The results are reported for an APMP.R(I)-K1.1 comparison that extends the regional comparison of standards for air kerma APMP.R(I)-K1 to several laboratories unable to participate earlier. The comparison was conducted with the goal of supporting the relevant calibration and measurement capabilities (CMCs) planned for publication by the participant laboratories. The comparison was conducted by the pilot laboratory, the Australian Radiation Protection and Nuclear Safety (ARPANSA), Australia, supported by the Institute of Nuclear Energy Research (INER), Taiwan, in a modified ring-shaped arrangement from September 2009 to November 2010, in parallel with an APMP.R(I)-K4 comparison being piloted by the INER. The laboratories that took part in the comparison were the ARPANSA, the Centre of Technology of Radiation Safety and Metrology (PTKMR-BATAN), Indonesia, the Division of Radiation and Medical Devices (DMSC), Thailand, the INER, the National Centre for Radiation Science (NCRS), New Zealand, the National Institute for Standards (NIS), Egypt and the National Metrology Institute of Japan (NMIJ/AIST), Japan. The two primary laboratories, ARPANSA and NMIJ, were chosen as the linking laboratories. Three ionization chambers were used as transfer instruments to be calibrated in terms of air kerma in 60Co radiotherapy beams. The comparison result is based on the ratio between the air kerma calibration coefficients (NK ) determined by the participants and the mean of the results of the linking laboratories. The mean comparison ratio was found to be within 0.5 % of the key comparison reference value KCRV. The largest deviation between any two comparison ratios for the three chambers in terms of air kerma was 2.0 %. An analysis of the participant uncertainty budgets enabled the calculation of degrees of equivalence (DoE) in terms of the deviations of the results and their associated uncertainties. As a result of this APMP comparison, the BIPM key comparison database (KCDB) should include three new entries since neither the PTKMR-BATAN nor the NCRS have yet been declared designated institutes and consequently their results cannot be entered. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).