Comparison of STOIC 1989 ground‐based lidar, microwave spectrometer, and Dobson spectrophotometer Umkehr ozone profiles with ozone profiles from balloon‐borne electrochemical concentration cell ozonesondes

Ground-based measurements of stratospheric ozone using a Jet Propulsion Laboratory (JPL) lidar, a NASA Goddard Space Flight Center (GSFC) lidar, a Millitech Corporation/NASA Langley Research Center (Millitech/LaRC) microwave spectrometer, and a NOAA Dobson ozone spectrophotometer were compared with in situ measurements made quasi-simultaneously with balloon-borne electrochemical concentration cell (ECC) ozonesondes during 10 days of the Stratospheric Ozone Intercomparison Campaign (STOIC). The campaign was conducted at Table Mountain Observatory, California, during the summer of 1989. ECC ozonesondes were flown by NOAA Climate Monitoring and Diagnostics Laboratory (CMDL) personnel as well as by personnel from the NASA Wallops Island Flight Facility (WFF). Within the altitude range of 20–32 km, ozone measurement precisions were estimated to be ±0.6 to ±1.2% for the JPL lidar, ±0.7% for the GSFC lidar, ±4% for the microwave spectrometer, and ±3% for the NOAA ECC ozonesonde instruments. These precisions decreased in the 32 to 38.6-km altitude range to ±1.3, ±1.5, and ±3% to ±10% for the JPL lidar, GSFC lidar, and the ECC sondes, respectively, but remained at ±4% for the microwave instrument. Ozone measurement accuracies in the 20 to 32 km altitude range were estimated to be ±1.2 to ±2.4% for the JPL lidar, ±1.4% for the GSFC lidar, ±6% for the microwave radiometer, and ±5% for the ECC ozonesondes. The accuracies decreased in the 32 to 38.6-km altitude range to ±2.6, ±3.0, ±7, and 1 ± 4% to −4 ± 10% for the JPL lidar, the GSFC lidar, the microwave spectrometer, and the ECC ozonesondes, respectively. While accuracy estimates for the ECC sondes were obtained by combining random and estimated bias errors, the accuracies for the lidar instruments were obtained by doubling the measurement precision figures, with the assumption that such doubling accounts for systematic errors. Within the altitude range of 20–36 km the mean ozone profiles produced by the JPL, GSFC, and the Millitech/LaRC groups did not differ from the mean ECC sonde ozone profile by more than about 2, 4, and 5%, respectively. Six morning Dobson instrument Umkehr observations yielded mean ozone amounts in layers 3 and 5–7 that agreed with comparison ECC ozonesonde data to within ±4%. In layer 4 the difference was 7.8%. (Less favorable comparison data were obtained for six afternoon Umkehr observations made in highly polluted near-surface air.) This good agreement in overall results obtained lends credence to the reliability of the ozone measurements made at Table Mountain Observatory during STOIC 1989.

[1]  Robert A. Barnes,et al.  Electrochemical concentration cell ozonesonde performance evaluation during STOIC 1989 , 1995 .

[2]  T. D. Walsh,et al.  STRATOSPHERIC OZONE INTERCOMPARISON CAMPAIGN (STOIC) 1989 : OVERVIEW , 1995 .

[3]  W. Komhyr,et al.  Effective Bass‐Paur 1985 ozone absorption coefficients for use with Dobson ozone spectrophotometers , 1993 .

[4]  B. Connor,et al.  Ground‐based microwave monitoring of stratospheric ozone , 1992 .

[5]  Brian J. Connor,et al.  Detection of stratospheric ozone trends by ground-based microwave observations , 1991, Defense, Security, and Sensing.

[6]  J. Butler,et al.  STROZ LITE: Stratospheric Ozone Lidar Trailer Experiment , 1991 .

[7]  William P. Chu,et al.  Comparison of ozone profiles from ground-based lidar, electrochemical concentration cell balloon sonde, ROCOZ-A rocket ozonesonde, and Stratospheric Aerosol and Gas Experiment satellite measurements , 1990 .

[8]  C. Rodgers Characterization and Error Analysis of Profiles Retrieved From Remote Sounding Measurements , 1990 .

[9]  W. Evans,et al.  Results from the balloon ozone intercomparison campaign (BOIC) , 1986 .

[10]  J. Pelon,et al.  Lidar measurements of ozone vertical profiles. , 1985, Applied Optics.

[11]  A. Bandy,et al.  Electrochemical concentration cell ozonesonde accuracy and precision , 1985 .

[12]  S. Oltmans,et al.  Performance Characteristics of High-Altitude ECC Ozonesondes , 1985 .

[13]  R. J. Paur,et al.  The ultraviolet cross-sections of ozone. I. The measurements. II - Results and temperature dependence , 1985 .

[14]  M. Proffitt,et al.  Fast‐response dual‐beam UV‐absorption ozone photometer suitable for use on stratospheric balloons , 1983 .

[15]  Shortened version of the Umkehr method for observing the vertical distribution of ozone. , 1979, Applied optics.

[16]  W. Komhyr,et al.  Development of an ECC Ozonesonde , 1971 .

[17]  E. Vigroux Détermination des coefficients moyens d'absorption de l'ozone en vue des observations concernant l'ozone atmosphérique à l'aide du spectromètre dobson , 1967 .

[18]  Ernest Vigroux,et al.  Contribution à l'étude expérimentale de l'absorption de l'ozone , 1953 .

[19]  A. R. Meetham,et al.  Vertical Distribution of Ozone in the Atmosphere , 1933, Nature.