Effect of Alteration in Pump Speed on Pump Output and Left Ventricular Filling with Continuous-Flow Left Ventricular Assist Device

Third-generation continuous-flow left ventricular assist devices (LVAD) provide reduced pulsatility flow. We examined the safe working range for LVAD pump speed and the effect on pump output and cardiac function in 13 stable outpatients with VentrAssist-LVAD (Ventracor Ltd, Australia). Pump speed was decreased from a baseline mean of 2,073 ± 86 revolutions per minute (RPM, with corresponding mean flow of 5.59 ± 1.18 L/min, mean ± standard deviation) to an average low-speed of 1,835 ± 55 RPM (corresponding flow 4.68 ± 0.99 L/min) and up to high-speed of 2,315 ± 66 RPM (corresponding flow 6.30 ± 1.29 L/min). There was a strong linear relationship between alteration in speed and flow rates (r2 = 0.89, p < 0.00001) but marked interpatient variation. Downward titration to preset minimum 1,800 RPM was achieved in 9/13 (69%) and upward titration to the preset maximum 2,400 RPM was achieved in 4/13 (31%). Upward titration was stopped due to ventricular suction or nonsustained ventricular tachycardia (VT) in 4/13 (31%). Ventricular suction or VT (in 4/13) tended to be more common in patients with poor right ventricular (RV) function (p = 0.07). In summary, pump flow is stable within a relatively small speed range and should not be altered without close monitoring due to variation in response between patients, particularly with concomitant RV impairment.

[1]  Andrew Boyle,et al.  Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular function. , 2010, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[2]  A. Wakatsuki,et al.  Statin attenuates increase in C-reactive protein during estrogen replacement therapy in postmenopausal women. , 2002, Circulation.

[3]  Yukihiko Nosé,et al.  Flow characteristics and required control algorithm of an implantable centrifugal left ventricular assist device , 2008, Heart and Vessels.

[4]  Robert L Kormos,et al.  Second INTERMACS annual report: more than 1,000 primary left ventricular assist device implants. , 2010, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[5]  David Kaye,et al.  A prospective, multicenter trial of the VentrAssist left ventricular assist device for bridge to transplant: safety and efficacy. , 2008, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[6]  David Kaye,et al.  Initial clinical experience with the VentrAssist left ventricular assist device: the pilot trial. , 2008, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[7]  J. Maessen,et al.  Suction due to left ventricular assist: implications for device control and management. , 2007, Artificial organs.

[8]  F. Pagani,et al.  The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. , 2008, Journal of the American College of Cardiology.

[9]  Nigel H. Lovell,et al.  Noninvasive Average Flow and Differential Pressure Estimation for an Implantable Rotary Blood Pump Using Dimensional Analysis , 2008, IEEE Transactions on Biomedical Engineering.

[10]  Kiyotaka Fukamachi,et al.  Predictors of Severe Right Ventricular Failure After Implantable Left Ventricular Assist Device Insertion: Analysis of 245 Patients , 2002, Circulation.

[11]  Nader Moazami,et al.  Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device. , 2009, Journal of the American College of Cardiology.

[12]  Seongjin Choi,et al.  Hemodynamic controller for left ventricular assist device based on pulsatility ratio. , 2007, Artificial organs.

[13]  J F Antaki,et al.  Rotary blood pump flow spontaneously increases during exercise under constant pump speed: results of a chronic study. , 1999, Artificial organs.

[14]  Mathew Williams,et al.  Assessment of synchrony relationships between the native left ventricle and the HeartMate left ventricular assist device. , 2002, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[15]  S. Russell,et al.  Advanced heart failure treated with continuous-flow left ventricular assist device. , 2009, The New England journal of medicine.