Resonant Mie scattering (RMieS) correction of infrared spectra from highly scattering biological samples.

Infrared spectra of single biological cells often exhibit the 'dispersion artefact' observed as a sharp decrease in intensity on the high wavenumber side of absorption bands, in particular the Amide I band at approximately 1655 cm(-1), causing a downward shift of the true peak position. The presence of this effect makes any biochemical interpretation of the spectra unreliable. Recent theory has shed light on the origins of the 'dispersion artefact' which has been attributed to resonant Mie scattering (RMieS). In this paper a preliminary algorithm for correcting RMieS is presented and evaluated using simulated data. Results show that the 'dispersion artefact' appears to be removed; however, the correction is not perfect. An iterative approach was subsequently implemented whereby the reference spectrum is improved after each iteration, resulting in a more accurate correction. Consequently the corrected spectra become increasingly more representative of the pure absorbance spectra. Using this correction method reliable peak positions can be obtained.

[1]  M. Diem,et al.  Mie-type scattering and non-Beer-Lambert absorption behavior of human cells in infrared microspectroscopy. , 2005, Biophysical journal.

[2]  Paul Dumas,et al.  Resonant Mie scattering in infrared spectroscopy of biological materials--understanding the 'dispersion artefact'. , 2009, The Analyst.

[3]  Paul Dumas,et al.  THE USE OF SYNCHROTRON INFRARED MICROSPECTROSCOPY IN BIOLOGICAL AND BIOMEDICAL INVESTIGATIONS , 2003 .

[4]  Ehsan Gazi,et al.  Optical artefacts in transflection mode FTIR microspectroscopic images of single cells on a biological support: the effect of back-scattering into collection optics. , 2007, The Analyst.

[5]  G. Laurie,et al.  Basement membrane complexes with biological activity. , 1986, Biochemistry.

[6]  A. Kohler,et al.  Estimating and Correcting Mie Scattering in Synchrotron-Based Microscopic Fourier Transform Infrared Spectra by Extended Multiplicative Signal Correction , 2008, Applied spectroscopy.

[7]  B. Beccard,et al.  Imaging capabilities of synchrotron infrared microspectroscopy. , 2004, Faraday discussions.

[8]  C. Tropea,et al.  Light Scattering from Small Particles , 2003 .

[9]  Max Diem,et al.  Spatially resolved IR microspectroscopy of single cells. , 2002, Biopolymers.

[10]  M. Diem,et al.  Infrared micro-spectroscopy of human cells: Causes for the spectral variance of oral mucosa (buccal) cells. , 2006, Vibrational Spectroscopy.

[11]  R. Kronig On the Theory of Dispersion of X-Rays , 1926 .

[12]  Benjamin Bird,et al.  Cytology by Infrared Micro-Spectroscopy: Automatic Distinction of Cell Types in Urinary Cytology. , 2008, Vibrational Spectroscopy.

[13]  Francis L Martin,et al.  Infrared spectroscopy with multivariate analysis potentially facilitates the segregation of different types of prostate cell. , 2006, Biophysical journal.

[14]  Ehsan Gazi,et al.  A study of cytokinetic and motile prostate cancer cells using synchrotron-based FTIR microspectroscopic imaging , 2005 .

[15]  M. Diem,et al.  Infrared microspectroscopy of individual human cervical cancer (HeLa) cells. , 2004, Biopolymers.

[16]  M. Diem,et al.  Fourier transform infrared (FTIR) spectral mapping of the cervical transformation zone, and dysplastic squamous epithelium. , 2004, Gynecologic oncology.

[17]  P. Lasch,et al.  FT-IR spectroscopic investigations of single cells on the subcellular level , 2002 .

[18]  W. Fridman,et al.  Highly resolved chemical imaging of living cells by using synchrotron infrared microspectrometry. , 1998, Proceedings of the National Academy of Sciences of the United States of America.

[19]  Max Diem,et al.  Correction of dispersive line shape artifact observed in diffuse reflection infrared spectroscopy and absorption/reflection (transflection) infrared micro-spectroscopy , 2005 .

[20]  Rainer Pepperkok,et al.  IR microspectroscopy of live cells , 2005 .

[21]  A. Kohler,et al.  Spectroscopic signatures of single, isolated cancer cell nuclei using synchrotron infrared microscopy. , 2009, The Analyst.

[22]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[23]  G. Mie Beiträge zur Optik trüber Medien, speziell kolloidaler Metallösungen , 1908 .

[24]  J Dwyer,et al.  Fixation protocols for subcellular imaging by synchrotron‐based Fourier transform infrared microspectroscopy , 2005, Biopolymers.

[25]  Miss A.O. Penney (b) , 1974, The New Yale Book of Quotations.

[26]  O. Chubar,et al.  Synchrotron infrared microscopy at the French Synchrotron Facility SOLEIL , 2006 .

[27]  M. Diem,et al.  Infrared spectroscopy of human tissue. V. Infrared spectroscopic studies of myeloid leukemia (ML-1) cells at different phases of the cell cycle. , 1999, Biospectroscopy.

[28]  N. Clarke,et al.  Direct evidence of lipid translocation between adipocytes and prostate cancer cells with imaging FTIR microspectroscopy Published, JLR Papers in Press, May 11, 2007. , 2007, Journal of Lipid Research.

[29]  H. Martens,et al.  Light scattering and light absorbance separated by extended multiplicative signal correction. application to near-infrared transmission analysis of powder mixtures. , 2003, Analytical chemistry.

[30]  A. Kohler,et al.  FT-IR microspectroscopy as a tool to assess lung cancer cells response to chemotherapy , 2005 .

[31]  Paul Dumas,et al.  Reflection contributions to the dispersion artefact in FTIR spectra of single biological cells. , 2009, The Analyst.

[32]  M. Hartmann,et al.  Light scattering by small particles. Von H. C. VANDE HULST. New York: Dover Publications, Inc. 1981. Paperback, 470 S., 103 Abb. und 46 Tab., US $ 7.50 , 1984 .

[33]  L. Fowler,et al.  Application of immunohistochemistry to cytology. , 2009, Archives of pathology & laboratory medicine.