A model‐based tracking method for measuring 3D dynamic joint motion using an alternating biplane x‐ray imaging system

PURPOSES To propose a new model-based tracking method for measuring three-dimensional (3D) dynamic joint kinematics using a clinical alternating biplane x-ray imaging system; and to quantify in vitro its errors in measuring ankle and knee motions at different motion speeds. METHODS A new model-based tracking method based on motion component partition and interpolation (MCPI) was developed for measuring 3D dynamic joint kinematics based on a clinical alternating biplane x-ray imaging system. Two detectors of the biplane imaging system placed perpendicular to each other were operated to collect alternating fluoroscopic images of the targeted joint during tasks. The CT data of the joint were also acquired for the reconstruction of volumetric and surface models of each of the associated bones. The CT-based models of the bones were first registered to the alternating images using a model-to-single-plane fluoroscopic image registration method, and the resulting bone poses were then refined using a two-level optimization with motion component partition and model vertex trajectory interpolation. The MCPI method was evaluated in vitro for measurement errors for an ankle and a knee specimen moving at different speeds against a standard reference provided by a highly accurate motion capture system. The positional and rotational errors of the measured bone poses were quantified in terms of the bias, precision, and root-mean-squared errors (RMSE), as well as the mean target registration error (mTRE), a final mTRE less than 2.5 mm indicating a successful registration. RESULTS The new method was found to have RMSE of bone pose measurements of less than 0.18 mm for translations and 0.72° for rotations for the ankle, and 0.33 mm and 0.74° for the knee with a high successful registration rate (>97%), and did not appear to be affected by joint motion speeds. Given the same alternating fluoroscopic images, the MCPI method outperformed the typical biplane analysis method assuming zero time offset between the two fluoroscopic views. The differences in performance between the methods were increased with increased joint motion speed. With the accurate bone pose data, the new method enabled talocrural, subtalar, and tibiofemoral kinematics measurements with submillimeter and subdegree accuracy, except for an RMSE of 1.04° for the internal/external rotation of the talocrural joint. CONCLUSIONS A new model-based tracking method based on MCPI has been developed for measuring dynamic joint motions using an alternating biplane x-ray imaging system widely available in medical centers. The MCPI method has been demonstrated in vitro to be highly accurate in determining the 3D kinematics of the bones of both the ankle joint complex and the knee. The current results suggest that the MCPI method would be an effective approach for measuring in vivo 3D kinematics of dynamic joint motion in a clinical setting equipped with an alternating biplane x-ray imaging system.

[1]  E R Valstar,et al.  Image-based RSA: Roentgen stereophotogrammetric analysis based on 2D-3D image registration. , 2008, Journal of biomechanics.

[2]  J. O'Connor,et al.  Ligaments and articular contact guide passive knee flexion. , 1998, Journal of biomechanics.

[3]  Kevin M. Johnson,et al.  Measurement of tibiofemoral kinematics using highly accelerated 3D radial sampling , 2013, Magnetic resonance in medicine.

[4]  Tung-Wu Lu,et al.  In vivo three-dimensional kinematics of the normal knee during active extension under unloaded and loaded conditions using single-plane fluoroscopy. , 2008, Medical engineering & physics.

[5]  Satoshi Hamai,et al.  Knee kinematics in medial osteoarthritis during in vivo weight‐bearing activities , 2009, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[6]  Michael J Bey,et al.  Three-dimensional motion analysis of the cervical spine for comparison of anterior cervical decompression and fusion versus artificial disc replacement in 17 patients: clinical article. , 2014, Journal of neurosurgery. Spine.

[7]  Atsushi Akaike,et al.  In Vivo Talocrural Joint Contact Mechanics With Functional Ankle Instability , 2015, Foot & ankle specialist.

[8]  Guoan Li,et al.  Validation of a non-invasive fluoroscopic imaging technique for the measurement of dynamic knee joint motion. , 2008, Journal of biomechanics.

[9]  Yoshinobu Sato,et al.  Kinematics of the Upper Cervical Spine in Rotation: In Vivo Three-Dimensional Analysis , 2004, Spine.

[10]  Edmund Y S Chao,et al.  Prediction of three-dimensional contact stress and ligament tension in the ankle during stance determined from computational modeling. , 2009, Foot & ankle international.

[11]  Tung-Wu Lu,et al.  Comparisons of surface vs. volumetric model-based registration methods using single-plane vs. bi-plane fluoroscopy in measuring spinal kinematics. , 2014, Medical engineering & physics.

[12]  Tung-Wu Lu,et al.  Intervertebral anticollision constraints improve out-of-plane translation accuracy of a single-plane fluoroscopy-to-CT registration method for measuring spinal motion. , 2013, Medical physics.

