Pelvic tilt measurement before and after total hip arthroplasty.

INTRODUCTION Most computer-assisted navigation systems used in total hip arthroplasty (THA) reference the anterior pelvic plane, which connects the anterior superior iliac spines and the pubic symphysis. The pelvic tilt is defined as the angle between this anterior pelvic plane (APP) and a vertical line in the standing position. Important interindividual variations of this angle have been reported and may affect final functional anteversion of the acetabular cup. The preoperative value of the pelvic tilt has been included in computer-assisted navigation systems to improve acetabular cup positioning. However, there is no data available which strongly confirms the consistency of this angle for each individual after hip prosthesis implantation. HYPOTHESIS The orientation of the APP in the standing position is not significantly modified after THA. OBJECTIVES To evaluate in a prospective manner, the reproducibility of pelvic tilt measurement and its variability between THA preoperative and 3-year postoperative measurements. MATERIALS AND METHODS A lateral teleradiograph of the pelvis and dorsolumbar spine was obtained in the standing position preoperatively and 3 years after THA. Fifty patients undergoing THA performed by a single operator via an anterolateral approach (26 males and 24 females) were included prospectively. The pelvic tilt was measured on radiographs by two independent observers. The angle was defined as positive in case of pelvis retroversion relative to the vertical plane and negative in case of anteversion. Bland-Altman analysis was used to assess levels of agreement between both operator measurements while preoperative and last follow-up measurements were compared using the Student t-test for unpaired samples. RESULTS The level of agreement between measurements of both operators was satisfactory. Mean preoperative pelvic tilt was 4.68 degrees +/-0.68 S.D. (-6 degrees to 14 degrees), and 4.78 degrees +/-0.64 S.D. (-5 degrees to 14 degrees) at last follow-up. The mean difference between preoperative and last follow-up measurements was 3 degrees +/-0.3 S.D. There was no statistically significant variation between preoperative and 3-year follow-up values (p>0.05). Ninety-five percent of the patients had less than a 5 degrees difference between both measurements while 5% had a difference ranging from 5 degrees to 10 degrees ; none of the patients reported a variation greater than 10 degrees . DISCUSSION Our findings show no significant variation in pelvic tilt between preoperative and 3-year follow-up values after THA. Therefore, the individual preoperative value of this angle should be integrated to achieve proper acetabular cup placement during THA especially when using computed assisted navigation based on the APP.

[1]  H. Migaud,et al.  [Low accuracy of the anterior pelvic plane to guide the position of the cup with imageless computer assistance: variation of position in 106 patients]. , 2007, Revue de chirurgie orthopedique et reparatrice de l'appareil moteur.

[2]  J. Argenson,et al.  Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. , 2007, The Journal of bone and joint surgery. American volume.

[3]  D. Altman,et al.  A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. , 1990, Computers in biology and medicine.

[4]  Y. Pinoit,et al.  Positionnement de la cupule d’une prothèse totale de hanche par navigation sans image basée sur la cinématique articulaire: Premiers résultats avec évaluation tomodensitométrique de l’orientation des implants , 2006 .

[5]  Karl-Heinz Widmer,et al.  A simplified method to determine acetabular cup anteversion from plain radiographs. , 2004, The Journal of arthroplasty.

[6]  R. Philippota,et al.  Pelvic balance in sagittal and Lewinnek reference planes in the standing, supine and sitting positions , 2009 .

[7]  O. Müller,et al.  [Quantification and visualization of the influence of pelvic tilt upon measurement of acetabular inclination and anteversion]. , 2005, Zeitschrift fur Orthopadie und ihre Grenzgebiete.

[8]  L. Lin,et al.  A concordance correlation coefficient to evaluate reproducibility. , 1989, Biometrics.

[9]  E. Thomasson,et al.  Rôle de la courbure rachidienne sagittale dans la survenue des luxations précoces après reprise de prothèse totale de hanche , 2004 .

[10]  Positionnement actabulaire assist par ordinateur dans les prothses totales de hanche , 2007 .

[11]  J. Fleiss The design and analysis of clinical experiments , 1987 .

[12]  西原 俊作 Measurements of Pelvic Flexion Angle Using Three-Dimensional Computed Tomography , 2003 .

[13]  James B Stiehl,et al.  Comparison of conventional versus computer-navigated acetabular component insertion. , 2007, The Journal of arthroplasty.

[14]  C. Lill,et al.  Implant position calculation for acetabular cup placement considering pelvic lateral tilt and inclination , 2006, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[15]  Y. Pinoit,et al.  Fiabilité limitée du plan pelvien antérieur pour l’implantation assistée par informatique de la cupule d’une prothèse totale de hanche , 2007 .

[16]  T. P. Ng,et al.  Sagittal pelvic mal‐rotation and positioning of the acetabular component in total hip arthroplasty: Three‐dimensional computer model analysis , 2007, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[17]  K. Kaufman,et al.  The 2008 Frank Stinchfield Award: Variation in Postoperative Pelvic Tilt May Confound the Accuracy of Hip Navigation Systems , 2009, Clinical orthopaedics and related research.

[18]  D. Vandevelde,et al.  Reduction in Variability of Acetabular Cup Abduction Using Computer Assisted Surgery: A Prospective and Randomized Study , 2002, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[19]  A. Cotten,et al.  Variations of pelvic anteversion in the lying and standing positions analysis of 24 control subjects and implications for CT measurement of position of a prosthetic cup , 2001, Surgical and Radiologic Anatomy.

[20]  N. Wülker,et al.  Quantifizierung und Visualisierung des Einflusses der Beckenkippung auf die Messung von Pfanneninklination und -anteversion , 2005 .

[21]  M Tingart,et al.  Imageless navigation for insertion of the acetabular component in total hip arthroplasty: is it as accurate as CT-based navigation? , 2006, The Journal of bone and joint surgery. British volume.

[22]  Brigitte M Jolles,et al.  Computer-assisted Cup Placement Techniques in Total Hip Arthroplasty Improve Accuracy of Placement , 2004, Clinical orthopaedics and related research.

[23]  Branislav Jaramaz,et al.  Functional Pelvic Orientation Measured from Lateral Standing and Sitting Radiographs , 2006, Clinical orthopaedics and related research.

[24]  Alon Wolf,et al.  Cup Alignment Error Model for Total Hip Arthroplasty , 2005, Clinical orthopaedics and related research.

[25]  J. Jouve,et al.  Étude de la fiabilité des mesures de l’incidence pelvienne dans l’analyse de l’équilibre sagittal du bassin , 2008 .

[26]  Burkhard Lembeck,et al.  Pelvic tilt makes acetabular cup navigation inaccurate , 2005, Acta orthopaedica.