Performance comparison of breast imaging modalities using a 4AFC human observer study

This work compares the visibility of spheres and simulated masses in 2D-mammography and tomosynthesis systems using human observer studies. Performing comparison studies between breast imaging systems poses a number of practical challenges within a clinical environment. We therefore adopted a simulation approach which included synthetic breast blocks, a validated lesion simulation model and a set of validated image modelling tools as a viable alternative to clinical trials. A series of 4-alternative forced choice (4AFC) human observer experiments has been conducted for signal detection tasks using masses and spheres as targets. Five physicists participated in the study viewing images with a 5mm target at a range of contrast levels and 60 trials per experimental condition. The results showed that tomosynthesis has a lower threshold contrast than 2D-mammography for masses and spheres, and that detection studies using spheres may produce overly-optimistic threshold contrast values.

[1]  Oliver Diaz,et al.  Image simulation and a model of noise power spectra across a range of mammographic beam qualities. , 2014, Medical physics.

[2]  Kenneth C. Young,et al.  Conversion of mammographic images to appear with the noise and sharpness characteristics of a different detector and x-ray system. , 2012, Medical physics.

[3]  Walter Huda,et al.  How do lesion size and random noise affect detection performance in digital mammography? , 2006, Academic radiology.

[4]  D R Dance,et al.  Simulation and assessment of realistic breast lesions using fractal growth models , 2013, Physics in medicine and biology.

[5]  D. M. Parkin,et al.  Use of Statistics to Assess the Global Burden of Breast Cancer , 2006, The breast journal.

[6]  D R Dance,et al.  Estimation of scattered radiation in digital breast tomosynthesis , 2014, Physics in medicine and biology.

[7]  M Ruschin,et al.  Visibility of microcalcification clusters and masses in breast tomosynthesis image volumes and digital mammography: a 4AFC human observer study. , 2012, Medical physics.

[8]  A. Burgess Comparison of receiver operating characteristic and forced choice observer performance measurement methods. , 1995, Medical physics.

[9]  Hilde Bosmans,et al.  Development and validation of a modelling framework for simulating 2D-mammography and breast tomosynthesis images , 2014, Physics in medicine and biology.

[10]  Alaleh Rashidnasab,et al.  Simulation of 3D DLA masses in digital breast tomosynthesis , 2013, Medical Imaging.

[11]  Kenneth C. Young,et al.  Comparison of Breast Doses for Digital Tomosynthesis Estimated from Patient Exposures and Using PMMA Breast Phantoms , 2012, Digital Mammography / IWDM.