Generalized monitor unit calculation for the Varian enhanced dynamic wedge field.

The generalized monitor unit (MU) calculation equation for the Varian enhanced dynamic wedge (EDW) is derived. The assumption of this MU calculation method is that the wedge factor of the EDW at the center of the field is a function of field size, the position of the center of the field in the wedge direction, and the final position of the moving jaw. The wedge factors at the center of the field in both symmetric and asymmetric fields are examined. The difference between calculated and measured wedge factors is within 1.0%. The method developed here is easy to implement. The only datum required in addition to the standard set of conventional physical wedge implementation data is the off-axis output factor for the open field in the reference condition. The off-center point calculation is also examined. For the off-center point calculation, the dose profile in the wedge direction for the largest EDW field is used to obtain the relative off-center ratio in any smaller wedge field. The accuracy of the off-center point calculation decreases when the point of calculation is too close to the field edge.

[1]  M. Miften,et al.  Implementation of enhanced dynamic wedge in the focus rtp system. , 2000, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[2]  J C Rosenwald,et al.  The 'equivalent wedge' implementation of the Varian Enhanced Dynamic Wedge (EDW) into a treatment planning system. , 1999, Physics in medicine and biology.

[3]  J. Gibbons,et al.  Calculation of enhanced dynamic wedge factors for symmetric and asymmetric photon fields. , 1998, Medical physics.

[4]  J A Purdy,et al.  Multiple machine implementation of enhanced dynamic wedge. , 1998, International journal of radiation oncology, biology, physics.

[5]  T. Zhu,et al.  Commissioning of enhanced dynamic wedge on a ROCS RTP system. , 1997, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[6]  A W Beavis,et al.  Implementation of enhanced dynamic wedge into the Multidata DSS radiotherapy treatment planning system. , 1997, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[7]  M. Karlsson Implementation of Varian's EDW 5.2 on a Clinac 2300C/D for use with Helax TMS 3.1 dose planning system. , 1997, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[8]  T. Zhu,et al.  Performance evaluation of a diode array for enhanced dynamic wedge dosimetry. , 1997, Medical physics.

[9]  A W Beavis,et al.  Implementation of the Varian EDW into a commercial RTP system. , 1996, Physics in medicine and biology.

[10]  R L Siddon,et al.  Effective wedge angles with a universal wedge. , 1985, Physics in medicine and biology.

[11]  C Liu,et al.  Characterizing output for the Varian enhanced dynamic wedge field. , 1998, Medical physics.