Ultrasoft microwire neural electrodes improve chronic tissue integration.

Chronically implanted neural multi-electrode arrays (MEA) are an essential technology for recording electrical signals from neurons and/or modulating neural activity through stimulation. However, current MEAs, regardless of the type, elicit an inflammatory response that ultimately leads to device failure. Traditionally, rigid materials like tungsten and silicon have been employed to interface with the relatively soft neural tissue. The large stiffness mismatch is thought to exacerbate the inflammatory response. In order to minimize the disparity between the device and the brain, we fabricated novel ultrasoft electrodes consisting of elastomers and conducting polymers with mechanical properties much more similar to those of brain tissue than previous neural implants. In this study, these ultrasoft microelectrodes were inserted and released using a stainless steel shuttle with polyethyleneglycol (PEG) glue. The implanted microwires showed functionality in acute neural stimulation. When implanted for 1 or 8weeks, the novel soft implants demonstrated significantly reduced inflammatory tissue response at week 8 compared to tungsten wires of similar dimension and surface chemistry. Furthermore, a higher degree of cell body distortion was found next to the tungsten implants compared to the polymer implants. Our results support the use of these novel ultrasoft electrodes for long term neural implants. STATEMENT OF SIGNIFICANCE One critical challenge to the translation of neural recording/stimulation electrode technology to clinically viable devices for brain computer interface (BCI) or deep brain stimulation (DBS) applications is the chronic degradation of device performance due to the inflammatory tissue reaction. While many hypothesize that soft and flexible devices elicit reduced inflammatory tissue responses, there has yet to be a rigorous comparison between soft and stiff implants. We have developed an ultra-soft microelectrode with Young's modulus lower than 1MPa, closely mimicking the brain tissue modulus. Here, we present a rigorous histological comparison of this novel ultrasoft electrode and conventional stiff electrode with the same size, shape and surface chemistry, implanted in rat brains for 1-week and 8-weeks. Significant improvement was observed for ultrasoft electrodes, including inflammatory tissue reaction, electrode-tissue integration as well as mechanical disturbance to nearby neurons. A full spectrum of new techniques were developed in this study, from insertion shuttle to in situ sectioning of the microelectrode to automated cell shape analysis, all of which should contribute new methods to the field. Finally, we showed the electrical functionality of the ultrasoft electrode, demonstrating the potential of flexible neural implant devices for future research and clinical use.

[1]  Miguel A. L. Nicolelis,et al.  Actions from thoughts , 2001, Nature.

[2]  Walter Voit,et al.  Thiol-click chemistries for responsive neural interfaces. , 2013, Macromolecular bioscience.

[3]  M. Ward,et al.  Toward a comparison of microelectrodes for acute and chronic recordings , 2009, Brain Research.

[4]  J. Neary,et al.  Activation of Extracellular Signal-Regulated Kinase by Stretch-Induced Injury in Astrocytes Involves Extracellular ATP and P2 Purinergic Receptors , 2003, The Journal of Neuroscience.

[5]  Kasey Catt,et al.  Evaluation of poly(3,4-ethylenedioxythiophene)/carbon nanotube neural electrode coatings for stimulation in the dorsal root ganglion , 2015, Journal of neural engineering.

[6]  Arthur James Lowery,et al.  Characteristics of electrode impedance and stimulation efficacy of a chronic cortical implant using novel annulus electrodes in rat motor cortex , 2013, Journal of neural engineering.

[7]  Greg Worrell,et al.  Long-Term Measurement of Impedance in Chronically Implanted Depth and Subdural Electrodes During Responsive Neurostimulation in Humans , 2013, Brain Stimulation.

[8]  Carl Lagenaur,et al.  Neuroadhesive L1 coating attenuates acute microglial attachment to neural electrodes as revealed by live two-photon microscopy. , 2017, Biomaterials.

[9]  Huanan Zhang,et al.  Nanostructured Coatings for Improved Charge Delivery to Neurons , 2014 .

[10]  Srikanth Vasudevan,et al.  Thiol-ene/acrylate substrates for softening intracortical electrodes. , 2014, Journal of biomedical materials research. Part B, Applied biomaterials.

[11]  R. Wm A Molecular Perspective on Understanding and Modulating the Performance of Chronic Central Nervous System (CNS) Recording Electrodes -- Indwelling Neural Implants: Strategies for Contending with the In Vivo Environment , 2008 .

[12]  D. Kipke,et al.  Insertion shuttle with carboxyl terminated self-assembled monolayer coatings for implanting flexible polymer neural probes in the brain , 2009, Journal of Neuroscience Methods.

[13]  Igor A. Lavrov,et al.  Flexible parylene-based multielectrode array technology for high-density neural stimulation and recording , 2008 .

