Numerical simulation of thrombus aspiration in two realistic models of catheter tips.

Thrombus aspiration catheters are devices used to remove a blood clot from a vessel, usually prior to angioplasty or stent implantation. However, in vitro results showed that the use of different commercial devices could produce very different thrombus removals, suggesting a primary dependence on the distal tip configuration of the catheter. A computational methodology based on realistic catheter tip modeling was developed to investigate the factors affecting the thrombus suction. Two different designs were considered, either with a single central lumen or a combination of central and side holes. First, steady-state aspiration of distilled water from a reservoir was simulated and compared with experimental tests. Subsequently, the aspiration of a totally occlusive thrombus, modeled as a high viscous fluid, was simulated solving a complex two-phase (blood and thrombus) problem. In particular, the benefit of additional openings was investigated. Good matching between the steady-state experimental and numerically simulated hydraulic behaviors allowed a validation of the numerical models. Numerical results of thrombus aspiration showed that the catheter with central and side holes had a worse performance if compared with the single central lumen catheter. Indeed, the inlets in contact with both blood and thrombus preferentially aspirate blood due to its much lower viscosity. This effect hindered the aspiration of thrombus. The amount of aspirated thrombus highly depends on the complex, two-phase fluid dynamics occurring across the catheter tips. Results suggested that location of additional holes is crucial in the catheter aspiration performance.

[1]  R. Puniyani,et al.  Thromboviscometry as a tool for evaluation of thrombotic risk in systemic hypertension , 2000, Journal of Human Hypertension.

[2]  K. Sugi,et al.  Comparison of the in vitro performance of 6 and 7 French aspiration catheters. , 2007, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[3]  J. Ottervanger,et al.  Incidence and clinical significance of distal embolization during primary angioplasty for acute myocardial infarction. , 2002, European heart journal.

[4]  B. Min,et al.  A numerical study on the effect of side hole number and arrangement in venous cannulae. , 2007, Journal of biomechanics.

[5]  E. Romagnoli,et al.  Manual thrombus-aspiration improves myocardial reperfusion: the randomized evaluation of the effect of mechanical reduction of distal embolization by thrombus-aspiration in primary and rescue angioplasty (REMEDIA) trial. , 2005, Journal of the American College of Cardiology.

[6]  M. Kern,et al.  Use of a new thrombus extraction catheter (the Pronto) in the treatment of acute myocardial infarction. , 2005, Journal of interventional cardiology.

[7]  J. Roelandt,et al.  Left atrial enlargement: an electrocardiographic misnomer? An electrocardiographic-echocardiographic study. , 1986, European heart journal.

[8]  J. Riley,et al.  In vitro comparison of cavoatrial (dual stage) cannulae for use during cardiopulmonary bypass , 1986 .

[9]  F. Grover,et al.  Comparison of flow differences among venous cannulas. , 1983, The Annals of thoracic surgery.

[10]  A. M. Leone,et al.  Feasibility of sequential thrombus aspiration and filter distal protection in the management of very high thrombus burden lesions. , 2007, The Journal of invasive cardiology.

[11]  P. Verdonck,et al.  Particle Image Velocimetry–Validated, Computational Fluid Dynamics–Based Design to Reduce Shear Stress and Residence Time in Central Venous Hemodialysis Catheters , 2007, ASAIO journal.

[12]  R. Bartlett,et al.  A Standardized System for Describing Flow/Pressure Relationships in Vascular Access Devices , 1991, ASAIO transactions.

[13]  N. Kaneko,et al.  Comparison of the effects of a distal embolic protection device and an aspiration catheter during percutaneous coronary intervention in patients with acute myocardial infarction. , 2005, Circulation journal : official journal of the Japanese Circulation Society.

[14]  P. Verdonck,et al.  Computational fluid dynamics-analysis of the Niagara hemodialysis catheter in a right heart model. , 2004, Artificial organs.

[15]  D. Sorysz,et al.  Percutaneous thrombectomy with the RESCUE system in acute myocardial infarction. , 2004, Kardiologia polska.

[16]  Yuji Ikari,et al.  Improved performance of a new thrombus aspiration catheter: Outcomes from in vitro experiments and a case presentation , 2004, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[17]  P. Colombo,et al.  CLINICAL RESEARCH Interventional Cardiology Thrombus Aspiration Before Primary Angioplasty Improves Myocardial Reperfusion in Acute Myocardial Infarction The DEAR-MI (Dethrombosis to Enhance Acute Reperfusion in Myocardial Infarction) Study , 2006 .

[18]  Kam-tim Chan,et al.  Thrombectomy with Rescue Aspiration Catheter and Distal Protection with EPI FilterWire in Primary Angioplasty for Acute Myocardial Infarction , 2003 .

[19]  G. Finet,et al.  Large tube section is the key to successful coronary thrombus aspiration: Findings of a standardized bench test , 2006, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[20]  D. Dudek,et al.  The role of thrombectomy and embolic protection devices , 2005 .

[21]  M Grigioni,et al.  Computational model of the fluid dynamics of a cannula inserted in a vessel: incidence of the presence of side holes in blood flow. , 2002, Journal of biomechanics.