Variable Power Combiner for RF Mode Shimming in 7-T MR Imaging

This contribution discusses the utilization of RF power in an MRI system with RF mode shimming which enables the superposition of circularly polarized modes of a transmit RF coil array driven by a Butler matrix. Since the required power for the individual modes can vary widely, mode-shimming can result in a significant underutilization of the total available RF power. A variable power combiner (VPC) is proposed to improve the power utilization: it can be realized as a reconfiguration of the MRI transmit system by the inclusion of one additional matrix network which receives the power from all transmit amplifiers at its input ports and provides any desired (combined) power distribution at its output ports by controlling the phase and amplitude of the amplifiers’ input signals. The power distribution at the output ports of the VPC is then fed into the “mode” ports of the coil array Butler matrix in order to superimpose the spatial modes at the highest achievable power utilization. The VPC configuration is compared to the standard configuration of the transmit chain of our MRI system with 8 transmit channels and 16 coils. In realistic scenarios, improved power utilization was achieved from 17% to 60% and from 14% to 55% for an elliptical phantom and a region of interest in the abdomen, respectively, and an increase of the power utilization of 1 dB for a region of interest in the upper leg. In general, it is found that the VPC allows significant improvement in power utilization when the shimming solution demands only a few modes to be energized, while the technique can yield loss in power utilization in cases with many modes required at high power level.

[1]  M. E. Ladd,et al.  16-Bit Vector Modulator for B 1 Shimming in 7 T MRI , 2008 .

[2]  Jürgen Hennig,et al.  Experimental analysis of parallel excitation using dedicated coil setups and simultaneous RF transmission on multiple channels , 2005, Magnetic resonance in medicine.

[3]  U Klose,et al.  Mapping of the radio frequency magnetic field with a MR snapshot FLASH technique. , 1992, Medical physics.

[4]  Markus Vester,et al.  Degenerate mode band‐pass birdcage coil for accelerated parallel excitation , 2007, Magnetic resonance in medicine.

[5]  Michael B. Smith,et al.  Exploring the limits of RF shimming for high‐field MRI of the human head , 2006, Magnetic resonance in medicine.

[6]  Klaus Solbach,et al.  Microstrip Butler matrix design and realization for 7 T MRI , 2011, Magnetic resonance in medicine.

[7]  T. T. Ha,et al.  Microwave power combining and graceful degradation , 1980 .

[8]  Yudong Zhu,et al.  Parallel excitation with an array of transmit coils , 2004, Magnetic resonance in medicine.

[9]  T. Bolz,et al.  An 8-channel add-on RF shimming system for whole-body 7 Tesla MRI including real-time SAR monitoring , 2008 .

[10]  Mark E Ladd,et al.  RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high‐field MRI , 2010, Magnetic resonance in medicine.

[11]  P. Börnert,et al.  Transmit SENSE , 2003, Magnetic resonance in medicine.

[12]  Wilfried Schnell,et al.  Verstärkereinrichtung für eine Modenantenne , 2006 .

[13]  Oliver Kraff,et al.  Planar butler matrix technology for 7 Tesla MRI , 2009 .

[14]  Steen Moeller,et al.  B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil , 2005, Magnetic resonance in medicine.

[15]  D. Hoult Sensitivity and Power Deposition in a High‐Field Imaging Experiment , 2000, Journal of magnetic resonance imaging : JMRI.

[16]  J. Butler,et al.  Beam-forming matrix simplifies design of electronically scanned antennas , 1961 .