Interference Reduction in Glucose Detection by Redox Potential Tuning: New Glucose Meter Development

A new glucose meter was developed employing a novel disposable glucose sensor strip comprising a nicotinamide adenine dinucleotide-glucose dehydrogenase (NAD-GDH) and a mixture of Fe compounds as a mediator. An iron complex, 5-(2,5-di(thiophen-2-yl)-1H-pyrrol-1-yl)-1,10-phenanthroline iron(III) chloride (Fe-PhenTPy), was synthesized as a new mediator for the NAD-GDH system. Due to the high oxidation potential of the mediator, the detection potential was tuned to be more closely fitted toward the enzyme reaction potential, less than 400 mV (vs. Ag/AgCl), by mixing with an additional iron mediator. The impedance spectrometry for the enzyme sensor containing the mixed mediators showed an enhanced charge transfer property. In addition, a new cartridge-type glucose meter was manufactured using effective aligned-electrodes, which showed an enhanced response compared with conventional electrode alignment. The proposed glucose sensor resulted in a wide dynamic range in the concentration range of 30 - 500 mg dL(-1) with a reduced interference effect and a good sensitivity of 0.57 μA mM(-1).

[1]  Lo Gorton,et al.  Electrocatalytic oxidation of NAD(P) H at mediator-modified electrodes. , 2002, Journal of biotechnology.

[2]  Itamar Willner,et al.  Reconstitution of apo-glucose dehydrogenase on pyrroloquinoline quinone-functionalized au nanoparticles yields an electrically contacted biocatalyst. , 2005, Journal of the American Chemical Society.

[3]  N. Evans,et al.  Fluorescence-based glucose sensors. , 2005, Biosensors & bioelectronics.

[4]  Joseph Wang Electrochemical glucose biosensors. , 2008, Chemical reviews.

[5]  Nigel J. Forrow,et al.  Development of a commercial amperometric biosensor electrode for the ketone D-3-hydroxybutyrate. , 2005, Biosensors & bioelectronics.

[6]  D. Cox,et al.  Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood Glucose , 1987, Diabetes Care.

[7]  J. Ellison,et al.  Review of Adverse Events Associated With False Glucose Readings Measured by GDH-PQQ–Based Glucose Test Strips in the Presence of Interfering Sugars , 2010, Diabetes Care.

[8]  Koji Sode,et al.  Direct electron transfer type disposable sensor strip for glucose sensing employing an engineered FAD glucose dehydrogenase. , 2013, Enzyme and microbial technology.

[9]  Y. Shim,et al.  Electron-transfer mediator for a NAD-glucose dehydrogenase-based glucose sensor. , 2013, Analytical chemistry.

[10]  M. Brownlee Biochemistry and molecular cell biology of diabetic complications , 2001, Nature.

[11]  Jaehwan Kim,et al.  Conductometric glucose biosensor made with cellulose and tin oxide hybrid nanocomposite , 2011 .