[13]  S. Woo,et al.  In Vivo Tibiofemoral Kinematics During 4 Functional Tasks of Increasing Demand Using Biplane Fluoroscopy , 2012, The American journal of sports medicine.

[14]  Jeffrey C. Lagarias,et al.  Convergence Properties of the Nelder-Mead Simplex Method in Low Dimensions , 1998, SIAM J. Optim..

[15]  Michael J Bey,et al.  Accuracy and feasibility of dual fluoroscopy and model-based tracking to quantify in vivo hip kinematics during clinical exams. , 2014, Journal of applied biomechanics.

[16]  L. Blankevoort,et al.  The relation between geometry and function of the ankle joint complex: a biomechanical review , 2010, Knee Surgery, Sports Traumatology, Arthroscopy.

[17]  Seungbum Koo,et al.  Plantar-flexion of the ankle joint complex in terminal stance is initiated by subtalar plantar-flexion: A bi-planar fluoroscopy study. , 2015, Gait & posture.

[18]  J Erik Giphart,et al.  Accuracy of a contour-based biplane fluoroscopy technique for tracking knee joint kinematics of different speeds. , 2012, Journal of biomechanics.

[19]  E S Grood,et al.  A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. , 1983, Journal of biomechanical engineering.

[20]  William E. Lorensen,et al.  Marching cubes: A high resolution 3D surface construction algorithm , 1987, SIGGRAPH.

[21]  D. B. Baier,et al.  X-ray reconstruction of moving morphology (XROMM): precision, accuracy and applications in comparative biomechanics research. , 2010, Journal of experimental zoology. Part A, Ecological genetics and physiology.

[22]  Knee Surgery,et al.  Knee surgery sports traumatology arthroscopy , 1993 .

[23]  K. Shelburne,et al.  A comparison of calibration methods for stereo fluoroscopic imaging systems. , 2011, Journal of biomechanics.

[24]  Hideki Yoshikawa,et al.  Kinematics of the Subaxial Cervical Spine in Rotation In Vivo Three-Dimensional Analysis , 2004, Spine.

[25]  Andrew E Anderson,et al.  In Vivo Kinematics of the Tibiotalar and Subtalar Joints in Asymptomatic Subjects: A High-Speed Dual Fluoroscopy Study. , 2016, Journal of biomechanical engineering.

[26]  Scott Tashman,et al.  Validation of three-dimensional model-based tibio-femoral tracking during running. , 2009, Medical engineering & physics.

[27]  Cyrille Lenders,et al.  In vivo three-dimensional kinematics of the cervical spine during maximal axial rotation. , 2013, Manual therapy.

[28]  D L Hill,et al.  Validation of a two- to three-dimensional registration algorithm for aligning preoperative CT images and intraoperative fluoroscopy images. , 2001, Medical physics.

[29]  Peter A Hardy,et al.  Three-dimensional in vivo motion of adult hind foot bones. , 2006, Journal of biomechanics.

[30]  A. Leardini,et al.  Rear-foot, mid-foot and fore-foot motion during the stance phase of gait. , 2007, Gait & posture.

[31]  A. Cappozzo,et al.  Human movement analysis using stereophotogrammetry. Part 3. Soft tissue artifact assessment and compensation. , 2005, Gait & posture.

[32]  Benjamin J Fregly,et al.  Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy. , 2005, Journal of biomechanical engineering.

[33]  Andrew E Anderson,et al.  Accuracy and feasibility of high-speed dual fluoroscopy and model-based tracking to measure in vivo ankle arthrokinematics. , 2015, Gait & posture.

[34]  Yoshinori Takakura,et al.  In vivo kinematics of two-component total ankle arthroplasty during non-weightbearing and weightbearing dorsiflexion/plantarflexion. , 2011, Journal of biomechanics.

[35]  Joseph J Crisco,et al.  Automatic determination of anatomical coordinate systems for three-dimensional bone models of the isolated human knee. , 2010, Journal of biomechanics.

[36]  Sonia Duprey,et al.  Influence of joint constraints on lower limb kinematics estimation from skin markers using global optimization. , 2010, Journal of biomechanics.

[37]  J J O'Connor,et al.  Bone position estimation from skin marker co-ordinates using global optimisation with joint constraints. , 1999, Journal of biomechanics.

[38]  Makoto Sakamoto,et al.  Accuracy of Single Plane X-Ray Image-Based Technique for Assessment of Knee Kinematics , 2009 .

[39]  Guoan Li,et al.  A novel dual fluoroscopic imaging method for determination of THA kinematics: in-vitro and in-vivo study. , 2013, Journal of biomechanics.