[14]  O. B. Ozdoganlar,et al.  Chronic tissue response to carboxymethyl cellulose based dissolvable insertion needle for ultra-small neural probes. , 2014, Biomaterials.

[15]  O. B. Ozdoganlar,et al.  An ultra-compliant, scalable neural probe with molded biodissolvable delivery vehicle , 2012, 2012 IEEE 25th International Conference on Micro Electro Mechanical Systems (MEMS).

[16]  A. Schwartz,et al.  High-performance neuroprosthetic control by an individual with tetraplegia , 2013, The Lancet.

[17]  Patrick A Tresco,et al.  Chronic response of adult rat brain tissue to implants anchored to the skull. , 2004, Biomaterials.

[18]  Carl F. Lagenaur,et al.  The surface immobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. , 2011, Biomaterials.

[19]  Jon A. Mukand,et al.  Neuronal ensemble control of prosthetic devices by a human with tetraplegia , 2006, Nature.

[20]  David C. Martin,et al.  A finite-element model of the mechanical effects of implantable microelectrodes in the cerebral cortex , 2005, Journal of neural engineering.

[21]  Diana George,et al.  Removable silicon insertion stiffeners for neural probes using polyethylene glycol as a biodissolvable adhesive , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[22]  Zhanhong Du,et al.  Comprehensive chronic laminar single-unit, multi-unit, and local field potential recording performance with planar single shank electrode arrays , 2015, Journal of Neuroscience Methods.

[23]  Garrett B Stanley,et al.  The impact of chronic blood-brain barrier breach on intracortical electrode function. , 2013, Biomaterials.

[24]  Christian Bénar,et al.  Conducting Polymer Electrodes for Electroencephalography , 2014, Advanced Healthcare Materials.

[25]  James P. Harris,et al.  In vivo deployment of mechanically adaptive nanocomposites for intracortical microelectrodes , 2011, Journal of neural engineering.

[26]  A. Lozano,et al.  Deep Brain Stimulation for Treatment-Resistant Depression , 2005, Neuron.

[27]  Andrew B Schwartz,et al.  Cortical neural prosthetics. , 2004, Annual review of neuroscience.

[28]  R V Bellamkonda,et al.  Agarose gel stiffness determines rate of DRG neurite extension in 3D cultures. , 2001, Biomaterials.

[29]  Justin A. Blanco,et al.  Dissolvable films of silk fibroin for ultrathin conformal bio-integrated electronics. , 2010, Nature materials.

[30]  Paras R. Patel,et al.  Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. , 2012, Nature materials.

[31]  Shashi K. Murthy,et al.  Bridging the Divide between Neuroprosthetic Design, Tissue Engineering and Neurobiology , 2009, Front. Neuroeng..

[32]  Jochen Guck,et al.  Viscoelastic properties of individual glial cells and neurons in the CNS , 2006, Proceedings of the National Academy of Sciences.

[33]  Emrullah Korkmaz,et al.  Ultra-miniature ultra-compliant neural probes with dissolvable delivery needles: design, fabrication and characterization , 2016, Biomedical microdevices.

[34]  A. Michael,et al.  Brain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies , 2014, ACS chemical neuroscience.

[35]  Xiliang Luo,et al.  Poly (3, 4-ethylenedioxythiophene)-ionic liquid coating improves neural recording and stimulation functionality of MEAs. , 2015, Journal of materials chemistry. C.

[36]  H. Lüders,et al.  Deep Brain Stimulation in Epilepsy , 2001, Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society.

[37]  David F Meaney,et al.  Matrices with compliance comparable to that of brain tissue select neuronal over glial growth in mixed cortical cultures. , 2006, Biophysical journal.

[38]  Lisa A Flanagan,et al.  Neurite branching on deformable substrates , 2002, Neuroreport.

[39]  D. Kipke,et al.  Long-term neural recording characteristics of wire microelectrode arrays implanted in cerebral cortex. , 1999, Brain research. Brain research protocols.

[40]  Henrik Jörntell,et al.  Implant Size and Fixation Mode Strongly Influence Tissue Reactions in the CNS , 2011, PloS one.

[41]  D. Szarowski,et al.  Cerebral Astrocyte Response to Micromachined Silicon Implants , 1999, Experimental Neurology.

[42]  P. Tresco,et al.  Response of brain tissue to chronically implanted neural electrodes , 2005, Journal of Neuroscience Methods.

[43]  X Tracy Cui,et al.  Dexamethasone retrodialysis attenuates microglial response to implanted probes in vivo. , 2016, Biomaterials.

[44]  Giuliano Iurilli,et al.  Flexible, all-polymer microelectrode arrays for the capture of cardiac and neuronal signals. , 2011, Biomaterials.

[45]  Brian Hyland,et al.  Cortical effects of subthalamic stimulation correlate with behavioral recovery from dopamine antagonist induced akinesia. , 2009, Cerebral cortex.

