A 7×7×2mm3 8.6-μ 500-kb/s Transmitter with Robust Injection-Locking Based Frequency-to-Amplitude Conversion Receiver Targeting for Implantable Applications

This paper presents the first transmitter (TX)/receiver (RX) pair using galvanic coupling to achieve In-body to On-body Body Channel Communication (I2O-BCC) for ultra-low power (ULP) miniaturized implanted applications. The measured channel loss characteristic is smaller than 40dB within $10 \sim 40$ MHz band when the implanted depth is no more than 10cm. Injection locking (IL) based frequency-to-amplitude (F2A) conversion demodulator is proposed in RX to handle large frequency variations of TX FSK signals due to the lack of off-chip reference. The implanted TX only requires two off-chip components, micro battery and tiny probes, achieving packaged size of $7 \times 7 \times 2$ mm3. The in-body TX consumes only $8.6 \mu \mathrm {W}$ at 500 kb/s while the on-body RX consumes $591.5 \mu \mathrm {W}$. The TX power are $8 \times $ lower than previous works and achieves energy efficiency at 17.2pJ/b with minimal bill of material (BOM).

[1]  Mohamad Sawan,et al.  Implantable MICS-based wireless solution for bladder pressure monitoring , 2017, 2017 IEEE Biomedical Circuits and Systems Conference (BioCAS).

[2]  Yu-Wen Huang,et al.  21.3 A sub-mm3 wireless implantable intraocular pressure monitor microsystem , 2017, 2017 IEEE International Solid-State Circuits Conference (ISSCC).

[3]  David Blaauw,et al.  26.7 A 10mm3 syringe-implantable near-field radio system on glass substrate , 2016, 2016 IEEE International Solid-State Circuits Conference (ISSCC).

[4]  Hoi-Jun Yoo,et al.  A 45 $\mu$W Injection-Locked FSK Wake-Up Receiver With Frequency-to-Envelope Conversion for Crystal-Less Wireless Body Area Network , 2015, IEEE Journal of Solid-State Circuits.

[5]  Teresa H. Y. Meng,et al.  A mm-sized implantable power receiver with adaptive link compensation , 2009, 2009 IEEE International Solid-State Circuits Conference - Digest of Technical Papers.

[6]  David Blaauw,et al.  A cubic-millimeter energy-autonomous wireless intraocular pressure monitor , 2011, 2011 IEEE International Solid-State Circuits Conference.

[7]  Shreyas Sen,et al.  A 6.3pJ/b 30Mbps −30dB SIR-tolerant broadband interference-robust human body communication transceiver using time domain signal-interference separation , 2018, 2018 IEEE Custom Integrated Circuits Conference (CICC).

[8]  Amin Arbabian,et al.  27.7 A 30.5mm3 fully packaged implantable device with duplex ultrasonic data and power links achieving 95kb/s with <10−4 BER at 8.5cm depth , 2017, 2017 IEEE International Solid-State Circuits Conference (ISSCC).