Calculations of the pH changes produced in body tissue by a spherical stimulation electrode

A mathematical description of pH excursions produced in interstitial fluid by a spherical stimulation electrode is presented. The pH is calculated as a function of current density, electrode radius, distance, time, and pulsing regimen for an electrode driven by biphasic current pulses. Calculations indicate that large pH excursions occur around electrodes pulsed at current densities used for neural stimulation. For an electrode with a radius of about 1 μm these transient pH changes extend only a few micrometers from the electrode surface. The practical importance of these pH changes remains to be determined.

[1]  F. Albert Cotton,et al.  Advanced Inorganic Chemistry: A Comprehensive Text , 1972 .

[2]  Calculations of temperature rise produced in body tissue by a spherical electrode , 2006, Annals of Biomedical Engineering.

[3]  J. Edsall Carbon Dioxide, Carbonic Acid and Bicarbonate Ion: Physical Properties and Kinetics of Interconversion , 1969 .

[4]  L. Welt Clinical disorders of hydration and acid-base equilibrium . Welt , 1961 .

[5]  S. B. Brummer,et al.  Activated Ir: An Electrode Suitable for Reversible Charge Injection in Saline Solution , 1983 .

[6]  L L Hench,et al.  An in vitro and in vivo analysis of anodized tantalum capacitive electrodes: corrosion response, physiology, and histology. , 1977, Journal of biomedical materials research.

[7]  L L Hench,et al.  An in vitro analysis of metal electrodes for use in the neural environment. , 1977, Brain, behavior and evolution.

[8]  William F. Agnew,et al.  Histological Evaluation of Neural Damage from Electrical Stimulation , 1981 .

[9]  J. C. Jaeger,et al.  Conduction of Heat in Solids , 1952 .

[10]  David L. Guyton,et al.  Theory and design of capacitor electrodes for chronic stimulation , 1974, Medical and biological engineering.

[11]  S. B. Brummer,et al.  Electrochemical Considerations for Safe Electrical Stimulation of the Nervous System with Platinum Electrodes , 1977, IEEE Transactions on Biomedical Engineering.

[12]  J. R. Hughes,et al.  Brief, noninjurious electric waveform for stimulation of the brain. , 1955, Science.

[13]  R. A. Robinson,et al.  Electrolyte solutions : the measurement and interpretation of conductance, chemical potential and diffusion in solutions of simple electrolytes , 1955 .

[14]  D. McCreery,et al.  Histological evaluation of neural damage from electrical stimulation: considerations for the selection of parameters for clinical application. , 1981, Neurosurgery.