A 90 element CdTe array detector

Abstract The fabrication of a CdTe array radiation detector and its radiation detection characteristics are described. In order to obtain high efficiency of charge collection and realize uniform detection sensitivity, current-voltage characteristics with the combination of large and small barrier height contacts and three kinds of CdTe crystals have been investigated. It was found that the Schottky barrier height of electroless Pt deposition was 0.97 eV, which effectively suppressed electron injection. By using the crystal grown by the travelling heater method with a Cl concentration of 2 ppm, carrier lifetimes for electrons and holes of 1.0 and 0.5 μs, respectively, were achieved. A 90 element array detector exhibited an energy resolution as low as 4.5 keV and a count rate variation of less than 5% for 60 keV γ-rays.