Biomechanical effects of dental implant diameter, connection type, and bone density on microgap formation and fatigue failure: A finite element analysis

BACKGROUND AND OBJECTIVE Understanding fatigue failure and microgap formation in dental implants, abutments, and screws under various clinical circumstances is clinically meaningful. In this study, these aspects were evaluated based on implant diameter, connection type, and bone density. METHODS Twelve three-dimensional finite element models were constructed by combining two bone densities (low and high), two connection types (bone and tissue levels), and three implant diameters (3.5, 4.0, and 4.5 mm). Each model was composed of cortical and cancellous bone tissues, the nerve canal, and the implant complex. After the screw was preloaded, vertical (100 N) and oblique (200 N) loadings were applied. The relative displacements at the interfaces between implant, abutment, and screw were analyzed. The fatigue lives of the titanium alloy (Ti-6Al-4V) components were calculated through repetitive mastication simulations. Mann-Whitney U and Kruskal-Wallis one-way tests were performed on the 50 highest displacement values of each model. RESULTS At the implant/abutment interface, large microgaps were observed under oblique loading in the buccal direction. At the abutment/screw interface, microgap formation increased along the implant diameter under vertical loading but decreased under oblique loading (p < 0.001); the largest microgap formation occurred in the lingual direction. In all cases, the bone-level connection induced larger microgap formation than the tissue-level connections. Moreover, only the bone-level connection models showed fatigue failure, and the minimum fatigue life was observed for the implant diameter of 3.5 mm. CONCLUSIONS Tissue-level implants possess biomechanical advantages compared to bone-level ones. Two-piece implants with diameters below 3.5 mm should be avoided in the posterior mandibular area.

[1]  P. Weigl,et al.  The micromechanical behavior of implant-abutment connections under a dynamic load protocol. , 2018, Clinical implant dentistry and related research.

[2]  E. Pellizzer,et al.  Influence of tapered and external hexagon connections on bone stresses around tilted dental implants: three-dimensional finite element method with statistical analysis. , 2014, Journal of periodontology.

[3]  Sang‐Wan Shin,et al.  Effect of Implant Diameter on Fatigue Strength , 2017, Implant dentistry.

[4]  Rui-Feng Wang,et al.  The dynamic natures of implant loading. , 2009, The Journal of prosthetic dentistry.

[5]  N. H. Abu Kasim,et al.  Effects of different implant-abutment connections on micromotion and stress distribution: prediction of microgap formation. , 2012, Journal of dentistry.

[6]  Shanshan Gao,et al.  Micromotion of implant-abutment interfaces (IAI) after loading: correlation of finite element analysis with in vitro performances , 2019, Medical & Biological Engineering & Computing.

[7]  Xiao Zhang,et al.  Influence of custom-made implant designs on the biomechanical performance for the case of immediate post-extraction placement in the maxillary esthetic zone: a finite element analysis , 2017, Computer methods in biomechanics and biomedical engineering.

[8]  H Vaillancourt,et al.  Finite element analysis of crestal bone loss around porous-coated dental implants. , 1995, Journal of applied biomaterials : an official journal of the Society for Biomaterials.

[9]  D. Papadogiannis,et al.  Dynamic and static mechanical analysis of resin luting cements. , 2012, Journal of the mechanical behavior of biomedical materials.

[10]  E. Pellizzer,et al.  Finite element analysis on influence of implant surface treatments, connection and bone types. , 2016, Materials science & engineering. C, Materials for biological applications.

[11]  E. Pellizzer,et al.  Short dental implants versus standard dental implants placed in the posterior jaws: A systematic review and meta-analysis. , 2016, Journal of dentistry.

[12]  D. Rittel,et al.  Effect of dental implant diameter on fatigue performance. Part II: failure analysis. , 2014, Clinical implant dentistry and related research.

[13]  Wei Li,et al.  Bone’s responses to different designs of implant-supported fixed partial dentures , 2014, Biomechanics and Modeling in Mechanobiology.

[14]  E. Pellizzer,et al.  Effect of different types of prosthetic platforms on stress-distribution in dental implant-supported prostheses. , 2017, Materials science & engineering. C, Materials for biological applications.

[15]  Kazuhiko Yamamoto,et al.  Influence of bone parameters on peri-implant bone strain distribution in the posterior mandible , 2014, Medicina oral, patologia oral y cirugia bucal.

[16]  B. Stawarczyk,et al.  Bond strength between a high-performance thermoplastic and a veneering resin. , 2020, The Journal of prosthetic dentistry.

[17]  Davood Toghraie,et al.  Finite element analysis and experimental evaluation on stress distribution and sensitivity of dental implants to assess optimum length and thread pitch , 2019, Comput. Methods Programs Biomed..

