Evaluation of an active magnetic resonance tracking system for interstitial brachytherapy.

PURPOSE In gynecologic cancers, magnetic resonance (MR) imaging is the modality of choice for visualizing tumors and their surroundings because of superior soft-tissue contrast. Real-time MR guidance of catheter placement in interstitial brachytherapy facilitates target coverage, and would be further improved by providing intraprocedural estimates of dosimetric coverage. A major obstacle to intraprocedural dosimetry is the time needed for catheter trajectory reconstruction. Herein the authors evaluate an active MR tracking (MRTR) system which provides rapid catheter tip localization and trajectory reconstruction. The authors assess the reliability and spatial accuracy of the MRTR system in comparison to standard catheter digitization using magnetic resonance imaging (MRI) and CT. METHODS The MRTR system includes a stylet with microcoils mounted on its shaft, which can be inserted into brachytherapy catheters and tracked by a dedicated MRTR sequence. Catheter tip localization errors of the MRTR system and their dependence on catheter locations and orientation inside the MR scanner were quantified with a water phantom. The distances between the tracked tip positions of the MRTR stylet and the predefined ground-truth tip positions were calculated for measurements performed at seven locations and with nine orientations. To evaluate catheter trajectory reconstruction, fifteen brachytherapy catheters were placed into a gel phantom with an embedded catheter fixation framework, with parallel or crossed paths. The MRTR stylet was then inserted sequentially into each catheter. During the removal of the MRTR stylet from within each catheter, a MRTR measurement was performed at 40 Hz to acquire the instantaneous stylet tip position, resulting in a series of three-dimensional (3D) positions along the catheter's trajectory. A 3D polynomial curve was fit to the tracked positions for each catheter, and equally spaced dwell points were then generated along the curve. High-resolution 3D MRI of the phantom was performed followed by catheter digitization based on the catheter's imaging artifacts. The catheter trajectory error was characterized in terms of the mean distance between corresponding dwell points in MRTR-generated catheter trajectory and MRI-based catheter digitization. The MRTR-based catheter trajectory reconstruction process was also performed on three gynecologic cancer patients, and then compared with catheter digitization based on MRI and CT. RESULTS The catheter tip localization error increased as the MRTR stylet moved further off-center and as the stylet's orientation deviated from the main magnetic field direction. Fifteen catheters' trajectories were reconstructed by MRTR. Compared with MRI-based digitization, the mean 3D error of MRTR-generated trajectories was 1.5 ± 0.5 mm with an in-plane error of 0.7 ± 0.2 mm and a tip error of 1.7 ± 0.5 mm. MRTR resolved ambiguity in catheter assignment due to crossed catheter paths, which is a common problem in image-based catheter digitization. In the patient studies, the MRTR-generated catheter trajectory was consistent with digitization based on both MRI and CT. CONCLUSIONS The MRTR system provides accurate catheter tip localization and trajectory reconstruction in the MR environment. Relative to the image-based methods, it improves the speed, safety, and reliability of the catheter trajectory reconstruction in interstitial brachytherapy. MRTR may enable in-procedural dosimetric evaluation of implant target coverage.

[1]  S. Souza,et al.  Real‐time position monitoring of invasive devices using magnetic resonance , 1993, Magnetic resonance in medicine.

[2]  P Boesiger,et al.  Biopsy needle susceptibility artifacts , 1996, Magnetic resonance in medicine.

[3]  C Kolotas,et al.  Catheter autoreconstruction in computed tomography based brachytherapy treatment planning. , 2000, Medical physics.

[4]  M. Zaider,et al.  Accuracy in catheter reconstruction in computed tomography planning of high dose rate prostate brachytherapy , 2000, Proceedings of the 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Cat. No.00CH37143).

[5]  M. Zaider,et al.  Accuracy in catheter reconstruction in computed tomography planning of high dose rate prostate brachytherapy. , 2000 .

