Spatial Distortion in MRI-Guided Stereotactic Procedures: Evaluation in 1.5-, 3- and 7-Tesla MRI Scanners

Background: Magnetic resonance imaging (MRI) is replacing computed tomography (CT) as the main imaging modality for stereotactic transformations. MRI is prone to spatial distortion artifacts, which can lead to inaccuracy in stereotactic procedures. Objective: Modern MRI systems provide distortion correction algorithms that may ameliorate this problem. This study investigates the different options of distortion correction using standard 1.5-, 3- and 7-tesla MRI scanners. Methods: A phantom was mounted on a stereotactic frame. One CT scan and three MRI scans were performed. At all three field strengths, two 3-dimensional sequences, volumetric interpolated breath-hold examination (VIBE) and magnetization-prepared rapid acquisition with gradient echo, were acquired, and automatic distortion correction was performed. Global stereotactic transformation of all 13 datasets was performed and two stereotactic planning workflows (MRI only vs. CT/MR image fusion) were subsequently analysed. Results: Distortion correction on the 1.5- and 3-tesla scanners caused a considerable reduction in positional error. The effect was more pronounced when using the VIBE sequences. By using co-registration (CT/MR image fusion), even a lower positional error could be obtained. In ultra-high-field (7 T) MR imaging, distortion correction introduced even higher errors. However, the accuracy of non-corrected 7-tesla sequences was comparable to CT/MR image fusion 3-tesla imaging. Conclusion: MRI distortion correction algorithms can reduce positional errors by up to 60%. For stereotactic applications of utmost precision, we recommend a co-registration to an additional CT dataset.

[1]  J. Michael Fitzpatrick,et al.  A technique for accurate magnetic resonance imaging in the presence of field inhomogeneities , 1992, IEEE Trans. Medical Imaging.

[2]  P F Hemler,et al.  Method for correcting magnetic resonance image distortion for frame-based stereotactic surgery, with preliminary results. , 1995, Journal of image guided surgery.

[3]  Yuta Shibamoto,et al.  Validation of accuracy in image co-registration with computed tomography and magnetic resonance imaging in Gamma Knife radiosurgery , 2014, Journal of radiation research.

[4]  Oliver Kraff,et al.  MRI at 7 tesla and above: Demonstrated and potential capabilities , 2015, Journal of magnetic resonance imaging : JMRI.

[5]  G. Glover,et al.  Characterization of spatial distortion in magnetic resonance imaging and its implications for stereotactic surgery. , 1994, Neurosurgery.

[6]  Gary H. Glover,et al.  MR susceptibility misregistration correction , 1993, IEEE Trans. Medical Imaging.

[7]  Essa Yacoub,et al.  Feasibility of Using Ultra-High Field (7 T) MRI for Clinical Surgical Targeting , 2012, PloS one.

[8]  Zbigniew Petrovich,et al.  An image fusion study of the geometric accuracy of magnetic resonance imaging with the Leksell stereotactic localization system1 , 2001, Journal of applied clinical medical physics.

[9]  J. Fitzpatrick,et al.  A universal method for geometric correction of magnetic resonance images for stereotactic neurosurgery. , 1996, Stereotactic and functional neurosurgery.

[10]  Deming Wang,et al.  Geometric distortion in clinical MRI systems Part II: correction using a 3D phantom. , 2004, Magnetic resonance imaging.

[11]  M. Revenu,et al.  MRI geometric distortion: A simple approach to correcting the effects of non‐linear gradient fields , 1999, Journal of magnetic resonance imaging : JMRI.

[12]  J. Vymazal,et al.  Does new magnetic resonance imaging technology provide better geometrical accuracy during stereotactic imaging? , 2005, Journal of neurosurgery.

[13]  E Kanal,et al.  A comparison between magnetic resonance imaging and computed tomography for stereotactic coordinate determination. , 1992, Neurosurgery.

[14]  Christian P Karger,et al.  Accuracy of device-specific 2D and 3D image distortion correction algorithms for magnetic resonance imaging of the head provided by a manufacturer , 2006, Physics in medicine and biology.

[15]  M van Herk,et al.  Magnetic resonance image-directed stereotactic neurosurgery: use of image fusion with computerized tomography to enhance spatial accuracy. , 1995, Journal of neurosurgery.

[16]  M. Ladd,et al.  Evaluation of hardware-related geometrical distortion in structural MRI at 7 Tesla for image-guided applications in neurosurgery. , 2011, Academic radiology.

[17]  C. Yu,et al.  A Phantom Study of the Geometric Accuracy of Computed Tomographic and Magnetic Resonance Imaging Stereotactic Localization with the Leksell Stereotactic System , 2001, Neurosurgery.

[18]  A. Lozano,et al.  Direct visualization of deep brain stimulation targets in Parkinson disease with the use of 7-tesla magnetic resonance imaging. , 2010, Journal of neurosurgery.

[19]  Deming Wang,et al.  Geometric distortion in clinical MRI systems Part I: evaluation using a 3D phantom. , 2004, Magnetic resonance imaging.

[20]  B. Gerbi,et al.  Geometrical accuracy of a 3-tesla magnetic resonance imaging unit in Gamma Knife surgery. , 2006, Journal of neurosurgery.