[46]  J. Muthuswamy,et al.  Compliant intracortical implants reduce strains and strain rates in brain tissue in vivo , 2015, Journal of neural engineering.

[47]  Wei Wang,et al.  Collaborative Approach in the Development of High‐Performance Brain–Computer Interfaces for a Neuroprosthetic Arm: Translation from Animal Models to Human Control , 2014, Clinical and translational science.

[48]  D. Kipke,et al.  Neural probe design for reduced tissue encapsulation in CNS. , 2007, Biomaterials.

[49]  James P. Harris,et al.  Mechanically adaptive intracortical implants improve the proximity of neuronal cell bodies , 2011, Journal of neural engineering.

[50]  Daryl R Kipke,et al.  Complex impedance spectroscopy for monitoring tissue responses to inserted neural implants , 2007, Journal of neural engineering.

[51]  J. Hetke,et al.  Surface modification of neural recording electrodes with conducting polymer/biomolecule blends. , 2001, Journal of biomedical materials research.

[52]  D W Moran,et al.  A chronic generalized bi-directional brain–machine interface , 2011, Journal of neural engineering.

[53]  M. Wadhwa,et al.  Detection of interleukin-1 and interleukin-6 in adult rat brain, following mechanical injury, by in vivo microdialysis: evidence of a role for microglia in cytokine production , 1991, Journal of Neuroimmunology.

[54]  A. Vazquez,et al.  Two-photon imaging of chronically implanted neural electrodes: Sealing methods and new insights , 2016, Journal of Neuroscience Methods.

[55]  Sanjay Anand,et al.  The upregulation of specific interleukin (IL) receptor antagonists and paradoxical enhancement of neuronal apoptosis due to electrode induced strain and brain micromotion. , 2012, Biomaterials.

[56]  Vikash Gilja,et al.  Long-term Stability of Neural Prosthetic Control Signals from Silicon Cortical Arrays in Rhesus Macaque Motor Cortex , 2010 .

[57]  Robin C. Ashmore,et al.  An Electrocorticographic Brain Interface in an Individual with Tetraplegia , 2013, PloS one.

[58]  M L Boninger,et al.  Ten-dimensional anthropomorphic arm control in a human brain−machine interface: difficulties, solutions, and limitations , 2015, Journal of neural engineering.

[59]  X. Tracy Cui,et al.  In Vivo Electrochemical Analysis of a PEDOT/MWCNT Neural Electrode Coating , 2015, Biosensors.

[60]  Victor Pikov,et al.  Correlations between histology and neuronal activity recorded by microelectrodes implanted chronically in the cerebral cortex , 2016, Journal of neural engineering.

[61]  Jochen Guck,et al.  The relationship between glial cell mechanosensitivity and foreign body reactions in the central nervous system. , 2014, Biomaterials.

[62]  David C. Martin,et al.  In vivo studies of polypyrrole/peptide coated neural probes. , 2003, Biomaterials.

[63]  Eva Syková,et al.  Diffusion barriers evoked in the rat cortex by reactive astrogliosis , 1999, Glia.

[64]  Karen L. Smith,et al.  Effects of insertion conditions on tissue strain and vascular damage during neuroprosthetic device insertion , 2006, Journal of neural engineering.

[65]  J. Melvin,et al.  Dynamic mechanical properties of human brain tissue. , 1969, Journal of biomechanics.

[66]  P. Janmey,et al.  Biomechanics and Mechanotransduction in Cells and Tissues Cell type-specific response to growth on soft materials , 2005 .

[67]  Wolfgang Eberle,et al.  Synaptic dysfunction in hippocampus of transgenic mouse models of Alzheimer's disease: A multi-electrode array study , 2011, Neurobiology of Disease.

[68]  M. Dembo,et al.  Cell movement is guided by the rigidity of the substrate. , 2000, Biophysical journal.

[69]  K Miller,et al.  Mechanical properties of brain tissue in-vivo: experiment and computer simulation. , 2000, Journal of biomechanics.

[70]  Euisik Yoon,et al.  Chronic In Vivo Evaluation of PEDOT/CNT for Stable Neural Recordings , 2016, IEEE Transactions on Biomedical Engineering.

[71]  A. Benabid,et al.  Effect on parkinsonian signs and symptoms of bilateral subthalamic nucleus stimulation , 1995, The Lancet.

[72]  Jessica K. Nguyen,et al.  Mechanically-compliant intracortical implants reduce the neuroinflammatory response , 2014, Journal of neural engineering.

[73]  G. Pfurtscheller,et al.  Brain-Computer Interfaces for Communication and Control. , 2011, Communications of the ACM.

[74]  D J Weber,et al.  In vivo effects of L1 coating on inflammation and neuronal health at the electrode-tissue interface in rat spinal cord and dorsal root ganglion. , 2012, Acta biomaterialia.

