Continuous glucose sensing with fluorescent thin-film hydrogels. 2. Fiber optic sensor fabrication and in vitro testing.

BACKGROUND There continues to be a need for better sensors for continuous glucose monitoring. We are working on a two-component sensing system based on a viologen boronic acid and a fluorescent dye, which are immobilized in a hydrogel. This system has the potential for further development into a real-time glucose-monitoring device. The current study reports the fabrication of sensors using preformed hydrogels and the first in vitro monitoring of glucose concentrations in a prototype sensor configuration. METHODS Glucose sensing hydrogels containing a fluorescent dye and viologen boronic acid quencher were preformed in a mold. These preformed hydrogels were then attached to the distal end of a plastic fiber optic cable using different adhesives to prepare the in vitro sensors. These sensors were connected to a flow cell and monitored using a fluorescence spectrometer. The fluorescence emitted by the hydrogel changes depending on the glucose concentration. Hydrogel components were modified in order to optimize the performance of the sensors. RESULTS A soft tissue adhesive used by veterinarians was found to be an effective adhesive for bonding the hydrogel to the fiber tip. This adhesive did not affect the glucose sensing ability of the hydrogels after fabrication. Several sensors were fabricated with varying composition of sensing elements, and all of them showed stable and reversible glucose response. The glucose signal was found to be stable over months on repeated testing. Glucose sensing studies using the sensors with hydrogels containing different compositions of sensing elements showed that the ratio of dye to quencher is an important parameter in determining the magnitude and linearity of glucose response in the biological range. The response time of the sensor was shown to be dependent on the hydrophilicity of the hydrogels. Modifying the hydrogels with ionic comonomers shortened the response time. CONCLUSIONS The combination of the anionic dye 2 and viologen-based boronic acid 1 immobilized in a 2-hydroxyethyl methacrylate hydrogel functions well in a fiber optic configuration. This preliminary study suggests that the two-component sensing system has several advantages in terms of stability and ease of fabrication. Improvement of the configuration of the sensor and further development of the sensor towards application for in vitro study are underway.

[1]  S. Shinkai,et al.  Artificial Receptors as Chemosensors for Carbohydrates , 2010 .

[2]  Ying Ma,et al.  One saccharide sensor based on the complex of the boronic acid and the monosaccharide using electrochemical impedance spectroscopy , 2005 .

[3]  Deqing Zhang,et al.  A new saccharide sensor based on a tetrathiafulvalene-anthracene dyad with a boronic acid group. , 2005, The Journal of organic chemistry.

[4]  D. Cordes,et al.  The interaction of boronic acid-substituted viologens with pyranine: the effects of quencher charge on fluorescence quenching and glucose response. , 2005, Langmuir : the ACS journal of surfaces and colloids.

[5]  H. Tao,et al.  Selective Glucose Sensing Utilizing Complexation with Fluorescent Boronic Acid on Polycation , 2005 .

[6]  D. Cordes,et al.  Evaluation of Pyranine Derivatives in Boronic Acid Based Saccharide Sensing: Significance of Charge Interaction Between Dye and Quencher in Solution and Hydrogel , 2004, Journal of Fluorescence.

[7]  S. Shinkai,et al.  A Study of Boronic Acid Based Fluorescent Glucose Sensors , 2004, Journal of Fluorescence.

[8]  D. Cordes,et al.  Continuous glucose sensing with a fluorescent thin-film hydrogel. , 2003, Angewandte Chemie.

[9]  M. Delverdier,et al.  Tolerance of the rabbit cornea to an n-butyl-ester cyanoacrylate adhesive (Vetbond). , 2001, Veterinary ophthalmology.

[10]  M Schetz,et al.  Intensive insulin therapy in critically ill patients. , 2001, The New England journal of medicine.

[11]  T. James,et al.  Fluorescent internal charge transfer (ICT) saccharide sensor , 2001 .

[12]  L. Heinemann,et al.  Sensors for glucose monitoring: technical and clinical aspects , 2001, Diabetes/metabolism research and reviews.

[13]  G. S. Wilson,et al.  Enzyme-based biosensors for in vivo measurements. , 2000, Chemical reviews.

[14]  G. Grunkemeier,et al.  Glucose control lowers the risk of wound infection in diabetics after open heart operations. , 1997, The Annals of thoracic surgery.

[15]  Anthony W. Czarnik,et al.  Fluorescent chemosensors of carbohydrates. A means of chemically communicating the binding of polyols in water based on chelation-enhanced quenching , 1992 .

[16]  W Kerner,et al.  Implantable glucose sensors: present status and future developments. , 2001, Experimental and clinical endocrinology & diabetes : official journal, German Society of Endocrinology [and] German Diabetes Association.