Interface pressures and shear stresses at thirteen socket sites on two persons with transtibial amputation.

Residual limb/prosthetic socket interface pressures and shear stresses were measured at 13 sites on two subjects with unilateral transtibial amputation (TTA) using total-contact patellar-tendon-bearing prostheses. Maximal interface stresses during stance phase for each of 13 transducer sites were determined, then means for all steps calculated. Maximal pressure and resultant shear stress during stance phase were shown at anterior distal or mid-limb sites and the maxima occurred during the first 50% of stance phase. Anterior medial and lateral proximal sites showed their greatest pressure during the second 50%. At lateral mid-limb and popliteal fossa sites, resultant shear stress directions suggest that soft tissue was displaced toward the socket brim during weight-bearing. Results also suggest that skin across the distal tibial crest was in tension at the times of the first and second peaks in the shank axial force-time curve in all sessions. Significant difference (p < 0.05) in maximal stresses between sessions conducted > 3 weeks apart were apparent for both subjects.

[1]  L A Leavitt,et al.  Gait Analysis and Tissue-Socket Interface Pressures in Above-Knee Amputees , 1972, Southern medical journal.

[2]  Grahn Ec A power unit for functional hand splints. , 1970 .

[3]  W. Robert Harris,et al.  CANADIAN EXPERIENCE WITH THE PATELLAR TENDON BEARING BELOW-KNEE PROSTHESIS , 1962 .

[4]  Koepke Gh,et al.  Effect of liner materials on interface pressures in below-knee prostheses. , 1970 .

[5]  L Bennett,et al.  Stump-socket pressure in lower extremity prostheses. , 1968, Journal of biomechanics.

[6]  H B Skinner,et al.  Analysis of a below-knee patellar tendon-bearing prosthesis: a finite element study. , 1991, Journal of rehabilitation research and development.

[7]  V. C. Roberts,et al.  Development of a non-linear finite element modelling of the below-knee prosthetic socket interface. , 1995, Medical engineering & physics.

[8]  D J Winarski,et al.  Least-squares matrix correlations between stump stresses and prosthesis loads for below-knee amputees. , 1987, Journal of biomechanical engineering.

[9]  J. R. Engsberg,et al.  Quantifying interface pressures in below-knee amputee sockets , 1993 .

[10]  R. Contini,et al.  Tangential pressure measurements in above-knee suction sockets. , 1970, Bulletin of prosthetics research.

[11]  L. Bennett,et al.  Shear vs pressure as causative factors in skin blood flow occlusion. , 1979, Archives of physical medicine and rehabilitation.

[12]  D. Childress,et al.  Normal and shear stresses on a residual limb in a prosthetic socket during ambulation: comparison of finite element results with experimental measurements. , 1993, Journal of rehabilitation research and development.

[13]  J R Pearson,et al.  Pressure variation in the below-knee, patellar tendon bearing suction socket prosthesis. , 1974, Journal of biomechanics.

[14]  K Oberg,et al.  Pressures in critical regions of the below-knee patellar-tendon-bearing prosthesis. , 1973, Bulletin of prosthetics research.

[15]  V T Rhyne,et al.  Quantitative method to measure the relationship between prosthetic gait and the forces produced at the stump-socket interface. , 1970, American journal of physical medicine.

[16]  G H Koepke,et al.  Effect of liner materials on interface pressures in below-knee prostheses. , 1970, Archives of physical medicine and rehabilitation.

[17]  E M Burgess,et al.  Interface shear stresses during ambulation with a below-knee prosthetic limb. , 1992, Journal of rehabilitation research and development.