[75]  X Tracy Cui,et al.  Effects of caspase-1 knockout on chronic neural recording quality and longevity: insight into cellular and molecular mechanisms of the reactive tissue response. , 2014, Biomaterials.

[76]  Huanan Zhang,et al.  Chronic in vivo stability assessment of carbon fiber microelectrode arrays , 2016, Journal of neural engineering.

[77]  Carl F. Lagenaur,et al.  Elastomeric and soft conducting microwires for implantable neural interfaces. , 2015, Soft matter.

[78]  Andrew B. Schwartz,et al.  Brain-Controlled Interfaces: Movement Restoration with Neural Prosthetics , 2006, Neuron.

[79]  A T Barker,et al.  Surface electromyography using electrode arrays: A study of motor neuron disease , 2001, Muscle & nerve.

[80]  R. Andersen,et al.  Cognitive neural prosthetics. , 2010, Annual review of psychology.

[81]  Huanan Zhang,et al.  Insertion of linear 8.4 μm diameter 16 channel carbon fiber electrode arrays for single unit recordings , 2015, Journal of neural engineering.

[82]  Takashi D Y Kozai,et al.  Photoelectric artefact from optogenetics and imaging on microelectrodes and bioelectronics: New Challenges and Opportunities. , 2015, Journal of materials chemistry. B.

[83]  C. McIntyre,et al.  Sources and effects of electrode impedance during deep brain stimulation , 2006, Clinical Neurophysiology.

[84]  Patrick A Tresco,et al.  The brain tissue response to implanted silicon microelectrode arrays is increased when the device is tethered to the skull. , 2007, Journal of biomedical materials research. Part A.

[85]  Kasey Catt,et al.  Mechanical failure modes of chronically implanted planar silicon-based neural probes for laminar recording. , 2015, Biomaterials.

[86]  L. Rubin,et al.  The cell biology of the blood-brain barrier. , 1999, Annual review of neuroscience.

[87]  D. Tyler,et al.  Stimuli-Responsive Polymer Nanocomposites Inspired by the Sea Cucumber Dermis , 2008, Science.

[88]  Justin C. Williams,et al.  Flexible polyimide-based intracortical electrode arrays with bioactive capability , 2001, IEEE Transactions on Biomedical Engineering.

[89]  D. Kipke,et al.  Repeated voltage biasing improves unit recordings by reducing resistive tissue impedances , 2005, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[90]  David C. Martin,et al.  Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays , 2005, Experimental Neurology.

[91]  Michelle C LaPlaca,et al.  High rate shear strain of three-dimensional neural cell cultures: a new in vitro traumatic brain injury model. , 2005, Journal of biomechanics.

[92]  Jens Schouenborg,et al.  Influence of Probe Flexibility and Gelatin Embedding on Neuronal Density and Glial Responses to Brain Implants , 2015, PloS one.

[93]  P. Leleux,et al.  Highly Conformable Conducting Polymer Electrodes for In Vivo Recordings , 2011, Advanced materials.

[94]  Y. Agid,et al.  Subthalamic nucleus stimulation in severe obsessive-compulsive disorder. , 2008, The New England journal of medicine.

[95]  P. Krack,et al.  Deep-brain stimulation of the subthalamic nucleus or the pars interna of the globus pallidus in Parkinson's disease. , 2001, The New England journal of medicine.

[96]  D R Kipke,et al.  Reduction of neurovascular damage resulting from microelectrode insertion into the cerebral cortex using in vivo two-photon mapping , 2010, Journal of neural engineering.

[97]  Jerald D. Kralik,et al.  Chronic, multisite, multielectrode recordings in macaque monkeys , 2003, Proceedings of the National Academy of Sciences of the United States of America.

[98]  J. Donoghue,et al.  Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates , 2013, Journal of neural engineering.

[99]  Daryl R. Kipke,et al.  Wireless implantable microsystems: high-density electronic interfaces to the nervous system , 2004, Proceedings of the IEEE.

[100]  Seong-Gi Kim,et al.  In vivo two-photon microscopy reveals immediate microglial reaction to implantation of microelectrode through extension of processes , 2012, Journal of neural engineering.

[101]  David C. Martin,et al.  Electrochemical deposition and characterization of poly(3,4-ethylenedioxythiophene) on neural microelectrode arrays , 2003 .

[102]  Karl A. Sillay,et al.  Long‐Term Measurement of Therapeutic Electrode Impedance in Deep Brain Stimulation , 2010, Neuromodulation : journal of the International Neuromodulation Society.

[103]  R. Bellamkonda,et al.  Biomechanical analysis of silicon microelectrode-induced strain in the brain , 2005, Journal of neural engineering.

[104]  Matthew D. Johnson,et al.  In vivo impedance spectroscopy of deep brain stimulation electrodes , 2009, Journal of neural engineering.