Quality assurance for a six degrees‐of‐freedom table using a 3D printed phantom

Abstract Purpose To establish a streamlined end‐to‐end test of a 6 degrees‐of‐freedom (6DoF) robotic table using a 3D printed phantom for periodic quality assurance. Methods A 3D printed phantom was fabricated with translational and rotational offsets and an imbedded central ball‐bearing (BB). The phantom underwent each step of the radiation therapy process: CT simulation in a straight orientation, plan generation using the treatment planning software, setup to offset marks at the linac, registration and corrected 6DoF table adjustments via hidden target test, delivery of a Winston‐Lutz test to the BB, and verification of table positioning via field and laser lights. The registration values, maximum total displacement of the combined Winston‐Lutz fields, and a pass or fail criterion of the laser and field lights were recorded. The quality assurance process for each of the three linacs were performed for the first 30 days. Results Within a 95% confidence interval, the overall uncertainty values for both translation and rotation were below 1.0 mm and 0.5° for each linac respectively. When combining the registration values and other uncertainties for all three linacs, the average deviations were within 2.0 mm and 1.0° of the designed translation and rotation offsets of the 3D print respectively. For all three linacs, the maximum total deviation for the Winston‐Lutz test did not exceed 1.0 mm. Laser and light field verification was within tolerance every day for all three linacs given the latest guidance documentation for table repositioning. Conclusion The 3D printer is capable of accurately fabricating a quality assurance phantom for 6DoF positioning verification. The end‐to‐end workflow allows for a more efficient test of the 6DoF mechanics while including other important tests needed for routine quality assurance.

[1]  Thierry Gevaert,et al.  Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery. , 2012, International journal of radiation oncology, biology, physics.

[2]  H. Dodziuk,et al.  Applications of 3D printing in healthcare , 2016, Kardiochirurgia i torakochirurgia polska = Polish journal of cardio-thoracic surgery.

[3]  Tatiana Kelil,et al.  3D printing from MRI Data: Harnessing strengths and minimizing weaknesses , 2017, Journal of magnetic resonance imaging : JMRI.

[4]  Jun Won Kim,et al.  Clinical application of 3D-printed-step-bolus in post-total-mastectomy electron conformal therapy , 2016, Oncotarget.

[5]  T Mizowaki,et al.  The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch system , 2010, Physics in medicine and biology.

[6]  G. Baroni,et al.  3D-printed applicators for high dose rate brachytherapy: Dosimetric assessment at different infill percentage. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[7]  James McInerney,et al.  The use of custom 3D printed stereotactic frames for laser interstitial thermal ablation: technical note. , 2016, Neurosurgical focus.

[8]  Michael K Fix,et al.  Assessment of patient setup errors in IGRT in combination with a six degrees of freedom couch. , 2014, Zeitschrift fur medizinische Physik.

[9]  C. Iannuzzi,et al.  Study of ExacTrac X‐ray 6D IGRT setup uncertainty for marker‐based prostate IMRT treatment , 2012, Journal of applied clinical medical physics.

[10]  K. Winston,et al.  A system for stereotactic radiosurgery with a linear accelerator. , 1986, International journal of radiation oncology, biology, physics.

[11]  Paul Keall,et al.  Six degrees-of-freedom prostate and lung tumor motion measurements using kilovoltage intrafraction monitoring. , 2015, International journal of radiation oncology, biology, physics.

[12]  M. Schell,et al.  Stereotactic body radiation therapy: the report of AAPM Task Group 101. , 2010, Medical physics.

[13]  Tom Depuydt,et al.  Towards 3D printed multifunctional immobilization for proton therapy: Initial materials characterization. , 2016, Medical physics.

[14]  Philip Tack,et al.  3D-printing techniques in a medical setting: a systematic literature review , 2016, BioMedical Engineering OnLine.

[15]  Fang-Fang Yin,et al.  Task Group 142 report: quality assurance of medical accelerators. , 2009, Medical physics.

[16]  L. Xing,et al.  Overview of image-guided radiation therapy. , 2006, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[17]  Tae-Suk Suh,et al.  A depth-sensing technique on 3D-printed compensator for total body irradiation patient measurement and treatment planning. , 2016, Medical physics.

[18]  A. Takemura,et al.  Comparison of the Motion Accuracy of a Six Degrees of Freedom Radiotherapy Couch with and without Weights , 2013 .

[19]  Ken Goldberg,et al.  Evaluation of PC‐ISO for customized, 3D printed, gynecologic 192Ir HDR brachytherapy applicators , 2015, Journal of applied clinical medical physics.

[20]  T. Zhu,et al.  Should image rotation be addressed during routine cone-beam CT quality assurance? , 2013, Physics in medicine and biology.

[21]  Martijn Kusters,et al.  Clinical implementation of 3D printing in the construction of patient specific bolus for electron beam radiotherapy for non-melanoma skin cancer. , 2016, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[22]  P. Munro,et al.  Evaluation of a new six degrees of freedom couch for radiation therapy. , 2013, Medical physics.

[23]  Fang-Fang Yin,et al.  Use of the BrainLAB ExacTrac X-Ray 6D system in image-guided radiotherapy. , 2008, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[24]  H. Shu,et al.  Six degrees of freedom CBCT‐based positioning for intracranial targets treated with frameless stereotactic radiosurgery , 2012, Journal of applied clinical medical physics.

[25]  Weiliang Du,et al.  A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator , 2010, Journal of applied clinical medical physics.

[26]  Matthew F Bieniosek,et al.  Technical Note: Characterization of custom 3D printed multimodality imaging phantoms. , 2015, Medical physics.

[27]  Vincenzo Valentini,et al.  Adaptive optimization by 6 DOF robotic couch in prostate volumetric IMRT treatment: rototranslational shift and dosimetric consequences , 2015, Journal of applied clinical medical physics.

[28]  C. L. Ventola Medical Applications for 3D Printing: Current and Projected Uses. , 2014, P & T : a peer-reviewed journal for formulary management.

[29]  Rao Khan,et al.  Characterizing 3D printing in the fabrication of variable density phantoms for quality assurance of radiotherapy. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[30]  Patrick D Higgins,et al.  Patient specific 3D printed phantom for IMRT quality assurance , 2014, Physics in medicine and biology.

[31]  Cristiano J. Miosso,et al.  Use of 3D-printers to create Intensity-Modulated Radiotherapy Compensator Blocks , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[32]  Shiqin Su,et al.  Design and production of 3D printed bolus for electron radiation therapy , 2014, Journal of applied clinical medical physics.

[33]  Matthias Guckenberger,et al.  Precision of Image-Guided Radiotherapy (IGRT) in Six Degrees of Freedom and Limitations in Clinical Practice , 2007, Strahlentherapie und Onkologie.

[34]  Dong-Woo Cho,et al.  Computer-aided multiple-head 3D printing system for printing of heterogeneous organ/tissue constructs , 2016, Scientific Reports.