Implementation of the Structural SIMilarity (SSIM) Index as a Quantitative Evaluation Tool for Dose Distribution Error Detection.

PURPOSE To apply an imaging metric of the Structural SIMilarity (SSIM) index to the radiotherapy dose verification field and evaluate its capability to reveal the different types of errors between two dose distributions. METHOD The SSIM index consists of three sub-indices: luminance, contrast, and structure. Given two images, luminance analysis compares the local mean result, contrast analysis compares the local standard deviation, and the structure index represents the local Pearson correlation. Three test error patterns (absolute dose error, dose gradient error, and dose structure error) were designed to characterize the response of SSIM and its sub-indices and establish the correlation between the indices and different dose error types. After establishing the correlation, four radiotherapy plans (one MLC picket fence test plan, one brain stereotactic radiotherapy plan, and two head-and-neck plans) were tested by computing each index and compared with the gamma analysis results to determine their similarities and differences. RESULTS Among the three test error patterns, the luminance index decreased from 1 to 0.1 when the absolute dose agreement fell from 100% to 5%, the contrast index decreased from 1 to 0.36 when the dose gradient agreement fell from 100% to 10%, and the structure index decreased from 1 to 0.23 when the periodical dose pattern shifted (leading to a lower correlation). Thus, the luminance, contrast and structure index can detect the absolute dose error, gradient discrepancy, and dose structure error, respectively. For the four clinical cases, the sub-indices can reveal the type of error when gamma analysis only provided limited information. CONCLUSIONS The correlation between the subcomponents of the SSIM index and the error types of the dose distribution was established. The SSIM index provides additional error information than that provided by gamma analysis.

[1]  Jean M. Moran,et al.  Verification of dynamic and segmental IMRT delivery by dynamic log file analysis , 2002, Journal of applied clinical medical physics.

[2]  T. K. Yeung,et al.  Quality assurance in radiotherapy: evaluation of errors and incidents recorded over a 10 year period. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  Tomas Kron,et al.  Sensitivity of Electronic Portal Imaging Device (EPID) Based Transit Dosimetry to Detect Inter-fraction Patient Variations , 2019 .

[4]  James C. H. Chu,et al.  The sensitivity of ArcCHECK‐based gamma analysis to manufactured errors in helical tomotherapy radiation delivery , 2015, Journal of applied clinical medical physics.

[5]  Benjamin E Nelms,et al.  Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. , 2011, Medical physics.

[6]  Xiance Jin,et al.  Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification. , 2015, The British journal of radiology.

[7]  J. Cygler,et al.  Commissioning and quality assurance of treatment planning computers. , 1993, International journal of radiation oncology, biology, physics.

[8]  Jonathan G. Li,et al.  Comparison of two commercial detector arrays for IMRT quality assurance , 2009, Journal of Applied Clinical Medical Physics.

[9]  Benjamin E Nelms,et al.  Evaluating IMRT and VMAT dose accuracy: practical examples of failure to detect systematic errors when applying a commonly used metric and action levels. , 2013, Medical physics.

[10]  Patrick A Kupelian,et al.  Characterization and use of EBT radiochromic film for IMRT dose verification. , 2006, Medical physics.

[11]  Verónica Alba-Escorihuela,et al.  On the new metrics for IMRT QA verification. , 2016, Medical physics.

[12]  Eero P. Simoncelli,et al.  Image quality assessment: from error visibility to structural similarity , 2004, IEEE Transactions on Image Processing.

[13]  E. Schreibmann,et al.  Patient-specific quality assurance method for VMAT treatment delivery. , 2009, Medical physics.

[14]  Jatinder R Palta,et al.  Quality assurance of intensity-modulated radiation therapy. , 2008, International journal of radiation oncology, biology, physics.

[15]  Jan-Jakob Sonke,et al.  3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[16]  Developing new extension of GafChromic RTQA2 film to patient quality assurance field using a plan-based calibration method. , 2015, Physics in medicine and biology.

[17]  Jon J Kruse,et al.  On the insensitivity of single field planar dosimetry to IMRT inaccuracies. , 2010, Medical physics.

[18]  Jiazhou Wang,et al.  A study on investigating the delivery parameter error effect on the variation of patient quality assurance during RapidArc treatment. , 2013, Medical physics.

[19]  Jean-Pierre Bissonnette,et al.  Error in the delivery of radiation therapy: results of a quality assurance review. , 2005, International journal of radiation oncology, biology, physics.

[20]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[21]  Sara Bresciani,et al.  Pretreatment patient-specific IMRT quality assurance: a correlation study between gamma index and patient clinical dose volume histogram. , 2012, Medical physics.

[22]  Vladimir Feygelman,et al.  Real-world examples of sensitivity failures of the 3%/3mm pass rate metric and published action levels when used in IMRT/VMAT system commissioning , 2013 .