Detection of nitrite by flow injection analysis using a novel paired emitter-detector diode (PEDD) as a photometric detector

An inexpensive flow injection analysis system for determining low concentration levels of nitrite employing the Griess reagent spectrophotometric method is reported. The novel photometric detector applied within this manifold is a highly sensitive, low cost, miniaturized light emitting diode (LED) based flow detector. This colorimetric detector employs two LEDs, operating one as a light source and the other as a light detector. The emitter LED is forward biased and the detector reverse biased. The emitter and detector LED had a λmax of 530 nm and 623 nm respectively. The emission spectrum of the emitter LED efficiently overlapped with the absorbance spectrum of 9 µM NO2 and Griess reagent complex. A simple timer circuit measures the time taken for the photocurrent generated by the emitter LED to discharge the detector LED from 5 V (logic 1) to 1.7 V (logic 0). The Griess reagent method employed for nitrite determination is based on the formation of an azo dye, the intensity of which is directly related to nitrite concentration. The linear range, reproducibility and limit of detection were investigated. Detection limits in the nanomolar range were achieved using the Paired Emitter-Detector Diode (PEDD) flow analysis device. For a comparative study the linear range and limit of detection were also investigated using a platewell reader. Higher sensitivity and improved precision were obtained from the PEDD compared to the commercially available plate well reader.

[1]  Martha B. Shinn,et al.  Colorimetric Method for Determination of Nitrate , 1941 .

[2]  Gillian M. Greenway,et al.  Microfluidic devices for environmental monitoring , 2005 .

[3]  Wolfgang Frenzel,et al.  Characterisation of a gas-diffusion membrane-based optical flow-through sensor exemplified by the determination of nitrite. , 2004, Talanta.

[4]  Eric Kaltenbacher,et al.  In situ nitrite measurements using a compact spectrophotometric analysis system , 2002 .

[5]  Y. Miura,et al.  Flow Injection Analysis of Nitrite Based on Spectrophotometric Measurements of Iodine Formed by Oxidation of Iodide with Nitrite , 1999 .

[6]  Pieter Telleman,et al.  Microsystem Engineering of Lab-on-a-chip devices. , 2003 .

[7]  R. Uglow,et al.  A simple simultaneous flow injection method based on phosphomolybdenum chemistry for nitrate and nitrite determinations in water and fish samples. , 2002, Talanta.

[8]  Robert D. Waterbury,et al.  Determination of nanomolar concentrations of nitrite and nitrate in natural waters using long path length absorbance spectroscopy , 1998 .

[9]  B. Reis,et al.  A downsized flow set up based on multicommutation for the sequential photometric determination of iron(II)/iron(III) and nitrite/nitrate in surface water. , 2005, Talanta.

[10]  Suchandra Biswas,et al.  A novel spectrofluorimetric method for the ultra trace analysis of nitrite and nitrate in aqueous medium and its application to air, water, soil and forensic samples. , 2004, Talanta.

[11]  R. Compton,et al.  Detection and determination of nitrate and nitrite: a review. , 2001, Talanta.

[12]  P. Hauser,et al.  A multi-wavelength photometer based on light-emitting diodes. , 1995, Talanta.

[13]  A. Aydin,et al.  A novel method for the spectrophotometric determination of nitrite in water. , 2005, Talanta.

[14]  M. J. Ahmed,et al.  Simultaneous spectrophotometric determination of nitrite and nitrate by flow-injection analysis. , 1996, Talanta.

[15]  X. Yue,et al.  Flow injection catalytic spectrophotometric simultaneous determination of nitrite and nitrate. , 2004, Talanta.

[16]  Boaventura F. Reis,et al.  A flow system exploiting multicommutation for speciation of inorganic nitrogen in waters. , 2000 .

[17]  V. Hessel,et al.  Buchbesprechung: Microsystem Engineering of Lab-on-a-Chip Devices. Von O. Geschke, H. Klank, P. Tellemann. , 2004 .

[18]  K. Lau,et al.  A low-cost optical sensing device based on paired emitter–detector light emitting diodes , 2006 .

[19]  Gillian M. Greenway,et al.  Characterisation of a micro-total analytical system for the determination of nitrite with spectrophotometric detection , 1999 .

[20]  Masao Sakuraba,et al.  Integration of Si p–i–n diodes for light emitter and detector with optical waveguides , 2005 .

[21]  T. W. Stanley,et al.  Comparison of fifty-two spectrophotometric methods for the determination of nitrite , 1963 .

[22]  V. M. Ivanov,et al.  The 125th Anniversary of the Griess Reagent , 2004 .

[23]  J. Martínez Calatayud,et al.  Flow-injection analysis-spectrophotometric determination of nitrite and nitrate in water samples by reaction with proflavin. , 1996 .

[24]  Dermot Diamond,et al.  Photometric detection in flow analysis systems using integrated PEDDs. , 2005, Talanta.

[25]  Shoji Motomizu,et al.  Simple flow-injection system for the simultaneous determination of nitrite and nitrate in water samples. , 2004, Talanta.

[26]  Leif G. Anderson,et al.  Simultaneous spectrophotometric determination of nitrite and nitrate by flow injection analysis , 1979 .

[27]  T. Attin,et al.  Suitability of a malachite green procedure to detect minimal amounts of phosphate dissolved in acidic solutions , 2005, Clinical Oral Investigations.