Scatter correction for cone-beam CT via scatter kernel superposition-inspired convolutional neural network

Objective. X-ray scatter leads to signal bias and degrades the image quality in Computed Tomography imaging. Conventional real-time scatter estimation and correction methods include the scatter kernel superposition (SKS) methods, which approximate x-ray scatter field as a convolution of the scatter sources and scatter propagation kernels to reflect the spatial spreading of scatter x-ray photons. SKS methods are fast to implement but generally suffer from low accuracy due to the difficulties in determining the scatter kernels. Approach. To address such a problem, this work describes a new scatter estimation and correction method by combining the concept of SKS methods and convolutional neural network. Unlike conventional SKS methods which estimate the scatter amplitude and the scatter kernel based on the value of an individual pixel, the proposed method generates the scatter amplitude maps and the scatter width maps from projection images through a neural network, from which the final estimated scatter field is calculated based on a convolution process. Main Results. By incorporating physics in the network design, the proposed method requires fewer trainable parameters compared with another deep learning-based method (Deep Scatter Estimation). Both numerical simulations and physical experiments demonstrate that the proposed SKS-inspired convolutional neural network outperforms the conventional SKS method and other deep learning-based methods in both qualitative and quantitative aspects. Significance. The proposed method can effectively correct the scatter-related artifacts with a SKS-inspired convolutional neural network design.

[1]  Karl Stierstorfer,et al.  Deep Learning-Based Forward and Cross-Scatter Correction in Dual Source CT. , 2021, Medical physics.

[2]  Christopher Syben,et al.  X-Ray Scatter Estimation Using Deep Splines , 2021, IEEE Transactions on Medical Imaging.

[3]  Jing Xiao,et al.  Dual Encoder Fusion U-Net (DEFU-Net) for Cross-manufacturer Chest X-ray Segmentation , 2020, 2020 25th International Conference on Pattern Recognition (ICPR).

[4]  Gregory C Sharp,et al.  Evaluation of CBCT scatter correction using deep convolutional neural networks for head and neck adaptive proton therapy , 2020, Physics in medicine and biology.

[5]  J. Fessler,et al.  A deep neural network for fast and accurate scatter estimation in quantitative SPECT/CT under challenging scatter conditions , 2020, European Journal of Nuclear Medicine and Molecular Imaging.

[6]  Christopher Kurz,et al.  CBCT correction using a cycle-consistent generative adversarial network and unpaired training to enable photon and proton dose calculation , 2019, Physics in medicine and biology.

[7]  Guang-Hong Chen,et al.  Task-driven optimization of the non-spectral mode of photon counting CT for intracranial hemorrhage assessment , 2019, Physics in medicine and biology.

[8]  Ke Sheng,et al.  Scatter correction of cone-beam CT using a deep residual convolution neural network (DRCNN) , 2019, Physics in medicine and biology.

[9]  Marc Kachelrieß,et al.  Real‐time scatter estimation for medical CT using the deep scatter estimation: Method and robustness analysis with respect to different anatomies, dose levels, tube voltages, and data truncation , 2018, Medical physics.

[10]  Jing Wang,et al.  Generating synthesized computed tomography (CT) from cone-beam computed tomography (CBCT) using CycleGAN for adaptive radiation therapy , 2018, Physics in medicine and biology.

[11]  Christopher Kurz,et al.  ScatterNet: A convolutional neural network for cone‐beam CT intensity correction , 2018, Medical physics.

[12]  Marc Kachelrieß,et al.  Deep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT Using a Deep Convolutional Neural Network , 2018, Journal of Nondestructive Evaluation.

[13]  Jan Kautz,et al.  Loss Functions for Image Restoration With Neural Networks , 2017, IEEE Transactions on Computational Imaging.

[14]  A. Sisniega,et al.  Technical Note: spektr 3.0—A computational tool for x-ray spectrum modeling and analysis , 2016, Medical physics.

[15]  K. Klingenbeck,et al.  A general framework and review of scatter correction methods in cone beam CT. Part 2: scatter estimation approaches. , 2011, Medical physics.

[16]  Klaus Klingenbeck,et al.  A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: Scatter compensation approaches. , 2011, Medical physics.

[17]  J. Star-Lack,et al.  Improved scatter correction using adaptive scatter kernel superposition , 2010, Physics in medicine and biology.

[18]  Willi A Kalender,et al.  A fast and pragmatic approach for scatter correction in flat-detector CT using elliptic modeling and iterative optimization , 2010, Physics in medicine and biology.

[19]  Radhe Mohan,et al.  Scatter kernel estimation with an edge-spread function method for cone-beam computed tomography imaging , 2008, Physics in medicine and biology.

[20]  W. Kalender,et al.  Efficiency of antiscatter grids for flat-detector CT , 2007, Physics in medicine and biology.

[21]  D. Jaffray,et al.  The influence of bowtie filtration on cone-beam CT image quality. , 2007, Medical physics.

[22]  Freek J. Beekman,et al.  Efficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT , 2006, IEEE Transactions on Medical Imaging.

[23]  D. Jaffray,et al.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. , 2001, Medical physics.

[24]  R. Kruger,et al.  Scatter estimation for a digital radiographic system using convolution filtering. , 1987, Medical physics.

[25]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[26]  G H Glover,et al.  Compton scatter effects in CT reconstructions. , 1983, Medical physics.

[27]  P M Joseph,et al.  The effects of scatter in x-ray computed tomography. , 1982, Medical physics.

[28]  W. Curran,et al.  Paired cycle-GAN-based image correction for quantitative cone-beam computed tomography , 2020 .

[29]  K. Klingenbeck-Regn,et al.  Efficient object scatter correction algorithm for third and fourth generation CT scanners , 1999, European Radiology.