Development of a high analytical performance-tyrosinase biosensor based on a composite graphite-Teflon electrode modified with gold nanoparticles.

The design of a new tyrosinase biosensor with improved stability and sensitivity is reported. The biosensor design is based on the construction of a graphite-Teflon composite electrode matrix in which the enzyme and colloidal gold nanoparticles are incorporated by simple physical inclusion. Experimental variables such as the colloidal gold loading into the composite matrix, the enzyme loading and the potential applied to the bioelectrode were optimized. The Tyr-Au(coll)-graphite-Teflon biosensor exhibited suitable amperometric responses at -0.10 V for the different phenolic compounds tested (catechol; phenol; 3,4-dimethylphenol; 4-chloro-3-methylphenol; 4-chlorophenol; 4-chloro-2-methylphenol; 3-methylphenol and 4-methylphenol). The limits of detection obtained were 3 nM for catechol, 3.3 microM for 4-chloro-2-methylphenol, and approximately 20 nM for the rest of phenolic compounds. The presence of colloidal gold into the composite matrix gives rise to enhanced kinetics of both the enzyme reaction and the electrochemical reduction of the corresponding o-quinones at the electrode surface, thus allowing the achievement of a high sensitivity. The biosensor exhibited an excellent renewability by simple polishing, with a lifetime of at least 39 days without apparent loss of the immobilized enzyme activity. The usefulness of the biosensor for the analysis of real samples was evaluated by performing the estimation of the content of phenolic compounds in water samples of different characteristics.

[1]  S. Campuzano,et al.  Amperometric flow-injection determination of phenolic compounds at self-assembled monolayer-based tyrosinase biosensors , 2003 .

[2]  Songqin Liu,et al.  Renewable phenol biosensor based on a tyrosinase-colloidal gold modified carbon paste electrode , 2003 .

[3]  P. Yáñez‐Sedeño,et al.  Gold nanoparticle-based electrochemical biosensors , 2005, Analytical and Bioanalytical Chemistry.

[4]  Songqin Liu,et al.  Renewable reagentless hydrogen peroxide sensor based on direct electron transfer of horseradish peroxidase immobilized on colloidal gold-modified electrode. , 2002, Analytical biochemistry.

[5]  M. L. Mena,et al.  Graphite-Teflon composite bienzyme amperometric biosensors for monitoring of alcohols. , 2003, Biosensors & bioelectronics.

[6]  J. Pingarrón,et al.  Graphite-teflon composite bienzyme electrodes for the determination of cholesterol in reversed micelles. Application to food samples. , 2001, Analytical chemistry.

[7]  S. Dong,et al.  Silica sol-gel composite film as an encapsulation matrix for the construction of an amperometric tyrosinase-based biosensor. , 2000, Biosensors & bioelectronics.

[8]  M. Natan,et al.  MORPHOLOGY-DEPENDENT ELECTROCHEMISTRY OF CYTOCHROME C AT AU COLLOID-MODIFIED SNO2 ELECTRODES , 1996 .

[9]  Itamar Willner,et al.  "Plugging into Enzymes": Nanowiring of Redox Enzymes by a Gold Nanoparticle , 2003, Science.

[10]  I. Willner,et al.  Nanostructured Gold Colloid Electrodes , 2000 .

[11]  J. Box Investigation of the Folin-Ciocalteau phenol reagent for the determination of polyphenolic substances in natural waters , 1983 .

[12]  J. Wang,et al.  Tyrosinase-based ruthenium dispersed carbon paste biosensor for phenols. , 1994, Biosensors & bioelectronics.

[13]  U. Wollenberger,et al.  Determination of phenolic compounds using recombinant tyrosinase from Streptomyces antibioticus , 2001 .

[14]  Aiguo Wu,et al.  A method to construct a third-generation horseradish peroxidase biosensor: self-assembling gold nanoparticles to three-dimensional sol-gel network. , 2002, Analytical chemistry.

[15]  J M Pingarrón,et al.  Composite electrochemical biosensors: a comparison of three different electrode matrices for the construction of amperometric tyrosinase biosensors. , 2002, Biosensors & bioelectronics.

[16]  L Campanella,et al.  Biosensors for determination of total and natural antioxidant capacity of red and white wines: comparison with other spectrophotometric and fluorimetric methods. , 2004, Biosensors & bioelectronics.

[17]  I. Willner,et al.  Organization of Au Colloids as Monolayer Films onto ITO Glass Surfaces: Application of the Metal Colloid Films as Base Interfaces To Construct Redox-Active Monolayers , 1995 .

[18]  Jenny Emnéus,et al.  Improved stability and altered selectivity of tyrosinase based graphite electrodes for detection of phenolic compounds , 1999 .

[19]  F. W. Gilcreas,et al.  Standard methods for the examination of water and waste water. , 1966, American journal of public health and the nation's health.

[20]  Silvia Fabiano,et al.  Amperometric tyrosinase based biosensor using an electrogenerated polythiophene film as an entrapment support. , 2003, Talanta.

[21]  Huangxian Ju,et al.  Reagentless glucose biosensor based on direct electron transfer of glucose oxidase immobilized on colloidal gold modified carbon paste electrode. , 2003, Biosensors & bioelectronics.

[22]  C. McNeil,et al.  Disposable tyrosinase-peroxidase bi-enzyme sensor for amperometric detection of phenols. , 2002, Biosensors & bioelectronics.

[23]  M. L. Mena,et al.  Development of a tyrosinase biosensor based on gold nanoparticles-modified glassy carbon electrodes: Application to the measurement of a bioelectrochemical polyphenols index in wines , 2005 .

[24]  Jenny Emnéus,et al.  Tyrosinase graphite-epoxy based composite electrodes for detection of phenols , 1995 .

[25]  J. Pingarrón,et al.  Detection of phenolic compounds in flow systems based on tyrosinase-modified reticulated vitreous carbon electrodes. , 2001, Talanta.

[26]  J. Pingarrón,et al.  Graphite‐Teflon‐Peroxidase Composite Electrochemical Biosensors. A Tool for the Wide Detection of Phenolic Compounds , 2001 .

[27]  Huangxian Ju,et al.  Mediator-free phenol sensor based on titania sol-gel encapsulation matrix for immobilization of tyrosinase by a vapor deposition method. , 2003, Biosensors & bioelectronics.

[28]  H. Ju,et al.  Direct electrochemistry of horseradish peroxidase immobilized on a colloid/cysteamine-modified gold electrode. , 2000, Analytical biochemistry.

[29]  Y. Hasebe,et al.  Highly sensitive electrochemical determination of Escherichia coli density using tyrosinase-based chemically amplified biosensor , 1997 .

[30]  J M Pingarrón,et al.  A comparison of different strategies for the construction of amperometric enzyme biosensors using gold nanoparticle-modified electrodes. , 2005, Analytical biochemistry.

[31]  Huangxian Ju,et al.  Application of Colloidal Gold in Protein Immobilization, Electron Transfer, and Biosensing , 2003 .

[32]  Aimin Yu,et al.  Direct electron transfer and characterization of hemoglobin immobilized on a Au colloid-cysteamine-modified gold electrode , 2001 .