Effect of deltoid tension and humeral version in reverse total shoulder arthroplasty: a biomechanical study.

BACKGROUND No clear recommendations exist regarding optimal humeral component version and deltoid tension in reverse total shoulder arthroplasty (TSA). MATERIALS AND METHODS A biomechanical shoulder simulator tested humeral versions (0°, 10°, 20° retroversion) and implant thicknesses (-3, 0, +3 mm from baseline) after reverse TSA in human cadavers. Abduction and external rotation ranges of motion as well as abduction and dislocation forces were quantified for native arms and arms implanted with 9 combinations of humeral version and implant thickness. RESULTS Resting abduction angles increased significantly (up to 30°) after reverse TSA compared with native shoulders. With constant posterior cuff loads, native arms externally rotated 20°, whereas no external rotation occurred in implanted arms (20° net internal rotation). Humeral version did not affect rotational range of motion but did alter resting abduction. Abduction forces decreased 30% vs native shoulders but did not change when version or implant thickness was altered. Humeral center of rotation was shifted 17 mm medially and 12 mm inferiorly after implantation. The force required for lateral dislocation was 60% less than anterior and was not affected by implant thickness or version. CONCLUSION Reverse TSA reduced abduction forces compared with native shoulders and resulted in limited external rotation and abduction ranges of motion. Because abduction force was reduced for all implants, the choice of humeral version and implant thickness should focus on range of motion. Lateral dislocation forces were less than anterior forces; thus, levering and inferior/posterior impingement may be a more probable basis for dislocation (laterally) than anteriorly directed forces.

[1]  D. Pupello,et al.  Range of impingement-free abduction and adduction deficit after reverse shoulder arthroplasty. Hierarchy of surgical and implant-design-related factors. , 2008, The Journal of bone and joint surgery. American volume.

[2]  J. Tibone,et al.  The effect of the long head of the biceps on glenohumeral kinematics. , 2009, Journal of shoulder and elbow surgery.

[3]  L. D. De Wilde,et al.  Prosthetic component relationship of the reverse Delta III total shoulder prosthesis in the transverse plane of the body. , 2008, Journal of shoulder and elbow surgery.

[4]  G. Walch,et al.  Reverse total shoulder arthroplasty. Survivorship analysis of eighty replacements followed for five to ten years. , 2006, The Journal of bone and joint surgery. American volume.

[5]  James A. Johnson,et al.  Humeral head translation decreases with muscle loading. , 2008, Journal of shoulder and elbow surgery.

[6]  Marcus G Pandy,et al.  Contributions of the individual muscles of the shoulder to glenohumeral joint stability during abduction. , 2008, Journal of biomechanical engineering.

[7]  Gilles Walch,et al.  Reverse total shoulder arthroplasty: a review of results according to etiology. , 2007, The Journal of bone and joint surgery. American volume.

[8]  W Plitz,et al.  A dynamic shoulder model: reliability testing and muscle force study. , 1995, Journal of biomechanics.

[9]  Christian Gerber,et al.  Biomechanical relevance of glenoid component positioning in the reverse Delta III total shoulder prosthesis. , 2005, Journal of shoulder and elbow surgery.

[10]  Jason S Scibek,et al.  Shoulder kinematics in patients with full-thickness rotator cuff tears after a subacromial injection. , 2008, Journal of shoulder and elbow surgery.

[11]  Freddie H. Fu,et al.  A Dynamic Analysis of Glenohumeral Motion after Simulated Capsulolabral Injury. A Cadaver Model* , 1998, The Journal of bone and joint surgery. American volume.

[12]  Russell F Warren,et al.  Biomechanics of massive rotator cuff tears: implications for treatment. , 2008, The Journal of bone and joint surgery. American volume.

[13]  A Terrier,et al.  Simulated joint and muscle forces in reversed and anatomic shoulder prostheses. , 2008, The Journal of bone and joint surgery. British volume.

[14]  P. Grammont,et al.  Delta shoulder prosthesis for rotator cuff rupture. , 1993, Orthopedics.

[15]  Christian Gerber,et al.  The effect of component positioning on intrinsic stability of the reverse shoulder arthroplasty. , 2010, Journal of shoulder and elbow surgery.

