A Least Mean-Square Filter for the Estimation of the Cardiopulmonary Resuscitation Artifact Based on the Frequency of the Compressions

Cardiopulmonary resuscitation (CPR) artifacts caused by chest compressions and ventilations interfere with the rhythm diagnosis of automated external defibrillators (AED). CPR must be interrupted for a reliable diagnosis. However, pauses in chest compressions compromise the defibrillation success rate and reduce perfusion of vital organs. The removal of the CPR artifacts would enable compressions to continue during AED rhythm analysis, thereby increasing the likelihood of resuscitation success. We have estimated the CPR artifact using only the frequency of the compressions as additional information to model it. Our model of the artifact is adaptively estimated using a least mean-square (LMS) filter. It was tested on 89 shockable and 292 nonshockable ECG samples from real out-of-hospital sudden cardiac arrest episodes. We evaluated the results using the shock advice algorithm of a commercial AED. The sensitivity and specificity were above 95% and 85%, respectively, for a wide range of working conditions of the LMS filter. Our results show that the CPR artifact can be accurately modeled using only the frequency of the compressions. These can be easily registered after small changes in the hardware of the CPR compression pads.

[1]  S. O. Aase,et al.  Feasibility of shock advice analysis during CPR through removal of CPR artefacts from the human ECG. , 2004, Resuscitation.

[2]  Trygve Eftestøl,et al.  Removal of cardiopulmonary resuscitation artifacts from human ECG using an efficient matching pursuit-like algorithm , 2002, IEEE Transactions on Biomedical Engineering.

[3]  Jo Kramer-Johansen,et al.  Quality of out-of-hospital cardiopulmonary resuscitation with real time automated feedback: a prospective interventional study. , 2006, Resuscitation.

[4]  A. Agresti,et al.  Approximate is Better than “Exact” for Interval Estimation of Binomial Proportions , 1998 .

[5]  Jo Kramer-Johansen,et al.  Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest. , 2006, Resuscitation.

[6]  B. Widrow,et al.  Adaptive noise cancelling: Principles and applications , 1975 .

[7]  Lark,et al.  OUTCOMES OF RAPID DEFIBRILLATION BY SECURITY OFFICERS AFTER CARDIAC ARREST IN CASINOS , 2000 .

[8]  Yoshiaki Tadokoro,et al.  LMS-based notch filter for the estimation of sinusoidal signals in noise , 1995, Signal Process..

[9]  Sven Ole Aase,et al.  Compression depth estimation for CPR quality assessment using DSP on accelerometer signals , 2002, IEEE Transactions on Biomedical Engineering.

[10]  Klaus Rheinberger,et al.  Removal of CPR Artifacts From the Ventricular Fibrillation ECG by Adaptive Regression on Lagged Reference Signals , 2008, IEEE Transactions on Biomedical Engineering.

[11]  W. Dick Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care , 2000, Notfall & Rettungsmedizin.

[12]  R W Koster,et al.  Out-of-hospital cardiac arrests in Amsterdam and its surrounding areas: results from the Amsterdam resuscitation study (ARREST) in 'Utstein' style. , 1998, Resuscitation.

[13]  Jo Kramer-Johansen,et al.  Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest , 2005 .

[14]  Greg Sinibaldi,et al.  Incidence of EMS-treated out-of-hospital cardiac arrest in the United States. , 2004, Resuscitation.

[15]  B J Biggerstaff,et al.  Comparing diagnostic tests: a simple graphic using likelihood ratios. , 2000, Statistics in medicine.

[16]  Jesus Ruiz,et al.  Detection of ventricular fibrillation in the presence of cardiopulmonary resuscitation artefacts. , 2007, Resuscitation.

[17]  P. Steen,et al.  Reducing CPR artefacts in ventricular fibrillation in vitro. , 2001, Resuscitation.

[18]  A Stefanovska,et al.  Reversible transitions between synchronization states of the cardiorespiratory system. , 2000, Physical review letters.

[19]  G. Perkins,et al.  European Resuscitation Council guidelines for resuscitation 2005. Section 2. Adult basic life support and use of automated external defibrillators. , 2005, Resuscitation.

[20]  P. Steen,et al.  Effects of Interrupting Precordial Compressions on the Calculated Probability of Defibrillation Success During Out-of-Hospital Cardiac Arrest , 2002, Circulation.

[21]  TingYu,et al.  Adverse Outcomes of Interrupted Precordial Compression During Automated Defibrillation , 2002 .

[22]  J. Glover Adaptive noise canceling applied to sinusoidal interferences , 1977 .

[23]  Jesus Ruiz,et al.  A method to remove CPR artefacts from human ECG using only the recorded ECG. , 2008, Resuscitation.

[24]  K. Lindner,et al.  Spectral analysis of ventricular fibrillation and closed-chest cardiopulmonary resuscitation. , 1996, Resuscitation.

[25]  Maaret Castrén,et al.  European Resuscitation Council Guidelines for Resuscitation 2010 Section 2. Adult basic life support and use of automated external defibrillators , 2010, Resuscitation.

[26]  David E. Snyder,et al.  Wide variation in cardiopulmonary resuscitation interruption intervals among commercially available automated external defibrillators may affect survival despite high defibrillation efficacy , 2004, Critical care medicine.

[27]  Trygve Eftestøl,et al.  CPR artifact removal from human ECG using optimal multichannel filtering , 2000, IEEE Transactions on Biomedical Engineering.

[28]  Robert A. Berg,et al.  Adverse Hemodynamic Effects of Interrupting Chest Compressions for Rescue Breathing During Cardiopulmonary Resuscitation for Ventricular Fibrillation Cardiac Arrest , 2001, Circulation.

[29]  B. Abella,et al.  Uniform reporting of measured quality of cardiopulmonary resuscitation (CPR). , 2007, Resuscitation.