a Novel Brain PET/CT Scanner , 2016, The Journal of Nuclear Medicine.

[4]  Alan C. Evans,et al.  Performance evaluation of the PC-2048: a new 15-slice encoded-crystal PET scanner for neurological studies. , 1991, IEEE transactions on medical imaging.

[5]  E. Hoffman,et al.  A positron-emission transaxial tomograph for nuclear imaging (PETT). , 1975, Radiology.

[6]  C. Knoess,et al.  The ECAT HRRT: NEMA NEC evaluation of the HRRT system, the new High Resolution Research Tomograph , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[7]  Z H Cho,et al.  Positron ranges obtained from biomedically important positron-emitting radionuclides. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  Z. Cho,et al.  Bismuth germanate as a potential scintillation detector in positron cameras. , 1977, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  Thomas Beyer,et al.  Performance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard , 2018, The Journal of Nuclear Medicine.

[10]  E. Hoffman,et al.  Application of annihilation coincidence detection to transaxial reconstruction tomography. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[11]  Simon R. Cherry,et al.  Techniques to improve the spatial sampling of MicroPET-a high resolution animal PET tomograph , 2000 .

[12]  Geor Bakker,et al.  123I-Iododexetimide Preferentially Binds to the Muscarinic Receptor Subtype M1 In Vivo , 2015, The Journal of Nuclear Medicine.

[13]  R. Muenchausen,et al.  Crystal growth and optical characterization of cerium-doped Lu1.8Y0.2SiO5 , 2000 .

[14]  Zang-Hee Cho,et al.  A new approach to very high resolution mini-brain PET using a small number of large detectors , 1990 .

[15]  J. K. Chan,et al.  Circular Ring Transverse Axial Positron Camera for 3-Dimensional Reconstruction of Radionuclides Distribution , 1976, IEEE Transactions on Nuclear Science.

[16]  Jong Beom Ra,et al.  High-resolution circular ring positron tomograph with dichotomic sampling: Dichotom-I , 1983 .

[17]  Z. Cho,et al.  General views on 3-D image reconstruction and computerized transverse axial tomography , 1974 .

[18]  S. Y. Lee,et al.  A New Sampling Scheme for the Ring Positron Camera: Dichotomic Ring Sampling , 1981, IEEE Transactions on Nuclear Science.

[19]  Nan Zhang,et al.  Basic imaging performance characteristics of a variable field of view PET camera using quadrant sharing detector design , 1998 .

[20]  Michel M. Ter-Pogossian,et al.  Performance Study of PETT VI, a Positron Computed Tomograph with 288 Cesium Fluoride Detectors , 1982, IEEE Transactions on Nuclear Science.

[21]  Jong Beom Ra,et al.  A New Sampling Scheme for the Ring Positron Camera: Dichotomic Ring Sampling , 1981 .

[22]  C J Thompson,et al.  Dependence of the coincidence aperture function of narrow BGO crystals on crystal shape and light encoding schemes. , 1986, Physics in medicine and biology.

[23]  M Honda,et al.  Development of a Brain PET System, PET-Hat: A Wearable PET System for Brain Research , 2011, IEEE Transactions on Nuclear Science.

[24]  H. Zaidi,et al.  Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system , 2011, Physics in medicine and biology.

[25]  Zang-Hee Cho,et al.  A fusion PET–MRI system with a high‐resolution research tomograph‐PET and ultra‐high field 7.0 T‐MRI for the molecular‐genetic imaging of the brain , 2008, Proteomics.

[26]  Thomas Beyer,et al.  Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard , 2015, EJNMMI Physics.

[27]  J. Uribe,et al.  Studies of a Next-Generation Silicon-Photomultiplier–Based Time-of-Flight PET/CT System , 2017, The Journal of Nuclear Medicine.

[28]  Julien Bert,et al.  The MINDVIEW project: First results , 2018, European Psychiatry.

[29]  Yong Choi,et al.  Development of a position decoder circuit for PET consisting of GAPD arrays , 2010 .

[30]  Zang-Hee Cho,et al.  Wobbling and LSF-based maximum likelihood expectation maximization reconstruction for wobbling PET , 2016 .

[31]  D. Townsend,et al.  Method for transforming CT images for attenuation correction in PET/CT imaging. , 2006, Medical physics.

[32]  Klaus Wienhard,et al.  The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph , 2000 .

[33]  David W. Townsend,et al.  An investigation of factors affecting detector and geometric correction in normalization of 3-D PET data , 1995 .

[34]  M. Singh,et al.  Performance and Evaluation of the Circular Ring Transverse Axial Positron Camera (CRTAPC) , 1977, IEEE Transactions on Nuclear Science.