[6]  David M. Doddrell,et al.  Geometric distortion in clinical MRI systems Part II: correction using a 3D phantom. , 2004, Magnetic resonance imaging.

[7]  Deming Wang,et al.  Geometric distortion in clinical MRI systems Part II: correction using a 3D phantom. , 2004, Magnetic resonance imaging.

[8]  Deming Wang,et al.  Geometric distortion in clinical MRI systems Part I: evaluation using a 3D phantom. , 2004, Magnetic resonance imaging.

[9]  Christian Kirisits,et al.  Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (I): concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  Phillip M Devlin,et al.  Magnetic resonance-guided interstitial therapy for vaginal recurrence of endometrial cancer. , 2006, International journal of radiation oncology, biology, physics.

[11]  J. Dimopoulos,et al.  The Vienna applicator for combined intracavitary and interstitial brachytherapy of cervical cancer: clinical feasibility and preliminary results. , 2006, International journal of radiation oncology, biology, physics.

[12]  M. Ranney,et al.  Beyond the bedside: Clinicians as guardians of public health, medicine and science , 2020, The American Journal of Emergency Medicine.

[13]  J. Dimopoulos,et al.  Recommendations from gynaecological (GYN) GEC ESTRO working group (II): concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy-3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. , 2006, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  Kari Tanderup,et al.  Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer , 2008 .

[15]  Marinus A Moerland,et al.  Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of cervical cancer. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[16]  Kari Tanderup,et al.  Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  G. Németh,et al.  Accelerated partial-breast irradiation using high-dose-rate interstitial brachytherapy: 12-year update of a prospective clinical study. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  Christian Kirisits,et al.  Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group: considerations and pitfalls in commissioning and applicator reconstruction in 3D image-based treatment planning of cervix cancer brachytherapy. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[19]  Peter R Seevinck,et al.  Highly localized positive contrast of small paramagnetic objects using 3D center‐out radial sampling with off‐resonance reception , 2011, Magnetic resonance in medicine.

[20]  J. Pouliot,et al.  American Brachytherapy Society consensus guidelines for high-dose-rate prostate brachytherapy. , 2012, Brachytherapy.

[21]  Jan Egger,et al.  3T MR-Guided Brachytherapy for Gynecologic Malignancies , 2012, Magnetic resonance imaging.

[22]  Christian Kirisits,et al.  Uncertainties in image guided adaptive cervix cancer brachytherapy: impact on planning and prescription. , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[23]  Jackie Szymonifka,et al.  A prospective trial of real-time magnetic resonance-guided catheter placement in interstitial gynecologic brachytherapy. , 2013, Brachytherapy.

[24]  C. Kirisits,et al.  Magnetic resonance image guided brachytherapy. , 2014, Seminars in radiation oncology.

[25]  C. Kirisits,et al.  Review of clinical brachytherapy uncertainties: Analysis guidelines of GEC-ESTRO and the AAPM☆ , 2014, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[26]  John H. Lewis,et al.  Dosimetric consequences of interobserver variability in delineating the organs at risk in gynecologic interstitial brachytherapy. , 2014, International journal of radiation oncology, biology, physics.

[27]  Robert A Cormack,et al.  A system to use electromagnetic tracking for the quality assurance of brachytherapy catheter digitization. , 2014, Medical physics.

[28]  Alireza Mehrtash,et al.  Real‐time active MR‐tracking of metallic stylets in MR‐guided radiation therapy , 2015, Magnetic resonance in medicine.

[29]  Sang-June Park,et al.  Early clinical outcomes of ultrasound-guided CT-planned high-dose-rate interstitial brachytherapy for primary locally advanced cervical cancer. , 2015, Brachytherapy.

[30]  Emmanuel Racine,et al.  Fast, automatic, and accurate catheter reconstruction in HDR brachytherapy using an electromagnetic 3D tracking system. , 2014, Medical physics.