Thimble ionization chambers in medium-energy x-ray beams and the role of constructive details of the central electrode: Monte Carlo simulations and measurements

This paper presents investigations of thimble ionization chamber response in medium-energy kilovoltage x-ray beams (70-280 kVp, 0.09-3.40 mm Cu HVL). Two thimble ionization chambers (PTW30015 and PTW30016) were investigated, regarding the influence of the central electrode dimensions made of aluminum. Measurements were carried out in photon fields of different beam quality. Corresponding Monte Carlo simulations employing the EGSnrc Monte Carlo code system were performed. The simulations included the modelling of the x-ray tube and measurement setup for generation of x-ray spectra. These spectra were subsequently used to calculate the absorbed energy in the air cavity of the two thimble ionization chamber models and the air kerma at the reference point of the chambers. Measurements and simulations revealed an optimal diameter of the central electrode, concerning an almost energy-independent response of the ionizaton chamber. The Monte Carlo simulations are in good agreement with the measured values, expressed in beam quality correction factors k(Q). The agreement was generally within 0.6% but could only be achieved with an accurate model of the central electrode including its exact shape. Otherwise, deviations up to 8.5% resulted, decreasing with higher photon energies, which can be addressed to the high yield of the photoelectric effect in the electrode material aluminum at low photon energies.

[1]  I. Kawrakow,et al.  Calculated And Measured Air-Kerma Response Of Ionization Chambers In Low- And Medium-Energy Photon Beams , 2002 .

[2]  C. Ma,et al.  BEAM: a Monte Carlo code to simulate radiotherapy treatment units. , 1995, Medical physics.

[3]  P. Andreo,et al.  Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water , 2001 .

[4]  Martin J. Berger,et al.  Bremsstrahlung energy spectra from electrons with kinetic energy 1 keV–10 GeV incident on screened nuclei and orbital electrons of neutral atoms with Z = 1–100 , 1986 .

[5]  Martin J. Berger,et al.  Bremsstrahlung spectra from electron interactions with screened atomic nuclei and orbital electrons , 1985 .

[6]  I. Kawrakow Accurate condensed history Monte Carlo simulation of electron transport. II. Application to ion chamber response simulations. , 2000, Medical physics.

[7]  I. Kawrakow,et al.  Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMC , 2000, Physics in medicine and biology.

[8]  J. H. Hubbell,et al.  Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest , 1995 .

[9]  H Zaidi,et al.  Monte carlo simulation of x-ray spectra in diagnostic radiology and mammography using MCNP4C. , 2004, Physics in medicine and biology.

[10]  M. Mcewen,et al.  An experimental and computational investigation of the standard temperature-pressure correction factor for ion chambers in kilovoltage x rays. , 2007, Medical physics.

[11]  A. Nahum,et al.  Calculated response and wall correction factors for ionization chambers exposed to 60Co gamma rays. , 1982, Medical physics.

[12]  S M Seltzer,et al.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology. , 2001, Medical physics.

[13]  A. Maio,et al.  Monte Carlo simulation of a typical 60Co therapy source. , 1999, Medical physics.

[14]  I. Kawrakow,et al.  Efficient photon beam dose calculations using DOSXYZnrc with BEAMnrc. , 2006, Medical physics.

[15]  Iwan Kawrakow,et al.  Efficient x-ray tube simulations. , 2006, Medical physics.

[16]  A. Nahum,et al.  Effect of size and composition of the central electrode on the response of cylindrical ionization chambers in high-energy photon and electron beams , 1993 .

[17]  Fritz Sauter,et al.  Über den atomaren Photoeffekt in der K-Schale nach der relativistischen Wellenmechanik Diracs , 1931 .

[18]  F. H. Attix Introduction to Radiological Physics and Radiation Dosimetry , 1991 .

[19]  I. Kawrakow Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version. , 2000, Medical physics.

[20]  J F Williamson,et al.  Monte Carlo evaluation of kerma at a point for photon transport problems. , 1987, Medical physics.

[21]  J. H. Hubbell,et al.  XCOM: Photon cross sections on a personal computer , 1987 .

[22]  F Verhaegen,et al.  Monte Carlo modelling of radiotherapy kV x-ray units. , 1999, Physics in medicine and biology.

[23]  Frank Verhaegen,et al.  Evaluation of the EGSnrc Monte Carlo code for interface dosimetry near high-Z media exposed to kilovolt and 60Co photons. , 2002, Physics in medicine and biology.

[24]  F Verhaegen,et al.  An investigation of entrance surface dose calculations for diagnostic radiology using Monte Carlo simulations and radiotherapy dosimetry formalisms. , 2003, Physics in medicine and biology.

[25]  David W. O. Rogers,et al.  Ion chamber response and Awall correction factors in a 60Co beam by Monte Carlo simulation , 1985 .

[26]  M. S. Yeom,et al.  Monte Carlo calculation of the ionization chamber response to 60Co beam using PENELOPE. , 2006, Medical Physics (Lancaster).

[27]  A. Nahum,et al.  The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL; 10-300 kV generating potential). Institution of Physics and Engineering in Medicine and Biology. , 1996, Physics in medicine and biology.

[28]  Iwan Kawrakow,et al.  Calculations for plane-parallel ion chambers in 60Co beams using the EGSnrc Monte Carlo code. , 2003, Medical physics.

[29]  Ravinder Nath,et al.  Calculated response and wall correction factors for ionization chambers exposed to 60Co gamma rays , 1981 .

[30]  D. Rogers,et al.  EGS4 code system , 1985 .

[31]  I. Kawrakow,et al.  Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting. , 2004, Medical physics.