In the paper the performances of the Virtual Instruments under conducted electromagnetic disturbances are investigated, in order to find out if and how their metrological characteristics change with respect to the ones measured in absence of electromagnetic threats. The analysis is carried out by means of experimental tests, subjecting the AC power port of the systems under test to the conducted disturbances considered by the IEC-61326 standard and following the test procedures prescribed in the IEC-61000-4-series standards. The results show that the tested measurements instruments are sufficiently immune to these disturbances, since no variations of the distinctive parameters have been observed. However particular attention is necessary when the conducted disturbances are present, given that the emissions radiated by the supply cables themselves can alter the instrument performances.
[1]
J. Roztocil,et al.
Influence of magnetic field on measurements using A/D plug-in boards
,
1997,
IEEE Instrumentation and Measurement Technology Conference Sensing, Processing, Networking. IMTC Proceedings.
[2]
G. Betta,et al.
EMC characterization of PC-based data acquisition systems
,
2002,
Industrial Electronics, 2002. ISIE 2002. Proceedings of the 2002 IEEE International Symposium on.
[3]
Carmine Landi,et al.
A COMPARATIVE ANALYSIS IN TERMS OF CONDUCTED SUSCEPTIBILITY OF PC-BASED DATA ACQUISITION SYSTEMS
,
2003
.