A Feasibility Study of a Prototype PET Insert Device to Convert a General-Purpose Animal PET Scanner to Higher Resolution

We developed a prototype system to evaluate the feasibility of using a PET insert device to achieve higher resolution from a general-purpose animal PET scanner. Methods: The system consists of a high-resolution PET detector, a computer-controlled rotation stage, and a custom mounting plate. The detector consists of a cerium-doped lutetium oxyorthosilicate array (12 × 12 crystals, 0.8 × 1.66 × 3.75 mm3 each) directly coupled to a position-sensitive photomultiplier tube (PS-PMT). The detector signals were fed into the scanner electronics to establish coincidences between the 2 systems. The detector was mounted to a rotation stage that is attached to the scanner via the custom mounting plate after removing the transmission source holder. The rotation stage was concentric with the center of the scanner. The angular offset of the insert detector was calibrated via optimizing point-source images. In all imaging experiments, coincidence data were collected from 9 angles to provide 180° sampling. A 22Na point source was imaged at different offsets from the center to characterize the in-plane resolution of the insert system. A 68Ge point source was stepped across the axial field of view to measure the sensitivity of the system. A 23.2-g mouse was injected with 38.5 MBq of 18F-fluoride and imaged at 3 h after injection for 2 h. Results: The transverse image resolution of the PET insert device ranges from 1.1- to 1.4-mm full width at half maximum (FWHM) without correction for the point-source dimension. This corresponds to approximately 33% improvement over the resolution of the original scanner (1.7- to 1.8-mm FWHM) in 2 of the 3 directions. The sensitivity of the device is 0.064% at the center of the field, 46-fold lower than the sensitivity of an existing animal PET scanner. The mouse bone scan had improved image resolution using the PET insert device over that of the existing animal PET scanner alone. Conclusion: We have demonstrated the feasibility of using a high-resolution insert device in an existing PET scanner to provide high-resolution PET. A PET insert device with more detector modules will improve sensitivity and may become an alternative to special-purpose PET systems for high-resolution PET.

[1]  A. Del Guerra,et al.  YAP-PET: first results of a small animal positron emission tomograph based on YAP:Ce finger crystals , 1998 .

[2]  M. Phelps,et al.  PET: the merging of biology and imaging into molecular imaging. , 2000, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[3]  B. Hasegawa,et al.  A multipinhole small animal SPECT system with submillimeter spatial resolution. , 2006, Medical physics.

[4]  James F. Young,et al.  MicroPET: a high resolution PET scanner for imaging small animals , 1996, IEEE Nuclear Science Symposium Conference Record.

[5]  Roger Lecomte,et al.  Initial results from the Sherbrooke avalanche photodiode positron tomograph , 1996 .

[6]  Yuan-Chuan Tai,et al.  2D linear and iterative reconstruction algorithms for a PET-insert scanner. , 2007, Physics in medicine and biology.

[7]  Donald W. Wilson,et al.  SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. , 2006, Medical physics.

[8]  M. Janecek,et al.  A simulation study for the design of a prototype insert for whole-body PET scanners , 2006, IEEE Transactions on Nuclear Science.

[9]  Peter Bruyndonckx,et al.  Performance study of a 3D small animal PET scanner based on BaF2 crystals and a photo sensitive wire chamber , 1997 .

[10]  T J Spinks,et al.  The design and physical characteristics of a small animal positron emission tomograph. , 1995, Physics in medicine and biology.

[11]  A. P. Jeavons,et al.  A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals , 1998 .

[12]  M. Schmand,et al.  Design of a small animal PET imaging system with 1 microliter volume resolution , 2004, IEEE Transactions on Nuclear Science.

[13]  M. Daube-Witherspoon,et al.  Treatment of axial data in three-dimensional PET. , 1987, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  Hiroyuki Okada,et al.  A high resolution animal PET scanner using compact PS-PMT detectors , 1996 .

[15]  P. V. van Rijk,et al.  U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[16]  S. Cherry,et al.  MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging. , 2003, Physics in medicine and biology.

[17]  S. Siegel,et al.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[18]  Michel Defrise,et al.  Exact and approximate rebinning algorithms for 3-D PET data , 1997, IEEE Transactions on Medical Imaging.