EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip.

Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse--femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be +/- 1.6 mm for the x-direction and +/- 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively. EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.

[1]  P B Pynsent,et al.  Metal-on-metal resurfacing of the hip in patients under the age of 55 years with osteoarthritis. , 2004, The Journal of bone and joint surgery. British volume.

[2]  P. Beaulé,et al.  Risk Factors Affecting Outcome of Metal-on-Metal Surface Arthroplasty of the Hip , 2004, Clinical orthopaedics and related research.

[3]  P. Campbell,et al.  Metal-on-metal hybrid surface arthroplasty: two to six-year follow-up study. , 2004, The Journal of bone and joint surgery. American volume.

[4]  R. Sinha Hip Replacement: Current Trends and Controversies , 2002 .

[5]  P. Beaulé,et al.  Surface Arthroplasty of the Hip Revisited: Current Indications And Surgical Technique , 2002 .

[6]  R. Porcher,et al.  Hydroxyapatite-coated versus grit-blasted femoral stems. a prospective, randomised study using EBRA-FCA. , 2001, The Journal of bone and joint surgery. British volume.

[7]  M. Krismer,et al.  Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement , 1999 .

[8]  M. Krismer,et al.  The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA , 1999 .

[9]  T. Ilchmann,et al.  Migration, Subsidence and Wear after Hip Replacement , 1999 .

[10]  C. Rieker,et al.  Analysis of 118 second-generation metal-on-metal retrieved hip implants , 1999 .

[11]  C. Rieker,et al.  Analysis of 118 second-generation metal-on-metal retrieved hip implants. , 1999, The Journal of bone and joint surgery. British volume.

[12]  I. Learmonth,et al.  Tantalum ball position after total hip arthroplasty. , 1998, The Journal of bone and joint surgery. British volume.

[13]  K. Hardinge,et al.  Migration and wear of long-term successful Charnley total hip replacements. , 1998, The Journal of bone and joint surgery. British volume.

[14]  J Kärrholm,et al.  Radiostereometry of hip prostheses. Review of methodology and clinical results. , 1997, Clinical orthopaedics and related research.

[15]  M. Freeman,et al.  Early radiological observations may predict the long-term survival of femoral hip prostheses. , 1997, The Journal of bone and joint surgery. British volume.

[16]  R Bauer,et al.  EBRA: a method to measure migration of acetabular components. , 1995, Journal of biomechanics.

[17]  P S Walker,et al.  Prediction of clinical outcome of THR from migration measurements on standard radiographs. A study of cemented Charnley and Stanmore femoral stems. , 1995, The Journal of bone and joint surgery. British volume.

[18]  J. Kärrholm,et al.  Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses. , 1994, The Journal of bone and joint surgery. British volume.

[19]  B. Mjöberg Theories of wear and loosening in hip prostheses. Wear-induced loosening vs loosening-induced wear--a review. , 1994, Acta orthopaedica Scandinavica.

[20]  M A Freeman,et al.  Early migration and late aseptic failure of proximal femoral prostheses. , 1994, The Journal of bone and joint surgery. British volume.

[21]  J. Kärrholm,et al.  Subsidence, tip, and hump micromovements of noncoated ribbed femoral prostheses. , 1993, Clinical orthopaedics and related research.

[22]  D W Hukins,et al.  Measurement of hip prostheses using image analysis. The maxima hip technique. , 1991, The Journal of bone and joint surgery. British volume.

[23]  G Selvik,et al.  In vivo measurements of relative motion between an uncemented femoral total hip component and the femur by roentgen stereophotogrammetric analysis. , 1991, Clinical orthopaedics and related research.

[24]  G. Selvik Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. , 1989, Acta orthopaedica Scandinavica. Supplementum.

[25]  K E Marks,et al.  A ten-year follow-up of one hundred consecutive Müller curved-stem total hip-replacement arthroplasties. , 1982, The Journal of bone and joint surgery. American volume.