The performance and evaluation of Abbott Architect Ci8200biochemical immunity conjunctedly ma-chine
暂无分享,去创建一个
To comprehend the fuction of Abbott Laboratories Architect ci8200biochemical immunity conjuctedly machine and evaluate the instrument.Evaluate the precision,accuracy,linear and sample carrying rate of the instrument by adopting RANDOX definite value serum and high density patient serum.The CV value of the same items measured in diferent analytic system is(0.4~0.8)%,the CV of values between deferent days is(0.5~0.85)%.The relative error to RANDOX definite value serum is(0.26~1.20)%.The instument has a good linearity and low sample carrying rate.Abbott Laboratories Architect Ci8200biochemical immunity conjuctedly machine are well designed and has good performance,which is suitable for the measuring the sample in enormous quantities.