[16]  Pierre Trouilloud,et al.  Etude et réalisation d'une nouvelle prothèse d'épaule , 1987 .

[17]  Matthew Vasey,et al.  The Reverse Shoulder Prosthesis for glenohumeral arthritis associated with severe rotator cuff deficiency. A minimum two-year follow-up study of sixty patients. , 2005, The Journal of bone and joint surgery. American volume.

[18]  Pascal Boileau,et al.  Grammont reverse prosthesis: design, rationale, and biomechanics. , 2005, Journal of shoulder and elbow surgery.

[19]  D D Robertson,et al.  Three-Dimensional Analysis of the Proximal Part of the Humerus: Relevance to Arthroplasty* , 2000, The Journal of bone and joint surgery. American volume.

[20]  P. McMahon,et al.  A novel cadaveric model for anterior-inferior shoulder dislocation using forcible apprehension positioning. , 2003, Journal of rehabilitation research and development.

[21]  C Gerber,et al.  Treatment of painful pseudoparesis due to irreparable rotator cuff dysfunction with the Delta III reverse-ball-and-socket total shoulder prosthesis. , 2005, The Journal of bone and joint surgery. American volume.

[22]  Gilles Walch,et al.  The Three-Dimensional Geometry of the Proximal Humerus: Implications for Surgical Technique and Prosthetic Design , 1997 .

[23]  M. Pandy,et al.  Moment arms of the shoulder musculature after reverse total shoulder arthroplasty. , 2010, The Journal of bone and joint surgery. American volume.

[24]  N. Wülker,et al.  The effect of negative intraarticular pressure and rotator cuff force on glenohumeral translation during simulated active elevation. , 2000, Clinical biomechanics.

[25]  H. Uhthoff,et al.  The effect of posterior capsular tightening on peak subacromial contact pressure during simulated active abduction in the scapular plane. , 2010, Journal of shoulder and elbow surgery.

[26]  T. Zantop,et al.  Effect of coracoacromial ligament resection on glenohumeral stability under active muscle loading in an in vitro model. , 2008, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[27]  James A. Johnson,et al.  The effect of muscle loading on the kinematics of in vitro glenohumeral abduction. , 2007, Journal of biomechanics.

[28]  Hiroaki Tsutsui,et al.  The influence of handheld weight on the scapulohumeral rhythm. , 2008, Journal of shoulder and elbow surgery.

[29]  W. Warme,et al.  A method for documenting the change in center of rotation with reverse total shoulder arthroplasty and its application to a consecutive series of 68 shoulders having reconstruction with one of two different reverse prostheses. , 2010, Journal of shoulder and elbow surgery.

[30]  Matthew Vasey,et al.  The reverse shoulder prosthesis for glenohumeral arthritis associated with severe rotator cuff deficiency. a minimum two-year follow-up study of sixty patients surgical technique. , 2006, The Journal of bone and joint surgery. American volume.

[31]  Freddie H. Fu,et al.  Experimental investigation of reaction forces at the glenohumeral joint during active abduction. , 2000, Journal of shoulder and elbow surgery.

[32]  R. Churchill,et al.  Glenoid size, inclination, and version: an anatomic study. , 2001, Journal of shoulder and elbow surgery.

[33]  Maxwell C. Park,et al.  Influence of distinct anatomic subregions of the supraspinatus on humeral rotation , 2009, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[34]  J. Lachin,et al.  Properties of permuted-block randomization in clinical trials. , 1988, Controlled clinical trials.

[35]  L. D. De Wilde,et al.  Reliability of the glenoid plane. , 2010, Journal of shoulder and elbow surgery.

[36]  I. Anderson,et al.  Biomechanical evaluation of different designs of glenospheres in the SMR reverse total shoulder prosthesis: range of motion and risk of scapular notching. , 2009, Journal of shoulder and elbow surgery.

[37]  D. D’Lima,et al.  Accuracy of CT-based measurements of glenoid version for total shoulder arthroplasty. , 2010, Journal of shoulder and elbow surgery.

[38]  D. Molé,et al.  [Excentered scapulohumeral osteoarthritis]. , 2007, Revue de chirurgie orthopedique et reparatrice de l'appareil moteur.