A time domain analysis of a calibration error model

Two calibration approaches are examined to compare the error coefficients in signal flow models in the time domain. Ideally, the same error-coefficients should be obtained from both calibration approaches. For the most part, the calculated error coefficients are essentially the same. The only major difference occurred in the source-match term near the time t=0 point. A potential reason for this discrepancy may be attributed to the assumption that the offset distances are exactly a ratio of 2:1, as required by the MMAVERIC calibration technique used.<<ETX>>