Sonography of the skin at 100 MHz enables in vivo visualization of stratum corneum and viable epidermis in palmar skin and psoriatic plaques.

A main drawback of 20-25 MHz ultrasound units for skin imaging is their limited resolution. We used a transducer with a center frequency of 95 MHz and a resolution of 8.5 microm axially and 27 microm laterally - an almost 10-fold increase compared with 20 MHz. By means of a new scanning technology we reached a depth of field of 3.2 mm. We examined normal palmar skin, normal glabrous skin on the abdomen, the upper back, the calf and the dorsal forearm, and 35 lesions of psoriasis vulgaris. From 11 psoriatic plaques biopsies were taken for correlation with the sonograms. In normal palmar skin, the horny layer is represented as an echopoor band below the skin entry echo, traversed by echorich coils, which correspond to eccrine sweat gland ducts. The thickness of this band significantly increases after occlusive application of petrolatum. Its lower border is defined by an echorich line, representing the stratum corneum/stratum Malpighii-interface. Underneath, a second echopoor band is visible, which corresponds to the viable epidermis plus the papillary dermis, bordered by the scattered echo reflexes of the reticular dermis. This band is also visible in glabrous skin; however, the stratum corneum cannot be detected. In psoriatic lesions, the thickened horny layer appears echorich; after application of petrolatum, its echodensity decreases. Below, the acanthotic epidermis plus the dermis with the inflammatory infiltrate are represented as an echopoor band. There is an excellent correlation between the sonometric thickness of this band and the histometric thickness of the acanthosis plus the infiltrated dermis. Our results show that 100 MHz sonography is a valuable tool for in vivo examination of the upper skin layers.

[1]  S Murakami,et al.  Human skin histology using high‐resolution echography , 1989, Journal of clinical ultrasound : JCU.

[2]  O Jolivet,et al.  In vivo hydration profile in skin layers by high-resolution magnetic resonance imaging. , 1994, Skin pharmacology : the official journal of the Skin Pharmacology Society.

[3]  F. Foster,et al.  The use of high-frequency ultrasound as a method of assessing the severity of a plaque of psoriasis. , 1996, Archives of dermatology.

[4]  S. Seidenari,et al.  B‐scanning evaluation with image analysis of psoriatic skin , 1992, Experimental dermatology.

[5]  L Pourcelot,et al.  ULTRASOUND IMAGING OF PSORIATIC SKIN: A NONINVASIVE TECHNIQUE TO EVALUATE TREATMENT OF PSORIASIS , 1994, International journal of dermatology.

[6]  J. Serup,et al.  Ultrasound for assessment of allergic and irritant patch test reactions , 1987, Contact dermatitis.

[7]  F. Dunn,et al.  Letter: Correlation of echographic visualizability of tissue with biological composition and physiological state. , 1973, The Journal of the Acoustical Society of America.

[8]  T. Carpenter,et al.  Short echo time MRI enables visualisation of the natural state of human stratum corneum water in vivo. , 1996, Magnetic resonance imaging.

[9]  P Altmeyer,et al.  Malignant melanoma in 20-MHz B scan sonography. , 1992, Dermatology.

[10]  L From,et al.  Does High‐Frequency (40–60 MHz) Ultrasound Imaging Play a Role in the Clinical Management of Cutaneous Melanoma? , 1995, Annals of plastic surgery.

[11]  A. Kligman,et al.  CHAPTER XX – The Biology of the Stratum Corneum , 1964 .

[12]  F. Lizzi,et al.  Diagnostic ultrasound in dermatology: current uses and future potential. , 1994, Cutis.

[13]  S. Seidenari,et al.  High-frequency sonography combined with image analysis: a noninvasive objective method for skin evaluation and description. , 1995, Clinics in dermatology.

[14]  J. Jones,et al.  Scanning acoustic microscopy of neoplastic and inflammatory cutaneous tissue specimens. , 1991, The Journal of investigative dermatology.

[15]  P. Altmeyer,et al.  20 MHz sonography, colorimetry and image analysis in the evaluation of psoriasis vulgaris. , 1995, Journal of dermatological science.

[16]  S. R. Schwartz,et al.  Assessment of epithelial thickness by ultrasonic imaging. , 1991, Decubitus.

[17]  F. Lizzi,et al.  DIAGNOSTIC HIGH‐RESOLUTION ULTRASOUND IN DERMATOLOGY , 1993, International journal of dermatology.

[18]  J Serup,et al.  Ultrasound structure and digital image analysis of the subepidermal low echogenic band in aged human skin: diurnal changes and interindividual variability. , 1994, The Journal of investigative dermatology.

[19]  H Ermert,et al.  Psoriasis vulgaris in 50 MHz B-scan ultrasound--characteristic features of stratum corneum, epidermis and dermis. , 1994, Acta dermato-venereologica. Supplementum.

[20]  L From,et al.  A 40-100 MHz B-scan ultrasound backscatter microscope for skin imaging. , 1995, Ultrasound in medicine & biology.

[21]  Thomas Dirschka,et al.  Twenty MHz B‐scan sonography for visualization and skin thickness measurement of human skin , 1994 .

[22]  S. Seidenari,et al.  B scanning evaluation of allergic reactions with binary transformation and image analysis. , 1992, Acta dermato-venereologica. Supplementum.

[23]  R H Guy,et al.  Homogeneous transport in a heterogeneous membrane: water diffusion across human stratum corneum in vivo. , 1996, Biophysical journal.

[24]  B. Fornage,et al.  Imaging of the skin with 20-MHz US. , 1993, Radiology.

[25]  C W Hanke,et al.  High Frequency Ultrasonic Imaging of Skin: Experimental Results , 1984, Ultrasonic imaging.

[26]  J C Bamber,et al.  High frequency, high resolution B‐scan ultrasound in the assessment of skin tumours , 1993, The British journal of dermatology.

[27]  K. Harasiewicz,et al.  High Frequency 40‐MHz Ultrasound A Possible Noninvasive Method for the Assessment of the Boundary of Basal Cell Carcinomas , 1996, Dermatologic surgery : official publication for American Society for Dermatologic Surgery [et al.].