Observations on retrieved glenoid components.

To obtain more information on the pattern of damage of prosthetic glenoid components, we analyzed 7 retrieved glenoid components. The consecutive series included 2 standard polyethylene components and 5 highly crystalline polyethylene glenoids (Hylamer; DePuy Dupont Orthopaedics, Warsaw, IN) retrieved 3 to 12 years after implantation. At revision, 4 of 5 Hylamer components were fractured. Common wear patterns were i) deformation and crumbling of the rim, particularly at the inferior hemicircumference, probably caused by direct contact of the humerus with the prosthetic component; ii) roughening (abrasion and scratching) of the adjacent articulating surface; and iii) concentric and congruous wear centered posteriorly. Available glenoid components may cover an excessive sector of the head. This can result in mechanical restriction of glenohumeral motion and abutment of the humerus against the glenoid rim. Abutment may cause major shear forces and therefore cause glenoid loosening. The value of articular surface mismatch is questionable because retrieved glenoids were worn to a conforming joint.

[1]  V C Mow,et al.  Articular geometry of the glenohumeral joint. , 1992, Clinical orthopaedics and related research.

[2]  G. Walch,et al.  Normal and Pathological Anatomy of the Glenoid: Effects on the Design, Preparation, and Fixation of the Glenoid Component , 1999 .

[3]  J. Iannotti,et al.  The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. , 1992, The Journal of bone and joint surgery. American volume.

[4]  R. L. Dooley,et al.  Anthropometric study of normal glenohumeral relationships. , 1997, Journal of shoulder and elbow surgery.

[5]  S M Howell,et al.  The glenoid-labral socket. A constrained articular surface. , 1989, Clinical orthopaedics and related research.

[6]  N. Poppen,et al.  Normal and abnormal motion of the shoulder. , 1976, The Journal of bone and joint surgery. American volume.

[7]  F A Matsen,et al.  Observations on retrieved polyethylene glenoid components. , 2001, The Journal of arthroplasty.

[8]  A H Burstein,et al.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. , 1983, Journal of biomedical materials research.

[9]  R. Cofield,et al.  Total shoulder arthroplasty with the Neer prosthesis: long-term results. , 1997, Journal of shoulder and elbow surgery.

[10]  R. J. Pawluk,et al.  Glenohumeral mechanics: a study of articular geometry, contact, and kinematics. , 2001, Journal of shoulder and elbow surgery.

[11]  J A Sidles,et al.  The effect of articular conformity and the size of the humeral head component on laxity and motion after glenohumeral arthroplasty. A study in cadavera. , 1995, The Journal of bone and joint surgery. American volume.

[12]  C. Sledge,et al.  Early failure of Hylamer acetabular inserts due to eccentric wear. , 1996, The Journal of arthroplasty.

[13]  R. J. Pawluk,et al.  Quantitation of in situ contact areas at the glenohumeral joint: A biomechanical study , 1992, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[14]  F A Matsen,et al.  Intrinsic stability of unused and retrieved polyethylene glenoid components. , 2001, Journal of shoulder and elbow surgery.

[15]  Jove Graham,et al.  Retrieved glenoid components: a classification system for surface damage analysis. , 2002, The Journal of arthroplasty.

[16]  J A Sidles,et al.  Glenohumeral stability from concavity-compression: A quantitative analysis. , 1993, Journal of shoulder and elbow surgery.

[17]  J. Kabo,et al.  The influence of conformity and constraint on translational forces and frictional torque in total shoulder arthroplasty. , 1993, Clinical orthopaedics and related research.