Spectral transmittance measurement nonlinearity of a (280-540) nm spectrophotometer based on a tunable femtosecond pulse laser

Femtosecond pulses tunable in (280-540) nm were generated via second and third harmonic generation processes from a Ti:Sapphire femtosecond laser with a 80 MHz pulse repetition rate. The temporal widths of the pulses are ~200 fs and the spectral bandwidths are <10 nm. These pulses can be used for spectral transmittance measurement using an integrating sphere photodetector with less than 0.1% fluorescence effect. The spectral transmittance measurement nonlinearity of the femtosecond pulse laser spectrophotometer was characterized using a double-path scheme for the incident power level from 1 mW to 10 μW. The nonlinearity results were compared with those obtained using a 532 nm continuous wave laser at similar power levels. The spectrophotometer can be well applied for the spectral transmittance measurement of a wide variety of neutral optical materials after nonlinearity corrections.

[1]  Masashi Kawasaki,et al.  Resonant Exciton Second-harmonic Generation in Self-assembled ZnO Microcrystallite Thin Films , 2003 .

[2]  T. Iwasaki A single-mode optical fiber attenuator for intercomparison of attenuation , 1994 .

[3]  Chao Zhang,et al.  Third-harmonic generation in a general two-component quasi-periodic optical superlattice. , 2001, Optics letters.

[4]  L. Hanssen,et al.  Integrating-sphere system and method for absolute measurement of transmittance, reflectance, and absorptance of specular samples. , 2001, Applied optics.

[5]  H. E. Bennett,et al.  Accurate method for determining photometric linearity. , 1966, Applied optics.

[6]  B H Kolner Intensity noise enhancement in the half-wave plate/polarizer attenuator. , 1991, Optics letters.

[7]  Lorne A. Whitehead,et al.  Novel fiber-based integrating sphere for luminous flux measurements , 2006 .

[8]  S. V. Tikhomirov,et al.  Linearity of photodetectors in pulsed operation , 1978 .

[9]  E. Ikonen,et al.  High-accuracy spectrometer for measurement of regular spectral transmittance. , 1995, Applied optics.

[10]  K. L. Eckerle,et al.  New reference spectrophotometer. , 1973, Applied optics.

[11]  Gao Xun Research on Femtosecond Pulse Width Measurement by Spectroscopy , 2010 .

[12]  Jaeho Shim,et al.  Fluorescence detection system with miniaturized integrating sphere , 2011, 16th International Conference on Optical MEMS and Nanophotonics.

[13]  C. L. Sanders A Photocell Linearity Tester , 1962 .

[14]  Hongchao Hui,et al.  Characteristic research and linearity measurement of photodetectors , 2014, Photoelectronic Technology Committee Conferences.

[15]  Y. Zhao,et al.  Influence of a half-wave plate–polarizer attenuator on broadband femtosecond laser pulses , 2007 .

[16]  Petri Kärhä,et al.  Measurement of the absolute linearity of photodetectors with a diode laser , 1999 .

[17]  Erkki Ikonen,et al.  High-accuracy measurement of specular spectral reflectance and transmittance , 1999 .

[18]  L. D. Wonnell Integrating sphere photodetector for measurement of continuous-wave and peak laser power , 1976 .

[19]  郑小兵 Zheng Xiaobing,et al.  Linearity Measurement of Accurate Photodetector , 2008 .