Adverse Anesthetic Outcomes Arising from Gas Delivery Equipment: A Closed Claims Analysis

Background: Anesthesia gas delivery equipment is a potentially important source of patient injury. To better define the contribution of gas delivery equipment to professional liability in anesthesia, the authors conducted an in‐depth analysis of cases from the database of the American Society of Anesthesiologists Closed Claims Project. Methods: The database of the Closed Claims Project is composed of closed US malpractice claims that have been collected in a standardized manner. All claims resulting from the use of gas delivery equipment were reviewed for recurrent patterns of injury. Results: Gas delivery equipment was associated with 72 (2%) of 3,791 claims in the database. Death and permanent brain damage accounted for almost all adverse outcomes (n = 55, 76%). Equipment misuse was defined as fault or human error associated with the preparation, maintenance, or deployment of a medical device. Equipment failure was defined as unexpected malfunction of a medical device, despite routine maintenance and previous uneventful use. Misuse of equipment (n = 54, 75%) was three times more common than equipment failure (n = 17, 24%). Misconnects and disconnects of the breathing circuit made the largest contribution to injury (n = 25, 35%). Reviewers judged that 38 of 72 claims (53%) could have been prevented by pulse oximetry, capnography, or a combination of these two monitors. Overall, 56 of 72 gas delivery claims (78%) were deemed preventable with the use or better use of monitors. The year of occurrence for claims involving gas delivery equipment ranged from 1962 to 1991 and did not differ significantly from claims involving other adverse respiratory events. Conclusions: Claims associated with gas delivery equipment are infrequent but severe and continue to occur in the 1990s. Educational and preventive strategies that focus on equipment misuse and breathing circuit configuration may have the greatest potential for enhancing the safety of anesthesia gas delivery equipment.

[1]  R. Holland Anesthesia‐Related Mortality in Australia , 1984, International anesthesiology clinics.

[2]  R L Keenan,et al.  Cardiac arrest due to anesthesia. A study of incidence and causes. , 1985, JAMA.

[3]  J. Cooper,et al.  An analysis of major errors and equipment failures in anesthesia management: considerations for prevention and detection. , 1984 .

[4]  J Spierdijk,et al.  Accidents, near accidents and complications during anaesthesia A retrospective analysis of a 10‐year period in a teaching hospital , 1990, Anaesthesia.

[5]  J. R. Otworth,et al.  Bilateral tension pneumothorax due to a defective anesthesia breathing circuit filter. , 1991, Journal of clinical anesthesia.

[6]  Karen B. Domino,et al.  Nerve injury associated with anesthesia: a closed claims analysis. , 1991, Anesthesiology.

[7]  K L Posner,et al.  Adverse respiratory events in anesthesia: a closed claims analysis. , 1990, Anesthesiology.

[8]  R. Caplan,et al.  Eye injuries associated with anesthesia , 1992 .

[9]  W J Russell,et al.  Problems with Ventilation: An Analysis of 2000 Incident Reports , 1993, Anaesthesia and intensive care.

[10]  C. Kingsley,et al.  Defective carbon dioxide absorber as a cause for a leak in a breathing circuit. , 1996, Anesthesiology.

[11]  D. Benhamou,et al.  Nitroglycerin for fetal head entrapment during vaginal breech delivery? , 1995, Anesthesia and analgesia.

[12]  P Mendel Oxygen supply failure revisited , 1994, Journal of clinical monitoring.

[13]  P. Slinger,et al.  Intraoperative awareness due to malfunction of a Siemens 900B ventilator , 1990, Canadian journal of anaesthesia = Journal canadien d'anesthesie.

[14]  J. Crowley,et al.  An evaluation of anesthesiologists' present checkout methods and the validity of the FDA checklist. , 1991, Anesthesiology.

[15]  R. K. Webb,et al.  The Australian Incident Monitoring Study. Crisis management--validation of an algorithm by analysis of 2000 incident reports. , 1993, Anaesthesia and intensive care.

[16]  Breathing system disconnections. , 1994, British journal of anaesthesia.

[17]  K L Posner,et al.  Adverse respiratory events infrequently leading to malpractice suits. A closed claims analysis. , 1991, Anesthesiology.

[18]  M. Hovi-Viander,et al.  Death associated with anaesthesia in Finland. , 1980, British journal of anaesthesia.

[19]  R. Kaplan,et al.  Unusual source of air leak in a pediatric anesthesia breathing circuit. , 1995, Anesthesia and Analgesia.

[20]  V. Kumar,et al.  An analysis of critical incidents in a teaching department for quality assurance A survey of mishaps during anaesthesia , 1988, Anaesthesia.

[21]  J. Utting,et al.  Human misadventure in anaesthesia , 1979, Canadian Anaesthetists' Society journal.

[22]  Standard of care and anesthesia liability. , 1989 .

[23]  P. Biro Unusual cause for a circle system leak. , 1996, Anesthesia and analgesia.

[24]  J. Spierdijk,et al.  Reported significant observations during anaesthesia: a prospective analysis over an 18-month period. , 1992, British journal of anaesthesia.

[25]  A. McEwan,et al.  Bilateral Tension Pneumothorax Caused by a Blocked Bacterial Filter in an Anesthesia Breathing Circuit , 1993, Anesthesia and analgesia.

[26]  P D Sampson,et al.  Measuring interrater reliability among multiple raters: an example of methods for nominal data. , 1990, Statistics in medicine.

[27]  Janet B. Craig,et al.  A survey of anaesthetic misadventures , 1981, Anaesthesia.

[28]  L. Tiret,et al.  Complications associated with anaesthesiaa — prospective survey in France , 1986, Canadian Anaesthetists' Society journal.

[29]  J. Charlton Checklists and patient safety , 1990, Anaesthesia.