[40]  N. Stergiou,et al.  Three-Dimensional Tibiofemoral Kinematics of the Anterior Cruciate Ligament-Deficient and Reconstructed Knee during Walking * , 2003, The American journal of sports medicine.

[41]  T. Gill,et al.  In Vivo Elongation of the Anterior Cruciate Ligament and Posterior Cruciate Ligament during Knee Flexion , 2004, The American journal of sports medicine.

[42]  Thomas S Buchanan,et al.  A method for measurement of joint kinematics in vivo by registration of 3-D geometric models with cine phase contrast magnetic resonance imaging data. , 2005, Journal of biomechanical engineering.

[43]  Joon Ho Wang,et al.  In Vivo Analysis of the Isolated Posterior Cruciate Ligament–Deficient Knee During Functional Activities , 2012, American Journal of Sports Medicine.

[44]  Juan M Santos,et al.  Feasibility of using real‐time MRI to measure joint kinematics in 1.5T and open‐bore 0.5T systems , 2008, Journal of magnetic resonance imaging : JMRI.

[45]  Meir Nyska,et al.  Deviations in gait metrics in patients with chronic ankle instability: a case control study , 2015, Journal of Foot and Ankle Research.

[46]  F. Veldpaus,et al.  A least-squares algorithm for the equiform transformation from spatial marker co-ordinates. , 1988, Journal of biomechanics.

[47]  A. Cappozzo,et al.  Human movement analysis using stereophotogrammetry. Part 1: theoretical background. , 2005, Gait & posture.

[48]  Cheng-Chung Lin,et al.  A slice-to-volume registration method based on real-time magnetic resonance imaging for measuring three-dimensional kinematics of the knee. , 2013, Medical physics.

[49]  Tae-Hong Lim,et al.  Biomechanical Study on the Effect of Cervical Spine Fusion on Adjacent-Level Intradiscal Pressure and Segmental Motion , 2002, Spine.

[50]  Hartmut Witte,et al.  ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion--part I: ankle, hip, and spine. International Society of Biomechanics. , 2002, Journal of biomechanics.

[51]  William Hoff,et al.  In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics. , 2005, Journal of biomechanics.

[52]  Mohamed R. Mahfouz,et al.  In Vivo Evaluation of Dynamic Characteristics of the Normal, Fused, and Disc Replacement Cervical Spines , 2007, Spine.

[53]  Louis E DeFrate,et al.  In Vivo Kinematics of the Tibiotalar Joint After Lateral Ankle Instability , 2009, The American journal of sports medicine.

[54]  Tung-Wu Lu,et al.  In vivo three-dimensional intervertebral kinematics of the subaxial cervical spine during seated axial rotation and lateral bending via a fluoroscopy-to-CT registration approach. , 2014, Journal of biomechanics.

[55]  Chung-Ming Chen,et al.  A volumetric model-based 2D to 3D registration method for measuring kinematics of natural knees with single-plane fluoroscopy. , 2010, Medical physics.

[56]  Mohamed R Mahfouz,et al.  Multicenter Determination of In Vivo Kinematics After Total Knee Arthroplasty , 2003, Clinical orthopaedics and related research.

[57]  Robert F. LaPrade,et al.  Normative rearfoot motion during barefoot and shod walking using biplane fluoroscopy , 2016, Knee Surgery, Sports Traumatology, Arthroscopy.

[58]  Masahiro Shimizu,et al.  Direct assessment of 3D foot bone kinematics using biplanar X-ray fluoroscopy and an automatic model registration method , 2015, Journal of Foot and Ankle Research.

[59]  Tung-Wu Lu,et al.  Effects of soft tissue artifacts on the calculated kinematics and kinetics of the knee during stair-ascent. , 2011, Journal of biomechanics.

[60]  Won Man Park,et al.  A combined numerical and experimental technique for estimation of the forces and moments in the lumbar intervertebral disc , 2013, Computer methods in biomechanics and biomedical engineering.

[61]  Jens Frahm,et al.  Real-time magnetic resonance imaging of temporomandibular joint dynamics. , 2011 .

[62]  Graeme P. Penney,et al.  Standardized evaluation methodology for 2-D-3-D registration , 2005, IEEE Transactions on Medical Imaging.

[63]  Remmet Jonges,et al.  In-vivo range of motion of the subtalar joint using computed tomography. , 2008, Journal of biomechanics.

[64]  S. Banks,et al.  Ankle and Subtalar Kinematics during Dorsiflexion-Plantarflexion Activities , 2009, Foot & ankle international.