Design and validation of a high-order weighted-frequency fourier linear combiner-based Kalman filter for parkinsonian tremor estimation

The design of a tremor estimator is an important part of designing mechanical tremor suppression orthoses. A number of tremor estimators have been developed and applied with the assumption that tremor is a mono-frequency signal. However, recent experimental studies have shown that Parkinsonian tremor consists of multiple frequencies, and that the second and third harmonics make a large contribution to the tremor. Thus, the current estimators may have limited performance on estimation of the tremor harmonics. In this paper, a high-order tremor estimation algorithm is proposed and compared with its lower-order counterpart and a widely used estimator, the Weighted-frequency Fourier Linear Combiner (WFLC), using 18 Parkinsonian tremor data sets. The results show that the proposed estimator has better performance than its lower-order counterpart and the WFLC. The percentage estimation accuracy of the proposed estimator is 85±2.9%, an average improvement of 13% over the lower-order counterpart. The proposed algorithm holds promise for use in wearable tremor suppression devices.

[1]  A S Arnold,et al.  Design of a controlled-energy-dissipation orthosis (CEDO) for functional suppression of intention tremors. , 1995, Journal of rehabilitation research and development.

[2]  Nitish V. Thakor,et al.  Adaptive Fourier modeling for quantification of tremor 1 Funding provided by National Institute on Disability and Rehabilitation Research (grant number H133G30064). 1 , 1997, Journal of Neuroscience Methods.

[3]  Rajnikant V. Patel,et al.  An active handheld device for compensation of physiological tremor using an ionic polymer metallic composite actuator , 2013, 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[4]  W. Marsden I and J , 2012 .

[5]  Philippe Poignet,et al.  Online pathological tremor characterization using extended Kalman filtering , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[6]  E. Rocon,et al.  Design and Validation of a Rehabilitation Robotic Exoskeleton for Tremor Assessment and Suppression , 2007, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[7]  D. Brooks,et al.  A five-year study of the incidence of dyskinesia in patients with early Parkinson's disease who were treated with ropinirole or levodopa. , 2000, The New England journal of medicine.

[8]  José Luis Pons Rovira,et al.  Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data , 2010, Sensors.

[9]  Win Tun Latt,et al.  Kalman filtering of accelerometer and electromyography (EMG) data in pathological tremor sensing system , 2008, 2008 IEEE International Conference on Robotics and Automation.

[10]  Philippe Poignet,et al.  Pathological Tremor and Voluntary Motion Modeling and Online Estimation for Active Compensation , 2011, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[11]  M. Brin,et al.  Consensus Statement of the Movement Disorder Society on Tremor , 2008, Movement disorders : official journal of the Movement Disorder Society.

[12]  Ana Luisa Trejos,et al.  The measurement and analysis of Parkinsonian hand tremor , 2016, 2016 IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI).

[13]  龚启勇,et al.  Task force urges CT lung cancer screening for people at high risk , 2014 .

[14]  Behzad Taheri,et al.  Theoretical Development and Experimental Validation of an Adaptive Controller for Tremor Suppression at Musculoskeletal Level , 2013 .

[15]  Suhail Kazi,et al.  Experimental implementation of smart glove incorporating piezoelectric actuator for hand tremor control , 2010 .

[16]  J.L. Pons,et al.  Upper limb tremor suppression in ADL via an orthosis incorporating a controllable double viscous beam actuator , 2005, 9th International Conference on Rehabilitation Robotics, 2005. ICORR 2005..

[17]  A. Prochazka,et al.  Attenuation of pathological tremors by functional electrical stimulation I: Method , 2006, Annals of Biomedical Engineering.

[18]  Behzad Taheri Real-time pathological tremor identification and suppression in human arm via active orthotic devices , 2013 .

[19]  N.V. Thakor,et al.  Adaptive cancelling of physiological tremor for improved precision in microsurgery , 1998, IEEE Transactions on Biomedical Engineering.

[20]  M. Hallett Tremor , 1991, Neurology.

[21]  M J Rosen,et al.  Evaluating manual control devices for those with tremor disability. , 1987, Journal of rehabilitation research and development.