Oxygen saturation estimation in brain tissue using diffuse reflectance spectroscopy along stereotactic trajectories.

Diffuse reflectance spectroscopy (DRS) can be used to estimate oxygen saturation (SO2) of hemoglobin and blood fraction (fB) in brain tissue. The aim of the study was to investigate the SO2 and fB in different positions along deep brain stimulation (DBS) trajectories and in specific target regions using DRS and a novel algorithm. DRS measurements were done at 166 well-defined anatomical positions in relation to stereotactic DBS-implantation along 20 trajectories toward 4 DBS targets (STN, Vim, GPi and Zi). The measurements were dived into groups (gray, white and light gray matter) related to anatomical position, and DBS targets, before comparison and statistical analysis. The median SO2 in gray, white and light gray matter were 52%, 24% and 20%, respectively. Median fB in gray matter (3.9%) was different from values in white (1.0%, p < 0.05) and light gray (0.9%, p < 0.001) matter. No significant difference in median SO2 and fB was found between DBS target regions. The novel algorithm allows for quick and reliable estimation of SO2 and fB in human brain tissue.

[1]  Peter Rejmstad,et al.  High-Resolution Laser Doppler Measurements of Microcirculation in the Deep Brain Structures: A Method for Potential Vessel Tracking , 2016, Stereotactic and Functional Neurosurgery.

[2]  Piotr Sawosz,et al.  Application of optical methods in the monitoring of traumatic brain injury: A review , 2016, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[3]  A. Spiotta,et al.  Brain Monitoring After Subarachnoid Hemorrhage: Lessons Learned , 2011, Neurosurgery.

[4]  A. Figaji,et al.  Methods of monitoring brain oxygenation , 2010, Child's Nervous System.

[5]  L. Kou,et al.  Refractive indices of water and ice in the 0.65- to 2.5-µm spectral range. , 1993, Applied optics.

[6]  Setsuo Takatani,et al.  Theoretical Analysis of Diffuse Reflectance from a Two-Layer Tissue Model , 1979, IEEE Transactions on Biomedical Engineering.

[7]  Oliver Ganslandt,et al.  Combined Laser‐Doppler Flowmetry and Spectrophotometry: Feasibility Study of a Novel Device for Monitoring Local Cortical Microcirculation during Aneurysm Surgery , 2017, Journal of neurological surgery. Part A, Central European neurosurgery.

[8]  Karin Wårdell,et al.  Intracerebral quantitative chromophore estimation from reflectance spectra captured during deep brain stimulation implantation , 2013, Journal of biophotonics.

[9]  Hanli Liu,et al.  A stereotactic near-infrared probe for localization during functional neurosurgical procedures: further experience. , 2009, Journal of neurosurgery.

[10]  Karin Wårdell,et al.  Optical Monitoring Techniques for Navigation during Stereotactic Neurosurgery , 2016 .

[11]  C. Epstein,et al.  Bedside Assessment of Tissue Oxygen Saturation Monitoring in Critically Ill Adults: An Integrative Review of the Literature , 2014, Critical care research and practice.

[12]  Peter Rejmstad,et al.  A laser Doppler system for monitoring cerebral microcirculation: implementation and evaluation during neurosurgery , 2015, Medical & Biological Engineering & Computing.

[13]  Hamid Dehghani,et al.  Near-Infrared Spectroscopy in the Monitoring of Adult Traumatic Brain Injury: A Review. , 2015, Journal of neurotrauma.

[14]  J. Severinghaus Simple, accurate equations for human blood O2 dissociation computations. , 1979, Journal of applied physiology: respiratory, environmental and exercise physiology.

[15]  Maureen Johns,et al.  Calculation of hemoglobin saturation from in vivo human brain tissues using a modified diffusion theory model , 2001, SPIE BiOS.

[16]  W. Hoffman,et al.  Measurement of ischemia by changes in tissue oxygen, carbon dioxide, and pH. , 1999, Surgical neurology.

[17]  P. L. Le Roux,et al.  Parenchymal brain oxygen monitoring in the neurocritical care unit. , 2013, Neurosurgery clinics of North America.

[18]  Johan Richter,et al.  Relationship Between Laser Doppler Signals and Anatomy During Deep Brain Stimulation Electrode Implantation Toward the Ventral Intermediate Nucleus and Subthalamic Nucleus , 2013, Neurosurgery.

[19]  Johan Richter,et al.  Diffuse reflectance spectroscopy measurements for tissue-type discrimination during deep brain stimulation , 2008, Journal of neural engineering.

[20]  Per Enblad,et al.  Brain tissue oxygen monitoring: a study of in vitro accuracy and stability of Neurovent-PTO and Licox sensors , 2010, Acta Neurochirurgica.

[21]  J. Johansson,et al.  Spectroscopic method for determination of the absorption coefficient in brain tissue. , 2010, Journal of biomedical optics.

[22]  Timothy C Ryken,et al.  Management of isolated fractures of the atlas in adults. , 2013, Neurosurgery.