[18]  M. A. Pérez,et al.  Life prediction of different commercial dental implants as influence by uncertainties in their fatigue material properties and loading conditions , 2012, Comput. Methods Programs Biomed..

[19]  Ching-Chi Hsu,et al.  Biomechanical analyses of static and dynamic fixation techniques of retrograde interlocking femoral nailing using nonlinear finite element methods , 2014, Comput. Methods Programs Biomed..

[20]  G. Noh,et al.  Effects of cementless fixation of implant prosthesis: A finite element study , 2019, The journal of advanced prosthodontics.

[21]  Sagar Kalburgi,et al.  Finite element analysis of immature teeth filled with MTA, Biodentine and Bioaggregate , 2020, Comput. Methods Programs Biomed..

[22]  Mitsuo Niinomi,et al.  Mechanical properties of biomedical titanium alloys , 1998 .

[23]  P. Noritomi,et al.  Biomechanical influence of crown-to-implant ratio on stress distribution over internal hexagon short implant: 3-D finite element analysis with statistical test. , 2015, Journal of biomechanics.

[24]  F. Mante,et al.  Evaluating parameters of osseointegrated dental implants using finite element analysis--a two-dimensional comparative study examining the effects of implant diameter, implant shape, and load direction. , 1998, The Journal of oral implantology.

[25]  Martin Rücker,et al.  Influence of lubricant on screw preload and stresses in a finite element model for a dental implant. , 2014, The Journal of prosthetic dentistry.

[26]  G. C. Silva,et al.  Biomechanical evaluation of screw- and cement-retained implant-supported prostheses: a nonlinear finite element analysis. , 2014, The Journal of prosthetic dentistry.

[27]  G. Noh,et al.  Biomechanical analysis of 4 types of short dental implants in a resorbed mandible , 2019, The Journal of prosthetic dentistry.

[28]  T. Guda,et al.  Probabilistic analysis of preload in the abutment screw of a dental implant complex. , 2008, The Journal of prosthetic dentistry.

[29]  Yasufumi Yamanishi,et al.  In vitro fatigue tests and in silico finite element analysis of dental implants with different fixture/abutment joint types using computer-aided design models. , 2018, Journal of prosthodontic research.

[30]  G. van der Perre,et al.  Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog. , 1998, Journal of oral rehabilitation.

[31]  M. Swain,et al.  Fracture resistance of titanium and zirconia abutments: an in vitro study. , 2013, The Journal of prosthetic dentistry.

[32]  Byungsik Kang,et al.  Finite element analysis to determine implant preload. , 2003, The Journal of prosthetic dentistry.

[33]  Valentim Adelino Ricardo Barão,et al.  Effect of different mucosa thickness and resiliency on stress distribution of implant-retained overdentures-2D FEA , 2008, Comput. Methods Programs Biomed..

[34]  G. Noh,et al.  Biomechanical stability of internal bone-level implant:Dependency on hex or non-hex structure , 2020 .

[35]  T. Eliades,et al.  In Vitro Simulation and In Vivo Assessment of Tooth Wear: A Meta-Analysis of In Vitro and Clinical Research , 2019, Materials.

[36]  M. Özcan,et al.  Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis. , 2016, Journal of the mechanical behavior of biomedical materials.

[37]  F. Geramipanah,et al.  Effect of increased crown height on stress distribution in short dental implant components and their surrounding bone: A finite element analysis. , 2015, The Journal of prosthetic dentistry.

[38]  W. Müller,et al.  Finite element analysis of the biomechanical effects of PEEK dental implants on the peri-implant bone. , 2015, Journal of biomechanics.

[39]  A. Rack,et al.  In vitro synchrotron-based radiography of micro-gap formation at the implant–abutment interface of two-piece dental implants , 2010, Journal of synchrotron radiation.

[40]  E. Pellizzer,et al.  Influence of parafunctional loading and prosthetic connection on stress distribution: a 3D finite element analysis. , 2015, The Journal of prosthetic dentistry.

[41]  A. Berbéri,et al.  Micromovement Evaluation of Original and Compatible Abutments at the Implant-abutment Interface. , 2016, The journal of contemporary dental practice.

[42]  D. Rittel,et al.  Fatigue failure of dental implants in simulated intraoral media. , 2016, Journal of the mechanical behavior of biomedical materials.

[43]  E. Pellizzer,et al.  Can the modeling for simplification of a dental implant surface affect the accuracy of 3D finite element analysis? , 2016, Computer methods in biomechanics and biomedical engineering.

[44]  M. Sperandio,et al.  Association Between Implant-Abutment Microgap and Implant Circularity to Bacterial Leakage: An In Vitro Study Using Tapered Connection Implants. , 2018, The International journal of oral & maxillofacial implants.